510(k) Report

510(k) · Cleared

Recalled · 2011

Powerheart Aed G3A

K040438510(k) TraditionalCleared Jul 1, 2004133-day reviewClass III21 CFR 870.5310Recalls 25

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cardiac Science, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Jul 1, 2004 as substantially equivalent to K031987 (Cardiac Science, 2003) and 4 other predicates after a 133-day review. Its family tree reaches back 6 generations to 1997. 6 later devices descend from it. It has been recalled 25 times.

Decided
Jul 1, 2004
Review
133 days
Class
Class III
Product code
MKJ
Recalls
25
Predicates
5

Its longest line of descent runs back 6 generations to K970741 (1997). 12 ancestors in all, the oldest K955628 (1996). 6 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 3 other ancestors1999: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor2011: 1 descendant cleared2014: 2 descendants cleared2012: 1 descendant cleared2008: 1 descendant cleared2015: 1 descendant clearedK970741 · Powerheart Aecd · Cardiac Science, Inc. · 1997 · recalledK982710 · Aecd Arrhythmia Detection Software · Cardiac Science, Inc. · 1998 · recalledK993533 · Powerheart Aecd · Cardiac Science, Inc. · 2000 · recalledK011901 · Firstsave Star Biphasic, Model 9210/9200 · Survivalink Corp. · 2002 · recalledK022929 · Pediatric Attenuated Defibrillation Electrodes, Model 9730, Powederheart Aed, Model 9200/9210 · Cardiac Science, Inc. · 2003K031987 · Powerheart Aed G3, Model 9300 · Cardiac Science · 2003 · recalled1995200020052010201520202025
  1. Root · 6 generations backK970741 Powerheart AecdCardiac Science, Inc. · 1997 · Recalled
  2. 5 generations backK982710 Aecd Arrhythmia Detection SoftwareCardiac Science, Inc. · 1998 · Recalled
  3. 4 generations backK993533 Powerheart AecdCardiac Science, Inc. · 2000 · Recalled
  4. 3 generations backK011901 Firstsave Star Biphasic, Model 9210/9200Survivalink Corp. · 2002 · Recalled
  5. 2 generations backK022929 Pediatric Attenuated Defibrillation Electrodes, Model 9730…Cardiac Science, Inc. · 2003
  6. 1 generation backK031987 Powerheart Aed G3, Model 9300Cardiac Science · 2003 · Recalled
  7. This deviceK040438 Powerheart Aed G3ACardiac Science, Inc. · 2004
  8. Descendant · cites it directly · 2 of its ownK102496 Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic…Cardiac Science Corporation · 2011 · Recalled
  9. Descendant · 2 generations on · 1 of its ownK122758 Powerheart G5 AedCardiac Science Corporation · 2014
  10. Descendant · 3 generations on · 0 of its ownK143714 Powerheart G5 AEDCardiac Science Corporation · 2015
  11. Descendant · cites it directly · 0 of its ownK141231 ZP9141, ZP9146Y, ZP9146W Non-Rechargeable BatteryZeller Power Products (Battery Beast, LLC) · 2014
  12. Descendant · cites it directly · 0 of its ownK120350 Extended Life BatteryAed Battery Exchange, LLC · 2012

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2008: 12011: 12012: 12014: 22015: 1
2008high 2 (2014) · low 0 (2009)2026

Most influential descendants

Showing 5 of 6 · See all
  • K102496Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic, Powerheart Aed G3 Pro…Cardiac Science Corporation · 2011 · cites it directly2descendants
  • K122758Powerheart G5 AedCardiac Science Corporation · 2014 · 2 generations on1descendant
  • K143714Powerheart G5 AEDCardiac Science Corporation · 2015 · 3 generations on0descendants
  • K141231ZP9141, ZP9146Y, ZP9146W Non-Rechargeable BatteryZeller Power Products (Battery Beast, LLC) · 2014 · cites it directly0descendants
  • K120350Extended Life BatteryAed Battery Exchange, LLC · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K970741Powerheart AecdCardiac Science, Inc.19971
−5K982710Aecd Arrhythmia Detection SoftwareCardiac Science, Inc.19981
−4K993533Powerheart AecdCardiac Science, Inc.20001
−3K011901Firstsave Star Biphasic, Model 9210/9200Survivalink Corp.20022
−2K022929Pediatric Attenuated Defibrillation Electrodes, Model 9730, Powederheart Aed…Cardiac Science, Inc.2003—
−1K031987Powerheart Aed G3, Model 9300Cardiac Science200333
0 (this device)K040438Powerheart Aed G3ACardiac Science, Inc.2004—
+1K102496Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic, Powerheart Aed…Cardiac Science Corporation20112
+2K122758Powerheart G5 AedCardiac Science Corporation2014—
+3K143714Powerheart G5 AEDCardiac Science Corporation2015—
+1K141231ZP9141, ZP9146Y, ZP9146W Non-Rechargeable BatteryZeller Power Products (Battery Beast, LLC)2014—
+1K120350Extended Life BatteryAed Battery Exchange, LLC2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199710
199910
200130

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201110
201210
201420
201510
FIG. 1 — Lineage of K040438, 1997–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MKJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031987Powerheart Aed G3, Model 9300
Summary page 1
…ivalent to the current Powerheart® AED and accessories in commercial distribution that was cleared under premarket 510(k) notifications K022929, K01 1901, 1,982710 and K031987. The reason for this premarket notification is to introduce a fully -automation version of the device_…
Cardiac Science200333
*K022929Pediatric Attenuated Defibrillation Electrodes, Model 9730, Powederheart Aed, Model…
Summary page 1
…device in this submission is equivalent to the current Powerheart® AED and accessories in commercial distribution that was cleared under premarket 510(k) notifications K022929, K01 1901, 1,982710 and K031987. The reason for this premarket notification is to introduce a fully -automation version of the…
Cardiac Science, Inc.2003—
*K011901Firstsave Star Biphasic, Model 9210/9200
Summary page 1
…this submission is equivalent to the current Powerheart® AED and accessories in commercial distribution that was cleared under premarket 510(k) notifications K022929, K01 1901, 1,982710 and K031987. The reason for this premarket notification is to introduce a fully -automation version of the device_…
Survivalink Corp.20022
*K982264Firstsave Biphasic Models 9200 and 9210
Summary page 2
…01/27/2003 Corporation (renamed Powerheart AED) K01 1901 02/01/2002 (Cardiac Science) K982264 01/25/1999 Medtronic Physio- LWEPAK CR PLUS AED KO 11144 12/03/2001 Control Corp. 16…
Survivalink Corp.1999—
*K011144Lifepak 600 Automated External DefibrillatorMedtronic Physio-Control Corp.20019

Also cleared against K031987: 5 other devices →

What descends from it? (6)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102496Powerheart Aed G3 Semi-Automatic, Powerheart Aed G3 Automatic, Powerheart Aed G3 Pro…Cardiac Science Corporation20112
K141231ZP9141, ZP9146Y, ZP9146W Non-Rechargeable BatteryZeller Power Products (Battery Beast, LLC)20140
K120350Extended Life BatteryAed Battery Exchange, LLC20120
K082861Amco Replacement Battery for Powerheart Aed, Non-Rechargeable, Model 2L561; G3…Amco International Manufacturing & Design, Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 30, 2011Not classifiedZ-3265-2011Device DesignCardiac Science Corporation (CSC) has identified two errors, 0x48 and 0x99, that have repeatedly occurred during a device self-test in the G3 installed base over the last 6 years. These self test errors cause the device to enter into a…
Feb 3, 2010Not classifiedZ-1248-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Feb 3, 2010Not classifiedZ-1247-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Feb 3, 2010Not classifiedZ-1249-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Feb 3, 2010Not classifiedZ-1244-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Feb 3, 2010Not classifiedZ-1246-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Feb 3, 2010Not classifiedZ-1243-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Feb 3, 2010Not classifiedZ-1245-2010Nonconforming Material/ComponentA component of the AED device may fail during a rescue attempt and the AED will not be able to deliver therapy.
Nov 13, 2009Not classifiedZ-0765-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0778-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0766-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0768-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0769-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0771-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0775-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0776-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0777-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0767-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0770-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0772-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0773-2010Process controlPotential for devices not to deliver therapy
Nov 13, 2009Not classifiedZ-0774-2010Process controlPotential for devices not to deliver therapy
Oct 13, 2008Not classifiedZ-0921-2009Software designPotential for AED to not deliver defibrillation therapy.
Jan 27, 2006Not classifiedZ-0623-06Process controlSome Powerheart AED Model 9300A may contain a suspect electronic component and there is a remote likelihood that the function of this device may be interrupted during the charge.
Nov 12, 2004Not classifiedZ-0399-05Nonconforming Material/ComponentAEDs may contain an electronic component that does not meet its full operating specifications. The AEDs with this suspect electronic component may not functionn properly and may interfere with the AED's ability to delivery therapy. In the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 12 months on median.
  • 8 of the 12 ancestors have a recall of any class.
  • Of its descendants, K102496 (Cardiac Science Corporation, 2011) has gone furthest, with 2 descendants of its own.
  • 5 other devices also cite its predicate K031987.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The Powerheart® AED G3A is intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response. The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for…

Read the full statement (145 words)

The Powerheart® AED G3A is intended to be used by personnel who have been trained in its operation. The user should be qualified by training in basic life support or other physician-authorized emergency medical response.

The device is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of the ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will charge automatically and advise the operator to deliver therapy.

When a patient is a child or infant up to 8 years of age, or up to 55 lbs (25kg), the Powerheart AED G3 should be used with the Model 9730 Pediatric Attentuated Defibrillation Electrodes. The therapy should not be delayed to determine the patient's exact age or weight.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040438: Powerheart Aed G3A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K040438

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026