Airis Elite Magnetic Resonance Imaging Device

K032232510(k) AbbreviatedCleared Sep 29, 200370-day reviewClass II21 CFR 892.1000Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hitachi Medical Systems America, Inc. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Sep 29, 2003 as substantially equivalent to K001334 (Hitachi Medical Systems America, Inc., 2000) and 1 other predicate after a 70-day review. Its family tree reaches back 1 generation to 1998. 6 later devices descend from it. It has been recalled 2 times.

Decided
Sep 29, 2003
Review
70 days
Class
Class II
Product code
LNH
Recalls
2
Predicates
2

Its longest line of descent runs back 1 generation to K974212 (1998). 2 ancestors in all. 6 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2005: 1 descendant cleared2024: 2 descendants cleared2026: 3 descendants clearedK974212 · Airis II · Hitachi Medical Systems America, Inc. · 1998 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant K974212 · Airis II · Hitachi Medical Systems America, Inc. · 1998 · recalled0
  1. Root · 1 generation backK974212 Airis IIHitachi Medical Systems America, Inc. · 1998 · Recalled
  2. This deviceK032232 Airis Elite Magnetic Resonance Imaging DeviceHitachi Medical Systems America, Inc. · 2003
  3. Descendant · cites it directly · 2 of its ownK233629 APERTO Lucent MRI SystemFUJIFILM Healthcare Americas Corporation · 2024
  4. Descendant · 2 generations on · 1 of its ownK253862 APERTO Lucent MRI SystemFujifilm Corporation · 2026
  5. Descendant · cites it directly · 1 of its ownK241133 Magnifico Open (100009900); Magnifico MSK ( 100009910)Esaote, S.p.A. · 2024
  6. Descendant · 3 generations on · 0 of its ownK261283 APERTO Lucent MRI SystemFujifilm Corporation · 2026
  7. Descendant · 2 generations on · 0 of its ownK251901 Magnifico Open (100009900)Esaote, S.p.A. · 2026

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2005: 12024: 22026: 3
2005high 3 (2026) · low 0 (2006)2026

Most influential descendants

Showing 5 of 6 · See all
  • K233629APERTO Lucent MRI SystemFUJIFILM Healthcare Americas Corporation · 2024 · cites it directly2descendants
  • K253862APERTO Lucent MRI SystemFujifilm Corporation · 2026 · 2 generations on1descendant
  • K241133Magnifico Open (100009900); Magnifico MSK ( 100009910)Esaote, S.p.A. · 2024 · cites it directly1descendant
  • K261283APERTO Lucent MRI SystemFujifilm Corporation · 2026 · 3 generations on0descendants
  • K251901Magnifico Open (100009900)Esaote, S.p.A. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K974212Airis IIHitachi Medical Systems America, Inc.19985
0 (this device)K032232Airis Elite Magnetic Resonance Imaging DeviceHitachi Medical Systems America, Inc.2003—
+1K233629APERTO Lucent MRI SystemFUJIFILM Healthcare Americas Corporation2024—
+2K253862APERTO Lucent MRI SystemFujifilm Corporation2026—
+1K241133Magnifico Open (100009900); Magnifico MSK ( 100009910)Esaote, S.p.A.2024—
+3K261283APERTO Lucent MRI SystemFujifilm Corporation2026—
+2K251901Magnifico Open (100009900)Esaote, S.p.A.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
202420
202630
FIG. 1 — Lineage of K032232, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001334Hitachi Airis IIHitachi Medical Systems America, Inc.20003
*K974212Airis II
Summary page 2
…ification Number: 90LNH Trade/Proprietary Name: AIRIS Elite Predicate Device(s): AIRIS I1 (KO01334, K9742 12) 3.0 Device Intended Use The MR system is an imaging device and is intended to provide the physician with physiologi…
Hitachi Medical Systems America, Inc.19985

Also cleared against K001334: 7 other devices →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233629APERTO Lucent MRI SystemFUJIFILM Healthcare Americas Corporation20242
K241133Magnifico Open (100009900); Magnifico MSK ( 100009910)Esaote, S.p.A.20241
K050307Airis Elite Rapid Body CoilHitachi Medical Systems America, Inc.20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 21, 2006Not classifiedZ-0910-06OtherThe device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.
Jun 30, 2005Not classifiedZ-0890-05OtherImage reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K233629 (FUJIFILM Healthcare Americas Corporation, 2024) has gone furthest, with 2 descendants of its own.
  • 7 other devices also cite its predicate K001334.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The AIRIS Elite system is an MR imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, obliq,ue, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the system reflect the spatial distribution of protons (hydrogen…

Read the full statement (135 words)

The AIRIS Elite system is an MR imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, obliq,ue, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (TI), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities

Nucleus excited: Proton

Diagnostic uses: 2D,’3D T1, T2-weighted imaging

T1, T2 proton density measurements

Diffusion weighted imaging

MF, Angiography

Image processing

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K032232: Airis Elite Magnetic Resonance Imaging Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K032232

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026