510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Innova 4100, Innova 3100

K042053510(k) TraditionalCleared Aug 26, 200427-day reviewClass II21 CFR 892.1650Recalls 34

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Aug 26, 2004 as substantially equivalent to K960575 (GE Medical Systems, 1996) after a 27-day review. Its family tree reaches back 1 generation to 1996. It has been recalled 34 times.

Decided
Aug 26, 2004
Review
27 days
Class
Class II
Product code
OWB
Recalls
34
Predicates
1

Its longest line of descent runs back 1 generation to K960575 (1996). No later device cites it as a predicate yet.

K960575 · Advantx LCV+ · GE Medical Systems · 1996 · recalled199619982000200220042006
K960575 · Advantx LCV+ · GE Medical Systems · 1996 · recalled0
  1. Root · 1 generation backK960575 Advantx LCV+GE Medical Systems · 1996 · Recalled
  2. This deviceK042053 Innova 4100, Innova 3100Ge Medical Systems, Inc. · 2004
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K960575Advantx LCV+GE Medical Systems19966
0 (this device)K042053Innova 4100, Innova 3100Ge Medical Systems, Inc.2004—
FIG. 1 — Lineage of K042053, 1996–2004Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K960575Advantx LCV+
Our OCR of the scanned summary read K860575, FDA’s number for Yeast Identification Panel (American Micro Scan). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “GE Advantx LCV+”.
GE Medical Systems19966

Also cleared against K960575: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 18, 2015Class IIZ-0763-2016Nonconforming Material/ComponentGE Healthcare has recently become aware of a reported incident in which a Mavig suspension arm with monitor assembly fell to the floor.
Dec 11, 2013Class IIZ-1163-2014Software designGE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional…
Jan 21, 2013Class IIZ-0947-2013Nonconforming Material/ComponentA locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems.
Dec 20, 2011Class IIZ-1318-2013Nonconforming Material/ComponentWhile performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries…
Oct 21, 2011Class IIZ-1142-2013Labeling designAs part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff…
Aug 24, 2011Not classifiedZ-0133-2012Nonconforming Material/ComponentA potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails.
May 16, 2011Class IIZ-1616-2013Software designGE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
Jun 3, 2010Not classifiedZ-1922-2010Device DesignGE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.
Jun 3, 2010Not classifiedZ-1923-2010Device DesignGE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.
Jun 3, 2010Not classifiedZ-1921-2010Device DesignGE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety.
Jul 31, 2009Not classifiedZ-2241-2009Component design/selectionGE Healthcare has recently became aware of a potential power supply failure with the Innova 2000, 3100 and 4100 systems that may impact patient safety. A potential power supply failure inside the GE equipment cabinet could occur with no…
Jul 31, 2009Not classifiedZ-2242-2009Component design/selectionGE Healthcare has recently became aware of a potential power supply failure with the Innova 2000, 3100 and 4100 systems that may impact patient safety. A potential power supply failure inside the GE equipment cabinet could occur with no…
Apr 16, 2009Not classifiedZ-0135-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0137-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Jul 25, 2008Not classifiedZ-2438-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jul 25, 2008Not classifiedZ-2437-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jun 24, 2008Not classifiedZ-2358-2008Employee errorGE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two…
Jun 24, 2008Not classifiedZ-2356-2008Employee errorGE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two…
Jun 17, 2008Not classifiedZ-1475-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1476-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
May 19, 2008Not classifiedZ-1521-2008Software designGE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).
Mar 6, 2008Not classifiedZ-1093-2009Software designGE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to…
Mar 6, 2008Not classifiedZ-1092-2009Software designGE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to…
Jan 30, 2008Not classifiedZ-0475-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0474-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Nov 15, 2007Not classifiedZ-0691-2008Device DesignFrozen Image Problems: During an acquisition (fluoro and/or record), there have been cases reported where an image became "frozen" on the DL (digital leader acquisition system) live monitor screen. In such cases, the system continued to…
Nov 15, 2007Not classifiedZ-0692-2008Device DesignFrozen Image Problems: During an acquisition (fluoro and/or record), there have been cases reported where an image became "frozen" on the DL (digital leader acquisition system) live monitor screen. In such cases, the system continued to…
Oct 21, 2007Not classifiedZ-1516-2008Nonconforming Material/ComponentIncorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option)…
Oct 21, 2007Not classifiedZ-1517-2008Nonconforming Material/ComponentIncorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option)…
Mar 26, 2007Not classifiedZ-0871-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-0870-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-0869-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-1044-2007OtherCaution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense…
Mar 26, 2007Not classifiedZ-1043-2007OtherCaution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 1 other device also cites its predicate K960575.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The two Digital Fluoroscopic Imaging Systems are indicated for use in diagnostic and interventional angiographic procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventionnal procedures. These devices are not intended for mammography applications. The Bolus Chasing option submitted here is intended to perform subtracted angiography in real time of the lower limbs for diagnostic purpose.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K042053: Innova 4100, Innova 3100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042053

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026