Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24, AL23NH28, AL27NH32, AL31NH36, AL35NH40

K040509510(k) SpecialCleared Apr 19, 200452-day reviewClass II21 CFR 876.5540Recalls 48

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spire Biomedical, Inc. · MSD Catheter, Hemodialysis, Implanted

Cleared on Apr 19, 2004 as substantially equivalent to K032061 (Spire Biomedical, Inc., 2003) after a 52-day review. Its family tree reaches back 2 generations to 2002. 1 later device descends from it. It has been recalled 48 times.

Decided
Apr 19, 2004
Review
52 days
Class
Class II
Product code
MSD
Recalls
48
Predicates
1

Its longest line of descent runs back 2 generations to K013160 (2002). 1 later device descends from it, 1 citing it directly.

  1. Root · 2 generations backK013160 Pourchez Xpresso Twin Limen Chronic Hemodialysis CatheterSpire Biomedical, Inc. · 2002
  2. 1 generation backK032061 Decathlon Twin Lumen Hemodialysis with Separated Tips, Models…Spire Biomedical, Inc. · 2003 · Recalled
  3. This deviceK040509 Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24…Spire Biomedical, Inc. · 2004
  4. Descendant · cites it directly · 0 of its ownK063441 15.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24…Spire Biomedical, Inc. · 2007 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K06344115.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24, ALC23SH28, ALC27SH32…Spire Biomedical, Inc. · 2007 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K013160Pourchez Xpresso Twin Limen Chronic Hemodialysis CatheterSpire Biomedical, Inc.2002—
−1K032061Decathlon Twin Lumen Hemodialysis with Separated Tips, Models DE19/24SH…Spire Biomedical, Inc.200348
0 (this device)K040509Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24, AL23NH28…Spire Biomedical, Inc.2004—
+1K06344115.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24, ALC23SH28…Spire Biomedical, Inc.200748

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
FIG. 1 — Lineage of K040509, 2002–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032061Decathlon Twin Lumen Hemodialysis with Separated Tips, Models DE19/24SH, DE23/28SH…
Summary page 1
…keted Devices to Which Substantial Equivalence is Claimed: 1) Spire Biomedical, Inc.'s Decathlon TM (Double-D cross sectional round lumen profile), 510(K) # K032061. Device Description: The polyurethane 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis Catheter is a product line extens…
Spire Biomedical, Inc.200348

Also cleared against K032061: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K06344115.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24, ALC23SH28, ALC27SH32…Spire Biomedical, Inc.20070

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 23, 2009Not classifiedZ-0288-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0289-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0290-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0299-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0301-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0302-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0304-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0306-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0309-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0311-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0312-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0294-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0295-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0296-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0297-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0300-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0303-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0307-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0308-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0314-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0316-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0317-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0318-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0319-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0321-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0330-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0291-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0292-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0293-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0298-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0305-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0310-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0313-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0315-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0320-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0322-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0323-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0324-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0325-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0327-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0328-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0332-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0334-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0335-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0329-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0331-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0333-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0326-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 2 years and passes through 1 company; a new generation was cleared every 1.7 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K063441 (Spire Biomedical, Inc., 2007) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K032061.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

The 15.5Fr Alta LR Twin Lumen Fixed Tip Hemodialysis Catheter is designed for chronic (long-term) hemodialysis and apheresis. It is designed for percutaneous insertion or insertion via cutdown.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040509: Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24, AL23NH28, AL27NH32, AL31NH36, AL35NH40.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K040509

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026