510(k) · Cleared
Recalled · 200915.5 FR Decathlon Gold, Funnel Tips, Coated Hemodialysis Catheter, DFC19SH24, DFC23SH28, DFC27SH32, DFC31SH36, DFC35SH40
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Spire Biomedical, Inc. · MSD Catheter, Hemodialysis, Implanted
Cleared on Jan 29, 2007 as substantially equivalent to K060155 (Spire Biomedical, Inc., 2006) and 4 other predicates after a 77-day review. Its family tree reaches back 2 generations to 1993. It has been recalled 48 times.
- Decided
- Jan 29, 2007
- Review
- 77 days
- Class
- Class II
- Product code
- MSD
- Recalls
- 48
- Predicates
- 5
Its longest line of descent runs back 2 generations to K925626 (1993). 6 ancestors in all. No later device cites it as a predicate yet.
- Root · 2 generations backK925626 Maxima Plus Hollow Fiber OxygenatorMedtronic Vascular · 1993
- 1 generation backK060155 15.5FR Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter…Spire Biomedical, Inc. · 2006 · Recalled
- This deviceK063431 15.5 FR Decathlon Gold, Funnel Tips, Coated Hemodialysis Catheter…Spire Biomedical, Inc. · 2007
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K925626 | Maxima Plus Hollow Fiber Oxygenator | Medtronic Vascular | 1993 | — |
| −1 | K060155 | 15.5FR Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter with… | Spire Biomedical, Inc. | 2006 | 47 |
| 0 (this device) | K063431 | 15.5 FR Decathlon Gold, Funnel Tips, Coated Hemodialysis Catheter, DFC19SH24… | Spire Biomedical, Inc. | 2007 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 2 | 0 |
| 1997 | 1 | 0 |
| 2000 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MSD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MSD device
- #1 K241581 Hemodialysis Catheterscore 12.4 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (5)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K060155 | 15.5FR Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter with Separated TipsSummary page 3…Technological Characteristics Comparison to Predicate Devices: The 15.5Fr Decathlon Gold, coated catheter with funnel tips is identical to the Coated Decathlon catheter (K060155) in physical characteristics, namely, materials of construction, dimensions, and the coating. The only difference is the distal t… | 2006 |
| * | K925626 | Maxima Plus Hollow Fiber OxygenatorSummary page 3…ly marketed devices to which substantial equivalence is claimed: 1. Spire Biomedical coated Decathlon (K060155) 2. Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K933586) 3. Diametrics Paratrend intravascular blood gas sensor catheter (K970906) 4. Cordis Bx Velocity coronary stent… | 1993 |
| * | K933586 | Minimax Plus Hollow Fiber Oxygenator (with or Without Medtronic/Carmeda Bioactive SurfaceSummary page 3…devices to which substantial equivalence is claimed: 1. Spire Biomedical coated Decathlon (K060155) 2. Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K933586) 3. Diametrics Paratrend intravascular blood gas sensor catheter (K970906) 4. Cordis Bx Velocity coronary stent (P900043/S0… | 1994 |
| * | K970906 | Paratrend 7 Multiparameter Senior and Satellite Monitor System with Paratrend 7 Plus…Summary page 3…oated Decathlon (K060155) 2. Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K933586) 3. Diametrics Paratrend intravascular blood gas sensor catheter (K970906) 4. Cordis Bx Velocity coronary stent (P900043/S024) Performance Data: The data from the coated Decathlon (K060155) were not… | 1997 |
| * | P900043 | Palmaz-Schatz (TM) Balloon Expandable StentSummary page 3…ardiopulmonary bypass circuit (K925626 and K933586) 3. Diametrics Paratrend intravascular blood gas sensor catheter (K970906) 4. Cordis Bx Velocity coronary stent (P900043/S024) Performance Data: The data from the coated Decathlon (K060155) were not duplicated for the Decathlon Gold with funnel tips… | 1994 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 23, 2009 | Not classified | Z-0288-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0289-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0290-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0299-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0301-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0302-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0304-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0306-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0309-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0311-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0312-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0294-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0295-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0296-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0297-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0300-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0303-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0307-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0308-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0314-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0316-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0317-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0318-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0319-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0321-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0330-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0291-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0292-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0293-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0298-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0305-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0310-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0313-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0315-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0320-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0322-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0323-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0324-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0325-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0327-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0328-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0332-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0334-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0335-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0329-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0331-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0333-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
| Oct 23, 2009 | Not classified | Z-0326-2010 | Process control | There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential… |
The 5 latest of 48 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 2 companies; a new generation was cleared every 13.3 years on median.
- 1 of the 6 ancestors has a recall of any class.
- 2 other devices also cite its predicate K060155.
How does its category compare?
All MSD devicesMSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · MSD
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MSD
153in 2021
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MSD
1in 2013
Class I (4 since 2007) Class II/III not classified
Adverse event reports per year · MSD
1,054in 2025
MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 92 | 0 | 118 |
| 2008 | 3 | 89 | 47 | 174 |
| 2009 | 6 | 81 | 49 | 173 |
| 2010 | 4 | 116 | 11 | 306 |
| 2011 | 7 | 70 | 24 | 268 |
| 2012 | 2 | — | 0 | 367 |
| 2013 | 5 | 105 | 1 | 363 |
| 2014 | 8 | 246 | 0 | 506 |
| 2015 | 1 | — | 0 | 550 |
| 2016 | 2 | — | 0 | 532 |
| 2017 | 2 | — | 0 | 484 |
| 2018 | 2 | — | 0 | 463 |
| 2019 | 1 | — | 0 | 645 |
| 2020 | 4 | 106 | 0 | 401 |
| 2021 | 4 | 153 | 0 | 523 |
| 2022 | 0 | — | 0 | 610 |
| 2023 | 0 | — | 0 | 542 |
| 2024 | 0 | — | 0 | 841 |
| 2025 | 1 | — | 0 | 1,054 |
| 2026 | 0 | — | 4 (4) | — |
What is it indicated for?
Spire Biomedical Inc's Decathlon Gold, Coated Twin Lumen Chronic Hemodialysis Catheter with separated funnel tips is designed for chronic hemodialysis and apheresis. It is a radiopaque polyurethane with a heparin coating, designed for percutaneous insertion or insertion via cutdown. The ability of the Carmeda® End point Bonded Heparin Coating to reduce clotting is supported by in-vitro testing. Catheters longer than 40cm are intended for femoral vein insertion.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K063431 · printed Nov 5, 2026
Cite this page
510(k) Report. “K063431: 15.5 FR Decathlon Gold, Funnel Tips, Coated Hemodialysis Catheter, DFC19SH24, DFC23SH28, DFC27SH32, DFC31SH36, DFC35SH40.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K063431