510(k) Report

510(k) · Cleared

Recalled · 2008

Modification to: on-Board Imager Device

K041519510(k) SpecialCleared Aug 4, 200458-day reviewClass II21 CFR 892.5050Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Varian Medical Systems, Inc. · IYE Accelerator, Linear, Medical

Cleared on Aug 4, 2004 as substantially equivalent to K040192 (Varian Medical Systems, Inc., 2004) after a 58-day review. Its family tree reaches back 3 generations to 1996. It has been recalled 2 times.

Decided
Aug 4, 2004
Review
58 days
Class
Class II
Product code
IYE
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K963624 (1996). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK963624 Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc. · 1996 · Recalled
  2. 2 generations backK032996 Elekta Synergy SystemElekta Oncology Systems, Ltd. · 2003 · Recalled
  3. 1 generation backK040192 ON-Board Imager DeviceVarian Medical Systems, Inc. · 2004 · Recalled
  4. This deviceK041519 Modification to: on-Board Imager DeviceVarian Medical Systems, Inc. · 2004
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K963624Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc.19962
−2K032996Elekta Synergy SystemElekta Oncology Systems, Ltd.20033
−1K040192ON-Board Imager DeviceVarian Medical Systems, Inc.20042
0 (this device)K041519Modification to: on-Board Imager DeviceVarian Medical Systems, Inc.2004—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
FIG. 1 — Lineage of K041519, 1996–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040192ON-Board Imager Device
Summary page 1
…Product Code: 90 IYE 3. Predicate Devices to claim substantial equivalence: a. Varian Medical Systems'On Board Imager, K040192 4. Device Description: The present device (K040192) is cleared to make corrections to a patient's position based on a c…
Varian Medical Systems, Inc.20042

Also cleared against K040192: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 3, 2008Not classifiedZ-1442-2008Software designIf used with a third party radiation therapy treatment planning software system, mistreatment may occur because of a misalignment.
Apr 18, 2007Not classifiedZ-1195-2007Device DesignPatient positioning errors may occur because of mechanical variations during product installation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 5 months on median.
  • 3 of the 4 ancestors have a recall of any class.
  • 3 other devices also cite its predicate K040192.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The On-Board Imager Device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041519: Modification to: on-Board Imager Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041519

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026