510(k) · Cleared
Recalled · Class I · 2020Medley Pca Module, Model 8120
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Alaris Medical Systems, Inc. · MEA Pump, Infusion, Pca
Cleared on Dec 10, 2004 as substantially equivalent to K032233 (Alaris Medical Systems, Inc., 2003) after a 10-day review. Its family tree reaches back 1 generation to 2003. 1 of the 2 devices in its family, counting itself, has a Class I recall. It has been recalled 3 times, including a Class I recall.
- Decided
- Dec 10, 2004
- Review
- 10 days
- Class
- Class II
- Product code
- MEA
- Recalls
- 3 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 1 generation to K032233 (2003). No later device cites it as a predicate yet. 1 of the 2 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backK032233 Medley Pca Module, Model 8120Alaris Medical Systems, Inc. · 2003 · Recalled
- This deviceK043299 Medley Pca Module, Model 8120Alaris Medical Systems, Inc. · 2004
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K032233 | Medley Pca Module, Model 8120 | Alaris Medical Systems, Inc. | 2003 | 3 |
| 0 (this device) | K043299 | Medley Pca Module, Model 8120 | Alaris Medical Systems, Inc. | 2004 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K032233 | Medley Pca Module, Model 8120Summary page 5…toring Protocol is of comparable type and is substantially equivalent to the following predicate device: Medley PCA Module ALARIS Medical K032233 Sept. 9, 2003 Systems, Inc. INTENDED USE The Medley PCA Module with PCA/Monitoring… | 2003 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 4, 2020 | Class I | Z-2882-2020 | Component design/selection | The Alaris PC unit can display incorrect syringe type and/or syringe sizes. This could result in delays in infusion, under-infusion, or over-infusion. |
| Mar 5, 2008 | Not classified | Z-1712-2008 | Device Design | A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the display board of the… |
| Mar 5, 2008 | Not classified | Z-1711-2008 | Device Design | A channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors on all Alaris System… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Its one ancestor has a recall.
- 2 other devices also cite its predicate K032233.
How does its category compare?
All MEA devicesMEA Pump, Infusion, Pca: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MEA
2in 2017
2026 to Oct 25, outlined.
Median 510(k) review, days · MEA
Too few MEA 510(k)s in any year since 2007 to give a median.
Recalls per year · MEA
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MEA
3,278in 2025
MAUDE reports naming product code MEA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 81 |
| 2008 | 1 | — | 1 | 237 |
| 2009 | 1 | — | 0 | 153 |
| 2010 | 0 | — | 4 | 219 |
| 2011 | 0 | — | 2 | 245 |
| 2012 | 0 | — | 0 | 446 |
| 2013 | 1 | — | 0 | 453 |
| 2014 | 0 | — | 1 | 382 |
| 2015 | 0 | — | 0 | 99 |
| 2016 | 1 | — | 0 | 57 |
| 2017 | 2 | — | 2 | 44 |
| 2018 | 0 | — | 0 | 256 |
| 2019 | 0 | — | 0 | 132 |
| 2020 | 0 | — | 0 | 2,331 |
| 2021 | 0 | — | 0 | 3,776 |
| 2022 | 0 | — | 1 | 6,443 |
| 2023 | 0 | — | 0 | 3,984 |
| 2024 | 0 | — | 1 | 4,559 |
| 2025 | 0 | — | 0 | 3,278 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Medley PCA Module is intended for use in today's growing professional healthcare environment for facilities that utilize infusion devices for the delivery of medications or fluids. The Medley PCA Module is indicated for use on adults, pediatrics, and neonates for continuous or intermittent delivery through clinically acceptable routes of administration; such as intravenous (IV), subcutaneous, or epidural. The addition of the PCA/Monitoring Protocol…
Read the full statement (115 words)
The Medley PCA Module is intended for use in today's growing professional healthcare environment for facilities that utilize infusion devices for the delivery of medications or fluids. The Medley PCA Module is indicated for use on adults, pediatrics, and neonates for continuous or intermittent delivery through clinically acceptable routes of administration; such as intravenous (IV), subcutaneous, or epidural.
The addition of the PCA/Monitoring Protocol provides an optional and hospital-configurable feature that is intended to align with healthcare facilities' current protocols that require monitoring of patients while on PCA therapy. All device programming, data entry and validation of PCA/Monitoring Protocol parameters is performed by the trained healthcare professional according to hospital-defined protocol or a physician's order.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K043299 · printed Nov 6, 2026
Cite this page
510(k) Report. “K043299: Medley Pca Module, Model 8120.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K043299
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026