K042292510(k) TraditionalCleared Sep 8, 200415-day reviewClass II21 CFR 892.2050Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stentor, Inc. · LLZ System, Image Processing, Radiological

Cleared on Sep 8, 2004 as substantially equivalent to K023557 (GE Medical Systems, 2002) and 1 other predicate after a 15-day review. Its family tree reaches back 2 generations to 1992. 15 later devices descend from it. It has been recalled 7 times.

Decided
Sep 8, 2004
Review
15 days
Class
Class II
Product code
LLZ
Recalls
7
Predicates
2

Its longest line of descent runs back 2 generations to K920319 (1992). 4 ancestors in all. 15 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1998: 1 other ancestor2005: 2 descendants cleared2006: 2 descendants cleared2013: 1 descendant cleared2009: 1 descendant cleared2011: 2 descendants cleared2012: 1 descendant cleared2018: 3 descendants cleared2016: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedK920319 · Sienet Image Display Workstations, Image Acquisiti · Siemens Gammasonics, Inc. · 1992K023557 · Centricity Pacs Plus · GE Medical Systems · 20021990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 5 descendants3 generations forward: 3 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant K023557 · Centricity Pacs Plus · GE Medical Systems · 2002K920319 · Sienet Image Display Workstations, Image Acquisiti · Siemens Gammasonics, Inc. · 19920+5
  1. Root · 2 generations backK920319 Sienet Image Display Workstations, Image AcquisitiSiemens Gammasonics, Inc. · 1992
  2. 1 generation backK023557 Centricity Pacs PlusGE Medical Systems · 2002
  3. This deviceK042292 Isite PacsStentor, Inc. · 2004
  4. Descendant · cites it directly · 8 of its ownK063267 Isite Pacs, V 4. XStentor Inc., A Philips Medical Systems Co. · 2006 · Recalled
  5. Descendant · 2 generations on · 5 of its ownK111804 Intellispace Pacs 4.XPhilips Healthcare Informatics, Inc. · 2011 · Recalled
  6. Descendant · 3 generations on · 3 of its ownK160315 I4 (Integrated Intelligent Imaging Informatics) systemPhilips Medical Systems Nederland B.V. · 2016
  7. Descendant · 4 generations on · 2 of its ownK173588 Illumeo SystemPhilips Medical Systems Technologies, Ltd. · 2018
  8. Descendant · cites it directly · 2 of its ownK051809 Easyviz, Easyviz System, Easyviz Pacs System, Easyviz Image and Data…Medical Insight A/S · 2005

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2005: 22006: 22009: 12011: 22012: 12013: 12016: 12018: 32025: 12026: 1
2005high 3 (2018) · low 0 (2007)2026

Most influential descendants

Showing 5 of 15 · See all
  • K063267Isite Pacs, V 4. XStentor Inc., A Philips Medical Systems Co. · 2006 · cites it directly8descendants
  • K111804Intellispace Pacs 4.XPhilips Healthcare Informatics, Inc. · 2011 · 2 generations on5descendants
  • K160315I4 (Integrated Intelligent Imaging Informatics) systemPhilips Medical Systems Nederland B.V. · 2016 · 3 generations on3descendants
  • K173588Illumeo SystemPhilips Medical Systems Technologies, Ltd. · 2018 · 4 generations on2descendants
  • K051809Easyviz, Easyviz System, Easyviz Pacs System, Easyviz Image and Data Management SystemMedical Insight A/S · 2005 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K920319Sienet Image Display Workstations, Image AcquisitiSiemens Gammasonics, Inc.1992—
−1K023557Centricity Pacs PlusGE Medical Systems2002—
0 (this device)K042292Isite PacsStentor, Inc.2004—
+1K063267Isite Pacs, V 4. XStentor Inc., A Philips Medical Systems Co.20064
+2K111804Intellispace Pacs 4.XPhilips Healthcare Informatics, Inc.20112
+3K160315I4 (Integrated Intelligent Imaging Informatics) systemPhilips Medical Systems Nederland B.V.2016—
+4K173588Illumeo SystemPhilips Medical Systems Technologies, Ltd.2018—
+1K051809Easyviz, Easyviz System, Easyviz Pacs System, Easyviz Image and Data…Medical Insight A/S2005—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200520
200620
200910
201120
201210
201310
201610
201830
202510
202610
FIG. 1 — Lineage of K042292, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K023557Centricity Pacs Plus
Summary page 1
…ode: LLZ 1.5 Substantial Equivalence Manufacturer: General Electric Medical Systems Model: Centricity PACS Plus 510(k) Number: K023557 Manufacturer: Eastman Kodak Company Model: Kodak DirectView PACS 510(k) Number: K030781 1.6 Device…
GE Medical Systems2002—
*K030781Kodak Directview Pacs/Image Management System, Kodak Directview Workflow Manager,Kodak…
Summary page 1
…Model: Centricity PACS Plus 510(k) Number: K023557 Manufacturer: Eastman Kodak Company Model: Kodak DirectView PACS 510(k) Number: K030781 1.6 Device Description and Intended Use iSite is an image management system intended to be used by trained professi…
Eastman Kodak Company2003—

Also cleared against K023557: 5 other devices →

What descends from it? (15)

See all 15

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063267Isite Pacs, V 4. XStentor Inc., A Philips Medical Systems Co.20068
K051809Easyviz, Easyviz System, Easyviz Pacs System, Easyviz Image and Data Management SystemMedical Insight A/S20052
K131424Miplatform Medical Imaging SuiteHinacom Software and Technology, Ltd.20131
K052618Imagecast PacsIdx Systems Corp.20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 23, 2011Not classifiedZ-1881-2011Software designA software anomaly that occurs when clinicians use Volume Vision to view images in 3D and uses the measurement tool to measure regions of interest. This could lead to an incorrect interpretation of the measurements which could cause an…
Aug 29, 2008Not classifiedZ-0118-2009Software designTwo defects have been identified: 1) potential to display a patient on the canvas page that is different from the patient whose images are displayed on diagnostic monitors when using Conference Presentation States and when opening two or…
Jul 3, 2008Not classifiedZ-2346-2008Software designScout line and localizer crosshair on MPR images will display in the incorrect position under certain circumstances.
Jan 29, 2008Not classifiedZ-1443-2008Software designWhen using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool.
Jan 4, 2008Not classifiedZ-1141-2008Software designIncorrect display of one of the patient's images. This malfunction could cause an image to be incorrectly displayed, for example flipped left to right and anterior to posterior. The original image is still presented in the correct…
Feb 14, 2007Not classifiedZ-0616-2007OtherA defect may cause patient image orientation markers to be incorrectly labeled on Multi-Planar Reformation (MPR) images. Left and Right Anatomical orientation markers may be reversed.
Dec 20, 2005Not classifiedZ-0399-06OtherWhen using a third party dictation accessory with the iSite PACS, a malfunction may occur that would result in the petient/exam selected and shown may be different on the radiologist console and diagnostic display screens

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 10.0 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K063267 (Stentor Inc., A Philips Medical Systems Co., 2006) has gone furthest, with 8 descendants of its own.
  • 5 other devices also cite its predicate K023557.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

iSite PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging device or modalities. Lossy compressed mammographic images and digitized film screen…

Read the full statement (103 words)

iSite PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging device or modalities.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042292: Isite Pacs.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042292

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026