510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Exactech Equinoxe Shoulder System

K042021510(k) TraditionalCleared Oct 26, 200490-day reviewClass II21 CFR 888.3660Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Exactech, Inc. · KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Cleared on Oct 26, 2004 as substantially equivalent to K980244 (Tornier, 1999) and 17 other predicates after a 90-day review. Its family tree reaches back 3 generations to 1991. 71 later devices descend from it. It has been recalled 10 times.

Decided
Oct 26, 2004
Review
90 days
Class
Class II
Product code
KWS
Recalls
10
Predicates
18

Its longest line of descent runs back 3 generations to K911686 (1991). 27 ancestors in all, the oldest K790987 (1979). 71 later devices descend from it, 14 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2003: 1 other ancestor2001: 5 other ancestors2000: 5 other ancestors1999: 3 other ancestors1998: 2 other ancestors1996: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor1992: 1 other ancestor1987: 1 other ancestor1979: 1 other ancestor1993: 1 other ancestor2010: 6 descendants cleared2011: 4 descendants cleared2018: 7 descendants cleared2006: 2 descendants cleared2017: 2 descendants cleared2015: 4 descendants cleared2014: 2 descendants cleared2012: 3 descendants cleared2013: 1 descendant cleared2021: 3 descendants cleared2025: 8 descendants cleared2023: 7 descendants cleared2022: 4 descendants cleared2019: 3 descendants cleared2020: 1 descendant cleared2024: 6 descendants cleared2016: 3 descendants cleared2026: 5 descendants clearedK911686 · Global Total Shoulder, Uncement Humeral Hemi-Shoul · Depuy, Inc. · 1991×2: 2 generations cleared close together, K984541 (1999), K992065 (1999) 19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 16 other ancestors2 generations back: 7 other ancestors3 generations back: 1 other ancestor 1 generation forward: 14 descendants2 generations forward: 17 descendants3 generations forward: 18 descendants4 generations forward: 8 descendants5 generations forward: 3 descendants6 generations forward: 5 descendants7 generations forward: 4 descendants8 generations forward: 2 descendants K992065 · Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage Eccentric Head · DePuy Orthopaedics, Inc. · 1999K984541 · Global FX Humeral Stem, Global Advantage Humeral Head · DePuy Orthopaedics, Inc. · 1999K911686 · Global Total Shoulder, Uncement Humeral Hemi-Shoul · Depuy, Inc. · 19910+5
  1. Root · 3 generations backK911686 Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc. · 1991
  2. 2 generations backK984541 Global FX Humeral Stem, Global Advantage Humeral HeadDePuy Orthopaedics, Inc. · 1999
  3. 1 generation backK992065 Global Advantage Shoulder, Global Advantage Humeral Stem, Global…DePuy Orthopaedics, Inc. · 1999
  4. This deviceK042021 Exactech Equinoxe Shoulder SystemExactech, Inc. · 2004
  5. Descendant · cites it directly · 24 of its ownK093430 Exactech Equinoxe XL Keeled Glenoid, Pegged Glenoid, Exactech…Exactech, Inc. · 2010 · Recalled
  6. Descendant · cites it directly · 22 of its ownK103419 Exactech Equinoxe R Uhmwpe Posterior Augment Pegged Glenoids TMExactech, Inc. · 2010 · Recalled
  7. Descendant · 2 generations on · 21 of its ownK111379 Equinoxe Uhmwpe 12* Posterior Augment Pegged Glenoid, Small / Medium…Exactech, Inc. · 2011 · Recalled
  8. Descendant · 3 generations on · 20 of its ownK121220 Equinoxe Glenoid, Uhmwpe 16 PS=Osterior Augment Pegged, Small…Exactech, Inc. · 2012 · Recalled
  9. Descendant · 4 generations on · 16 of its ownK150583 Aequalis PerFORM+ Shoulder SystemTornier, Inc. · 2015

Showing 5 of 71 descendants. See all descendants

Descendants cleared per year

2006: 22010: 62011: 42012: 32013: 12014: 22015: 42016: 32017: 22018: 72019: 32020: 12021: 32022: 42023: 72024: 62025: 82026: 5
2006high 8 (2025) · low 0 (2007)2026

Most influential descendants

Showing 5 of 71 · See all
  • K093430Exactech Equinoxe XL Keeled Glenoid, Pegged Glenoid, Exactech Equinoxe Cage, Glenoid…Exactech, Inc. · 2010 · cites it directly24descendants
  • K103419Exactech Equinoxe R Uhmwpe Posterior Augment Pegged Glenoids TMExactech, Inc. · 2010 · cites it directly22descendants
  • K111379Equinoxe Uhmwpe 12* Posterior Augment Pegged Glenoid, Small / Medium / Large /…Exactech, Inc. · 2011 · 2 generations on21descendants
  • K121220Equinoxe Glenoid, Uhmwpe 16 PS=Osterior Augment Pegged, Small, Left/RightExactech, Inc. · 2012 · 3 generations on20descendants
  • K150583Aequalis PerFORM+ Shoulder SystemTornier, Inc. · 2015 · 4 generations on16descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K911686Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc.1991—
−2K984541Global FX Humeral Stem, Global Advantage Humeral HeadDePuy Orthopaedics, Inc.1999—
−1K992065Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage…DePuy Orthopaedics, Inc.1999—
0 (this device)K042021Exactech Equinoxe Shoulder SystemExactech, Inc.2004—
+1K093430Exactech Equinoxe XL Keeled Glenoid, Pegged Glenoid, Exactech Equinoxe Cage…Exactech, Inc.20101
+1K103419Exactech Equinoxe R Uhmwpe Posterior Augment Pegged Glenoids TMExactech, Inc.20102
+2K111379Equinoxe Uhmwpe 12* Posterior Augment Pegged Glenoid, Small / Medium / Large /…Exactech, Inc.20112
+3K121220Equinoxe Glenoid, Uhmwpe 16 PS=Osterior Augment Pegged, Small, Left/RightExactech, Inc.20122
+4K150583Aequalis PerFORM+ Shoulder SystemTornier, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198710
199210
199310
199410
199510
199610
199820
199930
200050
200150
200310
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200620
201060
201140
201230
201310
201420
201540
201630
201720
201870
201930
202010
202130
202240
202370
202460
202580
202650
FIG. 1 — Lineage of K042021, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWS device

  1. #1 K220847 AETOS Shoulder Systemscore 21.1 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (18)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K980244Aequalis Shoulder Prosthesis
Summary page 2
…Specifically, the Equinoxe® press-fit humeral stems are substantially equivalent to the Tomier Aequalis (K980244), the Zimmer (Sulzer) Anatomica (K003801), the Plus Orthopedic PROMOS (K032126), and the Arthrex…
Tornier1999—
*K003801Anatomical Press-Fit Humeral Stem
Summary page 2
…ally, the Equinoxe® press-fit humeral stems are substantially equivalent to the Tomier Aequalis (K980244), the Zimmer (Sulzer) Anatomica (K003801), the Plus Orthopedic PROMOS (K032126), and the Arthrex (K010124) press-fit…
Sulzer Orthopedics, Inc.20011
*K032126Promos Shoulder
Summary page 2
…nt to the Tomier Aequalis (K980244), the Zimmer (Sulzer) Anatomica (K003801), the Plus Orthopedic PROMOS (K032126), and the Arthrex (K010124) press-fit humeral stems. The Equinoxe® cemented humeral (short & long…
Plus Orthopedics2004—
*K010124Arthrex Univers 30 Shoulder Prosthesis
Summary page 2
…the Tomier Aequalis (K980244), the Zimmer (Sulzer) Anatomica (K003801), the Plus Orthopedic PROMOS (K032126), and the Arthrex (K010124) press-fit humeral stems. The Equinoxe® cemented humeral (short & long) stems are…
Arthrex, Inc.2001—
*K992065Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage Eccentric Head
Summary page 2
…ort & long) stems are substantially equivalent to the Plus Orthopedic PROMOS (K032126), the Depuy Global (K992065), and the Arthrex (K010124) press-fit humeral stems. The Equinoxe® fracture humeral stems are sub…
DePuy Orthopaedics, Inc.1999—
*K994392Aequalis Shoulder Fracture System
Summary page 2
…humeral stems. The Equinoxe® fracture humeral stems are substantially equivalent to the Tornier Fracture (K994392, K003728, K032679), and the Depuy Global FX (K1011099 & K984541) fracture humeral stems. The modu…
Tornier2000—
*K003728Modification to Aequalis Shoulder Fracture System
Summary page 2
…tems. The Equinoxe® fracture humeral stems are substantially equivalent to the Tornier Fracture (K994392, K003728, K032679), and the Depuy Global FX (K1011099 & K984541) fracture humeral stems. The modular eccen…
Tornier2000—
*K032679Aequalis Shoulder Fracture System
Summary page 2
…Equinoxe® fracture humeral stems are substantially equivalent to the Tornier Fracture (K994392, K003728, K032679), and the Depuy Global FX (K1011099 & K984541) fracture humeral stems. The modular eccentric hume…
Tornier2003—
*K011099Global FX Porous-Coated Humeral Stem
Our reading of the summary gave K101099, which is not a number in FDA’s lists. Corrected to the number with two digits swapped whose FDA name carries the summary’s words for the device, “Depuy Global FX fracture humeral stems”.
DePuy Orthopaedics, Inc.2001—
*K984541Global FX Humeral Stem, Global Advantage Humeral Head
Summary page 2
…ially equivalent to the Tornier Fracture (K994392, K003728, K032679), and the Depuy Global FX (K1011099 & K984541) fracture humeral stems. The modular eccentric humeral heads are substantially equivalent to the…
DePuy Orthopaedics, Inc.1999—
*K974044Global Total Shoulder Eccentric Humeral Head
Summary page 2
…K984541) fracture humeral stems. The modular eccentric humeral heads are substantially equivalent to the Depuy Global (K974044 & K992065), the Stryker Howmedica Osteonics Solar (K001419), the Tomier Aequalis (K012212), and t…
DePuy Orthopaedics, Inc.1998—
*K001419Solar Shoulder Offset Humeral Head
Summary page 2
…substantially equivalent to the Depuy Global (K974044 & K992065), the Stryker Howmedica Osteonics Solar (K001419), the Tomier Aequalis (K012212), and the Zimmer (Sulzer) Anatomica (K990137) eccentric humeral he…
Howmedica Osteonics Corp.2000—
*K012212Modification to Aequalis Shoulder System
Summary page 2
…substantially equivalent to the Depuy Global (K974044 & K992065), the Stryker Howmedica Osteonics Solar (K001419), the Tomier Aequalis (K012212), and the Zimmer (Sulzer) Anatomica (K990137) eccentric humeral heads. Finally, the…
Tornier2001—
*K990137Anatomica Humeral Stem/Head Components
Summary page 2
…er Howmedica Osteonics Solar (K001419), the Tomier Aequalis (K012212), and the Zimmer (Sulzer) Anatomica (K990137) eccentric humeral heads. Finally, the Equinoxe® cemented pegged and keeled glenoids are substant…
Sulzer Orthopedics, Inc.1999—
*K994393Aequalis Universal Shoulder Glenoid
Summary page 2
…y, the Equinoxe® cemented pegged and keeled glenoids are substantially equivalent to the Torier Aequalis (K994393), the Encore Foundation (K960906), the Plus Orthopedic PROMOS (K032126), the Zimmer (Sulzer) Anat…
Tornier2000—
*K960906Foundation Glenoid Component
Summary page 2
…Equinoxe® cemented pegged and keeled glenoids are substantially equivalent to the Torier Aequalis (K994393), the Encore Foundation (K960906), the Plus Orthopedic PROMOS (K032126), the Zimmer (Sulzer) Anatomica (K990136),…
Encore Orthopedics, Inc.1996—
*K990136Anatomica Glenoid Component
Summary page 2
…the Torier Aequalis (K994393), the Encore Foundation (K960906), the Plus Orthopedic PROMOS (K032126), the Zimmer (Sulzer) Anatomica (K990136), and the Depuy Global (K981487) cemented glenoids.…
Sulzer Orthopedics, Inc.19992
*K981487Depuy Global Shoulder Glenoid
Summary page 2
…06), the Plus Orthopedic PROMOS (K032126), the Zimmer (Sulzer) Anatomica (K990136), and the Depuy Global (K981487) cemented glenoids.…
DePuy Orthopaedics, Inc.1998—

Also cleared against K980244: 2 other devices →

What descends from it? (71)

See all 71

14 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093430Exactech Equinoxe XL Keeled Glenoid, Pegged Glenoid, Exactech Equinoxe Cage, Glenoid…Exactech, Inc.201024
K103419Exactech Equinoxe R Uhmwpe Posterior Augment Pegged Glenoids TMExactech, Inc.201022
K061454Exactech Equinoxe Shoulder StemsExactech, Inc.200616
K093599Arrow Anatomical Shoulder SystemFournitures Hospitalieres Industrie201011
K140063Equinoxe Extra Short Humeral HeadExactech, Inc.201410
K173388Exactech Equinoxe Stemless ShoulderExactech, Inc.20189
K060535Spacer-S Temporary Shoulder SpacerTecres S.P.A.20068
K102951Exactech Equinox Fixed Angle Replicator Plate (0MM,1.5MM,4.5MM), Humeral Head-Short…Exactech, Inc.20107
K150568ARROW Humeral stems size 6 and 16Fournitures Hospitalieres Industrie20154
K162726Exactech® Equinoxe® Preserve StemExactech, Inc.20173

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 5, 2025Class IIZ-0535-2026Release of Material/Component prior to receiving test resultsDue to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Sep 5, 2025Class IIZ-0537-2026Release of Material/Component prior to receiving test resultsDue to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Sep 5, 2025Class IIZ-0536-2026Release of Material/Component prior to receiving test resultsDue to products that have not completed design updates and performance testing being released/distributed. Products may not conform to design and performance criteria.
Mar 6, 2024Class IIZ-1402-2024Nonconforming Material/ComponentThe packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer…
Mar 6, 2024Class IIZ-1403-2024Nonconforming Material/ComponentThe packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer…
Mar 6, 2024Class IIZ-1405-2024Nonconforming Material/ComponentThe packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer…
Mar 6, 2024Class IIZ-1404-2024Nonconforming Material/ComponentThe packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer…
Nov 9, 2020Class IIZ-0884-2021Employee errorPotential for incorrect Instructions For Use (IFU) included with an Equinoxe Humeral Stem, Primary, Press Fit, 11mm package.
Feb 13, 2015Class IIZ-1459-2015Labeling mix-upsIncorrectly packaged. Outer and inner labeling may incorrectly identify the size of the enclosed device.
Jun 15, 2012Class IIZ-2115-2012Labeling False and MisleadingExactech, Inc. initiated a recall of Equinoxe Replicator Plates, Equinoxe Locking Screws, Equinoxe Compression Screw Kits, Equinoxe Fracture Adapter Tray Screws, and Octane-C Intervertebral fusion devices due to the firm determining that…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 5.3 years on median.
  • 2 of the 27 ancestors have a recall of any class.
  • Of its descendants, K093430 (Exactech, Inc., 2010) has gone furthest, with 24 descendants of its own.
  • 2 other devices also cite its predicate K980244.

How does its category compare?

All KWS devices

KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KWS

3in 2025

2007: 22008: 42009: 42010: 72011: 112012: 72013: 122014: 72015: 82016: 102017: 72018: 42019: 92020: 32021: 52022: 52023: 42024: 22025: 32026 (partial year): 2
2007high 12 (2013) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWS

97in 2025

2008: 37 days median2009: 108 days median2010: 33 days median2011: 90 days median2012: 86 days median2013: 120 days median2014: 89 days median2015: 129 days median2016: 114 days median2017: 120 days median2018: 152 days median2019: 106 days median2020: 98 days median2021: 151 days median2022: 158 days median2023: 250 days median2025: 97 days median
2007high 250 (2023) · low 33 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWS

2in 2023

2007: 212008: 292010: 102011: 112012: 102014: 12016: 12017: 42018: 12019: 22020: 52023: 22026 (partial year): 1
2007high 29 (2008) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWS

2,281in 2025

2007: 1582008: 2382009: 2942010: 3932011: 4862012: 5422013: 8182014: 1,0662015: 1,1702016: 9702017: 1,6812018: 1,5452019: 1,7282020: 1,5672021: 1,3582022: 1,2682023: 1,7502024: 2,1222025: 2,281
2007high 2,281 (2025) · low 158 (2007)2025

MAUDE reports naming product code KWS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—21158
200843729238
200941080294
201073310393
2011119011486
201278610542
2013121200818
201478911,066
2015812901,170
2016101141970
2017712041,681
2018415211,545
2019910621,728
202039851,567
2021515101,358
2022515801,268
2023425021,750
20242—02,122
202539702,281
20262—1—

What is it indicated for?

The Equinoxe Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment. Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows: rheumatoid arthritis, osteoarthritis, osteonecrosis or…

Read the full statement (156 words)

The Equinoxe Shoulder System is indicated to relieve pain and restore function in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment.

Clinical indications for the PRIMARY (P), LONG/REVISION (L/R) and FRACTURE (F) humeral components are as follows:

rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems

congenital abnormalities in the skeletally mature

primary and secondary necrosis of the humeral head.

humeral head fracture with displacement of the tuberosities

pathologies where arthodesis or resectional arthroplasty of the humeral head are not acceptable

revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved)

displaced three-part and four-part upper humeral fractures

spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases)

revision of failed previous reconstructions when distal anchorage is required to restore mobility from previous procedures (e.g. previous fusion)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042021: Exactech Equinoxe Shoulder System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042021

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026