510(k) Report

510(k) · Cleared

Recalled · 2006

Airis II

K974212510(k) TraditionalCleared Feb 9, 199891-day reviewClass II21 CFR 892.1000Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hitachi Medical Systems America, Inc. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Feb 9, 1998 after a 91-day review. The FDA summary names its predicate without a number. 32 later devices descend from it. It has been recalled 5 times.

Decided
Feb 9, 1998
Review
91 days
Class
Class II
Product code
LNH
Recalls
5
Predicates
0

32 later devices descend from it, 10 citing it directly.

Descendants, by year cleared 2000: 1 descendant cleared1998: 1 descendant cleared2009: 5 descendants cleared2003: 4 descendants cleared1999: 3 descendants cleared2006: 5 descendants cleared2007: 2 descendants cleared2005: 1 descendant cleared2024: 2 descendants cleared2014: 1 descendant cleared2008: 2 descendants cleared2026: 3 descendants cleared2002: 1 descendant cleared2011: 1 descendant cleared1995200020052010201520202025
  1. This deviceK974212 Airis IIHitachi Medical Systems America, Inc. · 1998
  2. Descendant · cites it directly · 8 of its ownK030918 Superopen 0.35T, Model NSM-P035Neusoft Digital Medical Systems Co., Ltd. · 2003
  3. Descendant · cites it directly · 8 of its ownK984278 Magnetic Resonance Diagnostic Device, Airis IIHitachi Medical Systems America, Inc. · 1998
  4. Descendant · 2 generations on · 7 of its ownK992650 Airis, Airis IIHitachi Medical Systems America, Inc. · 1999
  5. Descendant · cites it directly · 6 of its ownK032232 Airis Elite Magnetic Resonance Imaging DeviceHitachi Medical Systems America, Inc. · 2003 · Recalled
  6. Descendant · 3 generations on · 5 of its ownK013854 Mark 5000 Quadrature Shoulder CoilUsa Instruments, Inc. · 2002

Showing 5 of 32 descendants. See all descendants

Descendants cleared per year

1998: 11999: 32000: 12002: 12003: 42005: 12006: 52007: 22008: 22009: 52011: 12014: 12024: 22026: 3
1998high 5 (2006) · low 0 (2001)2026

Most influential descendants

Showing 5 of 32 · See all
  • K030918Superopen 0.35T, Model NSM-P035Neusoft Digital Medical Systems Co., Ltd. · 2003 · cites it directly8descendants
  • K984278Magnetic Resonance Diagnostic Device, Airis IIHitachi Medical Systems America, Inc. · 1998 · cites it directly8descendants
  • K992650Airis, Airis IIHitachi Medical Systems America, Inc. · 1999 · 2 generations on7descendants
  • K032232Airis Elite Magnetic Resonance Imaging DeviceHitachi Medical Systems America, Inc. · 2003 · cites it directly6descendants
  • K013854Mark 5000 Quadrature Shoulder CoilUsa Instruments, Inc. · 2002 · 3 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K974212Airis IIHitachi Medical Systems America, Inc.1998—
+1K030918Superopen 0.35T, Model NSM-P035Neusoft Digital Medical Systems Co., Ltd.2003—
+1K984278Magnetic Resonance Diagnostic Device, Airis IIHitachi Medical Systems America, Inc.1998—
+2K992650Airis, Airis IIHitachi Medical Systems America, Inc.1999—
+1K032232Airis Elite Magnetic Resonance Imaging DeviceHitachi Medical Systems America, Inc.20032
+3K013854Mark 5000 Quadrature Shoulder CoilUsa Instruments, Inc.2002—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199810
199930
200010
200210
200340
200510
200650
200720
200820
200950
201110
201410
202420
202630
FIG. 1 — Lineage of K974212, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Hitachi AIRIS
Named without a submission number
—
*—Hitachi AIRIS with Version 4.0 Operating System Software
Named without a submission number
—

What descends from it? (32)

See all 32

10 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K030918Superopen 0.35T, Model NSM-P035Neusoft Digital Medical Systems Co., Ltd.20038
K984278Magnetic Resonance Diagnostic Device, Airis IIHitachi Medical Systems America, Inc.19988
K032232Airis Elite Magnetic Resonance Imaging DeviceHitachi Medical Systems America, Inc.20036
K984280Airis II Magnetic Resonance Diagnostic Device,Interventional MR PackageHitachi Medical Systems America, Inc.19991
K092899Oper Series Open Type Permanent-Magnet Mri System, Model: Oper-0.3, -0.35, -0.4Ningbo Xingaoyi Magnetism Co., Ltd.20090
K090873I Open 0.5TBeijing Wandong Medical Equipment Co., Ltd.20090
K090433I_Open 0.4TBeijing Wandong Medical Equipment Co., Ltd.20090
K090080Cirrus OpenMri-Tech Sp. Zo.O.20090
K994378Airis IIHitachi Medical Systems America, Inc.20000
K990153Virgo Mri SystemMillennium Technology Services19990

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 28, 2006Not classifiedZ-1104-06OtherSoftware anomaly-A software defect can cause the slice line indicator to become mis-positioned during Multi-Planar Reconstructions (MPR) with the potential for patient mis-diagnosis.
Apr 25, 2006Not classifiedZ-1115-06OtherImproper Assembly-The AIRIS/AIRIS II MRI system has the potential for the input wiring of a power transformer to overheat due to damaged wiring or possible loose connections.
Feb 21, 2006Not classifiedZ-0910-06OtherThe device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.
Dec 1, 2005Not classifiedZ-0429-06OtherThe bolts holding the lift mechanism may be damaged causing them to fail and the table top to drop several inches.
Jun 30, 2005Not classifiedZ-0889-05OtherImage reconstruction software problem that could potentially cause multiple patient images to be stored in one patient folder.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K030918 (Neusoft Digital Medical Systems Co., Ltd., 2003) has gone furthest, with 8 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei)…

Read the full statement (188 words)

The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.

Anatomical Region: Head, Body, Spine, Extremities

Nucleus excited: Proton

Diagnostic uses: 2D T1-/T2-weighted imaging

T1, T2, proton density measurements

MR Angiography

image processing

Imaging capabilities:

2D, 3D Spin Echo (SE)

2D, 3D Fast Spin Echo (FSE)

2D Inversion Recovery (IR)

2D, 3D Fast Inversion Recovery (FIR)

2D,3D Gradient Echo (GE)

2D, 3D Steady state acquisition with rewinded GE (SARGE)

2D, 3D RF-spoiled SARGE (RSSG)

2D, 3D Time-reversed SARGE (TRSG)

MR Angiography

2D, 3D TOF

2D, 3D TOF RSSG

ECG, Peripheral Gating

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K974212: Airis II.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K974212

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026