Infusomat Space Volumetric Infusion Pump System

K083689510(k) TraditionalCleared Jul 1, 2009201-day reviewClass II21 CFR 880.5725Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

B.Braun Medical, Inc. · FRN Pump, Infusion

Cleared on Jul 1, 2009 as substantially equivalent to K062700 (B.Braun Medical, Inc., 2006) and 2 other predicates after a 201-day review. Its family tree reaches back 4 generations to 1990. 7 of the 26 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
Jul 1, 2009
Review
201 days
Class
Class II
Product code
FRN
Recalls
3 (1 Class I)
Predicates
3

Its longest line of descent runs back 4 generations to K904518 (1990). 21 ancestors in all, the oldest K820133 (1982). 4 later devices descend from it, 2 citing it directly. 7 of the 26 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 2 other ancestors2001: 1 other ancestor, 1 with a Class I recall11997: 1 other ancestor1995: 1 other ancestor, 1 with a Class I recall11994: 3 other ancestors1993: 1 other ancestor1992: 2 other ancestors1991: 1 other ancestor1989: 1 other ancestor1986: 1 other ancestor1984: 2 other ancestors1982: 1 other ancestor2015: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall12022: 2 descendants clearedK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalledK994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK023189 · Vista Basic with FM System · B.Braun Medical, Inc. · 2003K062700 · Infusomat Space Volumetric Infusion Pump System · B.Braun Medical, Inc. · 2006 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 3 other ancestors, 1 with a Class I recall13 generations back: 12 other ancestors 1 generation forward: 2 descendants, 2 with a Class I recall22 generations forward: 2 descendants K062700 · Infusomat Space Volumetric Infusion Pump System · B.Braun Medical, Inc. · 2006 · Class I recallK023189 · Vista Basic with FM System · B.Braun Medical, Inc. · 2003K994375 · Horizon Light · B.Braun Medical, Inc. · 2000 · Class I recallK904518 · Volumetric Infusion Pump · Kendall Mcgaw Laboratories, Inc. · 1990 · recalled0
  1. Root · 4 generations backK904518 Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc. · 1990 · Recalled
  2. 3 generations backK994375 Horizon LightB.Braun Medical, Inc. · 2000 · Recalled (Class I)
  3. 2 generations backK023189 Vista Basic with FM SystemB.Braun Medical, Inc. · 2003
  4. 1 generation backK062700 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Recalled (Class I)
  5. This deviceK083689 Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2009
  6. Descendant · cites it directly · 2 of its ownK121613 Agilia Infusion SystemFresenius Kabi AG · 2013 · Recalled (Class I)
  7. Descendant · 2 generations on · 0 of its ownK210075 Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG · 2022
  8. Descendant · 2 generations on · 0 of its ownK210073 Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat…Fresenius Kabi AG · 2022
  9. Descendant · cites it directly · 0 of its ownK142596 Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K121613Agilia Infusion SystemFresenius Kabi AG · 2013 · cites it directly2descendants
  • K210075Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG · 2022 · 2 generations on0descendants
  • K210073Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration…Fresenius Kabi AG · 2022 · 2 generations on0descendants
  • K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K904518Volumetric Infusion PumpKendall Mcgaw Laboratories, Inc.199037
−3K994375Horizon LightB.Braun Medical, Inc.2000Class I
−2K023189Vista Basic with FM SystemB.Braun Medical, Inc.2003—
−1K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2006Class I
0 (this device)K083689Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc.2009—
+1K121613Agilia Infusion SystemFresenius Kabi AG2013Class I
+2K210075Vigilant Software Suite – Vigilant Master MedFresenius Kabi AG2022—
+2K210073Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines…Fresenius Kabi AG2022—
+1K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc.2015Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198420
198610
198910
199110
199220
199310
199430
199511
199710
200111
200320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201311
201511
202220
FIG. 1 — Lineage of K083689, 1990–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062700Infusomat Space Volumetric Infusion Pump System
Summary page 1
…original submission of the Infusomat® Space Volumetric Infusion Pump System marketed under cleared 510(k) K062700. The predicate device for barcode and wireless data transmission functionalities is B. Braun Medi…
B.Braun Medical, Inc.2006Class I
*K011975Horizon Outlook with Dosecom
Summary page 1
…n functionalities is B. Braun Medical Inc. Outlook® Safety Infusion System, marketed under cleared 510(k) K011975. The predicate device for PCA functionality is the MEDLEY TM PCA Module marketed by…
B.Braun Medical, Inc.2001Class I
*K032233Medley Pca Module, Model 8120
Summary page 1
…functionality is the MEDLEY TM PCA Module marketed by ALARIS Medical Systems Inc. under cleared 510(k) K032233. B. Braun Medical Inc. 510(k) Premarket Notification DESCRIPTION: The Infusomat…
Alaris Medical Systems, Inc.20033

Also cleared against K062700: 4 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121613Agilia Infusion System Class IFresenius Kabi AG20132
K142596Infusomat Space Volumetric Infusion Pump Class IB.Braun Medical, Inc.20150

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 23, 2012Class IIZ-0623-2013Device DesignFirm has become aware of the potential for breakage of the pole clamp locking grids, or rasters, in the Space Pole Clamps.
Mar 23, 2012Class IZ-1992-2012Device DesignFirm became aware of the potential for breakage of the anti free flow clip catch, located on the inside of the pump door, when the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed. If the…
Jul 12, 2011Not classifiedZ-1015-2012Device DesignIf the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed, the clip catch on the inside of the pump door may break. If the clip catch is broken and the door opened, free flow protection is…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062700Infusomat Space Volumetric Infusion Pump SystemB.Braun Medical, Inc. · 2006 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K011975Horizon Outlook with DosecomB.Braun Medical, Inc. · 2001 · Direct predicate · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall
  • K994375Horizon LightB.Braun Medical, Inc. · 2000 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K950419The Imed Orion Infusion Pump and Administration SetsImed Corp. · 1995 · 2 generations back · ◆ in the 1995 ancestors bar of FIG. 14 Class I recalls
  • K142596Infusomat Space Volumetric Infusion PumpB.Braun Medical, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K121613Agilia Infusion SystemFresenius Kabi AG · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 19 years and passes through 2 companies; a new generation was cleared every 3.6 years on median.
  • 12 of the 21 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K121613 (Fresenius Kabi AG, 2013) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K062700.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Infusomat® Space Volumetric Infusion Pump System is an electrical, external, volumetric infusion pump system indicated for use with adults, pediatrics and neonates and is intended to provide infusions of parenteral fluids/medications, blood and blood products indicated for infusion through FDA approved routes of administration.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K083689: Infusomat Space Volumetric Infusion Pump System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K083689

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026