510(k) Report

510(k) · Cleared

Recalled · 2010

G5 I HBA1C Test, G5 II HBA1C Test

K041635510(k) TraditionalCleared Aug 16, 200461-day reviewClass II21 CFR 864.7470Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Provalis Diagnostics, Ltd. · LCP Assay, Glycosylated Hemoglobin

Cleared on Aug 16, 2004 as substantially equivalent to K952720 (Aalto Scientific, Ltd., 1995) after a 61-day review. Its family tree reaches back 1 generation to 1995. 1 later device descends from it. It has been recalled 4 times.

Decided
Aug 16, 2004
Review
61 days
Class
Class II
Product code
LCP
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K952720 (1995). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2009: 1 descendant clearedK952720 · Aalto Scientific LTD. Glycated Hemoglobin Control Levels I & II · Aalto Scientific, Ltd. · 1995 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant K952720 · Aalto Scientific LTD. Glycated Hemoglobin Control Levels I & II · Aalto Scientific, Ltd. · 1995 · recalled0
  1. Root · 1 generation backK952720 Aalto Scientific LTD. Glycated Hemoglobin Control Levels I & IIAalto Scientific, Ltd. · 1995 · Recalled
  2. This deviceK041635 G5 I HBA1C Test, G5 II HBA1C TestProvalis Diagnostics, Ltd. · 2004
  3. Descendant · cites it directly · 0 of its ownK081269 Avie A1C Test SystemMec Dynamics Corp · 2009

Most influential descendants

Showing 1 of 1 · See all
  • K081269Avie A1C Test SystemMec Dynamics Corp · 2009 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K952720Aalto Scientific LTD. Glycated Hemoglobin Control Levels I & IIAalto Scientific, Ltd.19953
0 (this device)K041635G5 I HBA1C Test, G5 II HBA1C TestProvalis Diagnostics, Ltd.2004—
+1K081269Avie A1C Test SystemMec Dynamics Corp2009—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
FIG. 1 — Lineage of K041635, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCP device

  1. #1 K221326 Nova Allegro HbA1c Assay, Nova Allegro Analyzerscore 9.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K952720Aalto Scientific LTD. Glycated Hemoglobin Control Levels I & II
Our reading of the summary gave “K952720.” beside “510(k)”; FDA’s K952720 is the only number that fits it and carries the summary’s name for the device.
Summary page 8
…ple addition G5 HbAl, Quality Control Kit Provalis use commercially available controls from Aalto scientific Ltd; Glycohemoglobin controls normal and abnormal. 510(k) K952720. 7) Conclusion: These performance characteristics clearly indicate substantial equivalence of…
Aalto Scientific, Ltd.19953
*—Glycosal II HbA1c Test
Named without a submission number
—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081269Avie A1C Test SystemMec Dynamics Corp20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 7, 2010Not classifiedZ-0109-2011Component design/selectionTest Cartridges may leak and have the potential to generate error messages and biased inaccurate results.
Jul 26, 2010Not classifiedZ-0122-2011Device DesignInaccurate results with high bias-- Inaccurate results may be read by the machine, creating the risk for untreated hypoglycemia at elevated altitudes.
Aug 3, 2006Not classifiedZ-2460-2008Nonconforming Material/ComponentCholestech Corp. was notified by the foreign manufacturer of the A1C test cartridge that stability data indicated that the product would not meet performance claims through the end of its shelf life.
May 20, 2005Not classifiedZ-0171-06OtherProvalis Diagnostics discovered, following its internal stability testing program (mid-point batch testing), that occasional low values were being found across multiple lots of in2it HbA1C Test Cartridges.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K081269 (Mec Dynamics Corp, 2009) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LCP devices

LCP Assay, Glycosylated Hemoglobin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LCP

1in 2024

2007: 72008: 52009: 62010: 22011: 102012: 92013: 22014: 82015: 32016: 22017: 52018: 22019: 42020: 12023: 22024: 1
2007high 10 (2011) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCP

269in 2019

2007: 133 days median2008: 206 days median2009: 158 days median2011: 176 days median2012: 143 days median2014: 127 days median2015: 239 days median2017: 218 days median2019: 269 days median
2007high 269 (2019) · low 127 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LCP

1in 2025

2007: 12008: 32009: 62010: 52011: 512013: 42014: 22015: 22016: 62017: 12018: 12022: 12024: 12025: 1
2007high 51 (2011) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCP

22in 2025

2007: 52008: 112009: 152010: 42011: 272012: 512013: 1932014: 1822015: 642016: 292017: 1492018: 1,9112019: 1142020: 132021: 72022: 82023: 272024: 392025: 22
2007high 1,911 (2018) · low 4 (2010)2025

MAUDE reports naming product code LCP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007713315
20085206311
20096158615
20102—54
2011101765127
20129143051
20132—4193
201481272182
20153239264
20162—629
201752181149
20182—11,911
201942690114
20201—013
20210—07
20220—18
20232—027
20241—139
20250—122
20260—0—

What is it indicated for?

The G5 II HbAlc Test System is intended for testing blood taken from a fingerprick. G5 II HbAl c Test System shows how good glucose control has been over a two to three month period. The G5 II HbAlc Test System consists of the HbAlc test cartridge, the G5 Instrument, the G5 System Check Cartridge and the 05 HbAlc Quality Controls. The G5 II HbAlc Test System is intended for prescription home use. The G5 I HbAlc Test System is an affinity chromatography…

Read the full statement (178 words)

The G5 II HbAlc Test System is intended for testing blood taken from a fingerprick.

G5 II HbAl c Test System shows how good glucose control has been over a two to three month period.

The G5 II HbAlc Test System consists of the HbAlc test cartridge, the G5 Instrument, the G5 System Check Cartridge and the 05 HbAlc Quality Controls.

The G5 II HbAlc Test System is intended for prescription home use.

The G5 I HbAlc Test System is an affinity chromatography method and is intended for the in vitro quantitative determination of HbAl c in capillary blood taken from a fingerprick or whole blood in EDTA.

G5 I HbAl c Test System is indicated for monitoring the time averaged blood glucose levels of known diabetics, for professional use as an indicator of overall Glycaemic control.

The G5 I HbAlc Test System consists of the HbAlc test cartridge, the G5 Instrument, the G5 System Check Cartridge and the G5 HbAlc Quality Controls.

The 05 I HbAlc Test System is intended for use in a physicians/ doctors office.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041635: G5 I HBA1C Test, G5 II HBA1C Test.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K041635

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026