510(k) Report

510(k) · Cleared

Recalled · 2009

15.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24, ALC23SH28, ALC27SH32, ALC31SH36, ALC35SH40, ALC50SH55

K063441510(k) TraditionalCleared Feb 12, 200790-day reviewClass II21 CFR 876.5540Recalls 48

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spire Biomedical, Inc. · MSD Catheter, Hemodialysis, Implanted

Cleared on Feb 12, 2007 as substantially equivalent to K060155 (Spire Biomedical, Inc., 2006) and 5 other predicates after a 90-day review. Its family tree reaches back 3 generations to 2002. It has been recalled 48 times.

Decided
Feb 12, 2007
Review
90 days
Class
Class II
Product code
MSD
Recalls
48
Predicates
6

Its longest line of descent runs back 3 generations to K013160 (2002). 8 ancestors in all, the oldest K925626 (1993). No later device cites it as a predicate yet.

  1. Root · 3 generations backK013160 Pourchez Xpresso Twin Limen Chronic Hemodialysis CatheterSpire Biomedical, Inc. · 2002
  2. 2 generations backK032061 Decathlon Twin Lumen Hemodialysis with Separated Tips, Models…Spire Biomedical, Inc. · 2003 · Recalled
  3. 1 generation backK040509 Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24…Spire Biomedical, Inc. · 2004 · Recalled
  4. This deviceK063441 15.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24…Spire Biomedical, Inc. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K013160Pourchez Xpresso Twin Limen Chronic Hemodialysis CatheterSpire Biomedical, Inc.2002—
−2K032061Decathlon Twin Lumen Hemodialysis with Separated Tips, Models DE19/24SH…Spire Biomedical, Inc.200348
−1K040509Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24, AL23NH28…Spire Biomedical, Inc.200448
0 (this device)K06344115.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24, ALC23SH28…Spire Biomedical, Inc.2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199410
199710
200010
200610
FIG. 1 — Lineage of K063441, 2002–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K06015515.5FR Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter with Separated Tips
Summary page 1
…hich Substantial Equivalence is Claimed: * Spire Biomedical, Inc.'s 15.5Fr Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter with Separated Tips (K060155) * Spire Biomedical, Inc.'s 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis Catheter (K040509) *…
Spire Biomedical, Inc.200647
*K040509Alta LR Fixed Tip Chronic Hemodialysis Catheter, Models AL19NH24, AL23NH28, AL27NH32…
Summary page 1
…emodialysis Catheter with Separated Tips (K060155) * Spire Biomedical, Inc.'s 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis Catheter (K040509) * Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K 933586) * Diametrics Paratrend intravascular bloo…
Spire Biomedical, Inc.200448
*K925626Maxima Plus Hollow Fiber Oxygenator
Summary page 1
…Spire Biomedical, Inc.'s 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis Catheter (K040509) * Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K 933586) * Diametrics Paratrend intravascular blood gas sensor catheter (K970906) * Cordis Bx Velocity coronar…
Medtronic Vascular1993—
*K933586Minimax Plus Hollow Fiber Oxygenator (with or Without Medtronic/Carmeda Bioactive Surface
Summary page 1
…edical, Inc.'s 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis Catheter (K040509) * Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K 933586) * Diametrics Paratrend intravascular blood gas sensor catheter (K970906) * Cordis Bx Velocity coronary stent (P900…
Medtronic Vascular1994—
*K970906Paratrend 7 Multiparameter Senior and Satellite Monitor System with Paratrend 7 Plus…
Summary page 1
…Catheter (K040509) * Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K 933586) * Diametrics Paratrend intravascular blood gas sensor catheter (K970906) * Cordis Bx Velocity coronary stent (P90043/S024) 16 Snjr~~~ Spire Biomedical, Inc., a One…
Diametrics Medical, Ltd.1997—
*P900043/S024BX Velocity (TM) with Hepacoat (TM) on Raptor (TM) Stent System for Coronary Arteries
Summary page 3
…opulmonary bypass circuit (K925626 and K933586) 3. Diametrics Paratrend intravascular blood gas sensor catheter (K970906) 4. Cordis Bx Velocity coronary stent (P900043/S024) Performance Data: A series of physical, mechanical, and biological tests were conducted on the predicate, coated Decathlon cath…
Cordis Corp.2000—

Also cleared against K060155: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 23, 2009Not classifiedZ-0288-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0289-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0290-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0299-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0301-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0302-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0304-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0306-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0309-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0311-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0312-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0294-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0295-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0296-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0297-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0300-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0303-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0307-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0308-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0314-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0316-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0317-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0318-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0319-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0321-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0330-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0291-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0292-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0293-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0298-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0305-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0310-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0313-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0315-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0320-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0322-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0323-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0324-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0325-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0327-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0328-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0332-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0334-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0335-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0329-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0331-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0333-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…
Oct 23, 2009Not classifiedZ-0326-2010Process controlThere is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 1.7 years on median.
  • 3 of the 8 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K060155.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

Spire Biomedical Inc's Alta Gold, Coated Twin Lumen Chronic Hemodialysis Catheter with fixed tips is designed for chronic hemodialysis and apheresis. It is a radiopaque polyurethane with a heparin coating, designed for percutaneous insertion or insertion via cutdown. The ability of the Carmeda End point Bonded Heparin Coating to reduce clotting is supported by in-vitro testing. Catheters longer than 40cm are intended for femoral vein insertion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063441: 15.5FR Alta Gold, Coated Hemodialysis Catheter, Model ALC19SH24, ALC23SH28, ALC27SH32, ALC31SH36, ALC35SH40, ALC50SH55.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K063441

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026