510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Smith & Nephew, Inc. Revision Knee System

K043440510(k) AbbreviatedCleared Feb 18, 200566-day reviewClass II21 CFR 888.3560Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew, Inc. · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on Feb 18, 2005 as substantially equivalent to K041106 (Smith & Nephew, Inc., 2004) and 3 other predicates after a 66-day review. Its family tree reaches back 2 generations to 1996. 31 later devices descend from it. It has been recalled 7 times.

Decided
Feb 18, 2005
Review
66 days
Class
Class II
Product code
JWH
Recalls
7
Predicates
4

Its longest line of descent runs back 2 generations to K962137 (1996). 4 ancestors in all, the oldest K951987 (1995). 31 later devices descend from it, 11 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 1 other ancestor2005: 1 descendant cleared2015: 3 descendants cleared2018: 1 descendant cleared2023: 2 descendants cleared2021: 2 descendants cleared2012: 3 descendants cleared2011: 2 descendants cleared2006: 3 descendants cleared2010: 1 descendant cleared2019: 2 descendants cleared2017: 1 descendant cleared2013: 1 descendant cleared2014: 1 descendant cleared2022: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant cleared2020: 1 descendant cleared2025: 3 descendants clearedK962137 · Genesis II Constrained System · Smith & Nephew Richards, Inc. · 1996 · recalledK041106 · Revision Knee System · Smith & Nephew, Inc. · 2004 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 11 descendants2 generations forward: 12 descendants3 generations forward: 6 descendants4 generations forward: 2 descendants K041106 · Revision Knee System · Smith & Nephew, Inc. · 2004 · recalledK962137 · Genesis II Constrained System · Smith & Nephew Richards, Inc. · 1996 · recalled0
  1. Root · 2 generations backK962137 Genesis II Constrained SystemSmith & Nephew Richards, Inc. · 1996 · Recalled
  2. 1 generation backK041106 Revision Knee SystemSmith & Nephew, Inc. · 2004 · Recalled
  3. This deviceK043440 Smith & Nephew, Inc. Revision Knee SystemSmith & Nephew, Inc. · 2005
  4. Descendant · cites it directly · 10 of its ownK060742 Legion Cobalt Chrome Revision Knee SystemSmith & Nephew, Inc. · 2006 · Recalled
  5. Descendant · cites it directly · 5 of its ownK121393 Smith & Nephew, Inc. Total Knee System InstrumentationSmith & Nephew, Inc. · 2012 · Recalled
  6. Descendant · cites it directly · 5 of its ownK053390 Columbus (CR) Total Knee SystemAesculap, Inc. · 2006 · Recalled
  7. Descendant · cites it directly · 4 of its ownK180334 LEGION Knee SystemSmith & Nephew, Inc. · 2018
  8. Descendant · cites it directly · 4 of its ownK151118 Smith and Nephew Legion Hinge Knee SystemSmith & Nephew · 2015

Showing 5 of 31 descendants. See all descendants

Descendants cleared per year

2005: 12006: 32010: 12011: 22012: 32013: 12014: 12015: 32017: 12018: 12019: 22020: 12021: 22022: 12023: 22024: 22025: 32026: 1
2005high 3 (2006) · low 0 (2007)2026

Most influential descendants

Showing 5 of 31 · See all
  • K060742Legion Cobalt Chrome Revision Knee SystemSmith & Nephew, Inc. · 2006 · cites it directly10descendants
  • K121393Smith & Nephew, Inc. Total Knee System InstrumentationSmith & Nephew, Inc. · 2012 · cites it directly5descendants
  • K053390Columbus (CR) Total Knee SystemAesculap, Inc. · 2006 · cites it directly5descendants
  • K180334LEGION Knee SystemSmith & Nephew, Inc. · 2018 · cites it directly4descendants
  • K151118Smith and Nephew Legion Hinge Knee SystemSmith & Nephew · 2015 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K962137Genesis II Constrained SystemSmith & Nephew Richards, Inc.19961
−1K041106Revision Knee SystemSmith & Nephew, Inc.20041
0 (this device)K043440Smith & Nephew, Inc. Revision Knee SystemSmith & Nephew, Inc.2005—
+1K060742Legion Cobalt Chrome Revision Knee SystemSmith & Nephew, Inc.20063
+1K121393Smith & Nephew, Inc. Total Knee System InstrumentationSmith & Nephew, Inc.20124
+1K053390Columbus (CR) Total Knee SystemAesculap, Inc.20061
+1K180334LEGION Knee SystemSmith & Nephew, Inc.2018—
+1K151118Smith and Nephew Legion Hinge Knee SystemSmith & Nephew2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200630
201010
201120
201230
201310
201410
201530
201710
201810
201920
202010
202120
202210
202320
202420
202530
202610
FIG. 1 — Lineage of K043440, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JWH device

  1. #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041106Revision Knee System
Summary page 1
…with bone cement. These indications are the same as currently used for the Genesis IIConstrained Knee System and Revision Knee System cleared via K962137 and K041106. B. DEVICE DESCRIPTION: The Smith & Nephew Revision Knee System contains femoral knee components, tibial components, angle…
Smith & Nephew, Inc.20041
*K951987Genesis II Total Knee System
Summary page 1
…edges. The designs of these devices are based upon existing components of the Genesis 1l Total Knee System and Revision System, previously cleared for market under K951987, K953274, K962137, and K041106. C. SUBSTANTIAL EQUIVALENCE INFORMATION: The Smith & Nephew Revision Knee System Component…
Smith & Nephew Richards, Inc.19956
*K953274Genesis Porous Tibial Trays & Revision Knee System
Summary page 1
…e designs of these devices are based upon existing components of the Genesis 1l Total Knee System and Revision System, previously cleared for market under K951987, K953274, K962137, and K041106. C. SUBSTANTIAL EQUIVALENCE INFORMATION: The Smith & Nephew Revision Knee System Components are sim…
Smith & Nephew Richards, Inc.1996—
*K962137Genesis II Constrained System
Summary page 1
…implantation with bone cement. These indications are the same as currently used for the Genesis IIConstrained Knee System and Revision Knee System cleared via K962137 and K041106. B. DEVICE DESCRIPTION: The Smith & Nephew Revision Knee System contains femoral knee components, tibial componen…
Smith & Nephew Richards, Inc.19961
*—Zimmer Legacy Constrained Condylar Knee (L-CCK)
Named without a submission number
—
*—Biomet Oncology Salvage System (OSS)
Named without a submission number
—
*—Sulzer Orthopedics MOST System
Named without a submission number
—

Also cleared against K041106: 4 other devices →

What descends from it? (31)

See all 31

11 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060742Legion Cobalt Chrome Revision Knee SystemSmith & Nephew, Inc.200610
K121393Smith & Nephew, Inc. Total Knee System InstrumentationSmith & Nephew, Inc.20125
K053390Columbus (CR) Total Knee SystemAesculap, Inc.20065
K180334LEGION Knee SystemSmith & Nephew, Inc.20184
K151118Smith and Nephew Legion Hinge Knee SystemSmith & Nephew20154
K230653Smith & Nephew, Inc. ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and…Smith & Nephew, Inc.20233
K100981United Stem, Psa TypeUnited Orthopedic Corp.20103
K202924LinkSymphoKnee SystemWaldemar Link GmbH & Co. KG20212
K052265Smith & Nephew Competitor Duo Knee FemoralSmith & Nephew, Inc.20052
K103837Balanced Knee System Revision Offset TibiaOrtho Development20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 2, 2024Class IIZ-0278-2025Packaging change controlComplaint received indicating that a package contained a LGN OX CONSTRAINED FEM 3 LT implant instead of a LGN OX CONSTRAINED FEM 4 LT as described on the product label.
Oct 2, 2024Class IIZ-0277-2025Packaging change controlComplaint received indicating that a package contained a LGN OX CONSTRAINED FEM 3 LT implant instead of a LGN OX CONSTRAINED FEM 4 LT as described on the product label.
May 19, 2021Class IIZ-1951-2021Nonconforming Material/ComponentThe device components experienced a manufacturing process error that resulted in surface damage from product handling.
Jan 15, 2018Class IIZ-2538-2019Process controlThe affected products were inadvertently packaged with an incorrect screw component.
Jan 18, 2016Class IIZ-0816-2016Packaging change controlThe package for catalog item, a LEGION Hemi Stepped Tibial Wedge, 5 MM Size 5-6 LT-LAT/ RT-MDL, should include both a screw-on hemi stepped tibial wedge and two tibial wedge screws to affix the wedge to the tibial baseplate. A complaint…
Jan 22, 2009Not classifiedZ-1819-2009Process controlProducts were mispackaged.
Jan 22, 2009Not classifiedZ-1820-2009Process controlProducts were mispackaged.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 7.9 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K060742 (Smith & Nephew, Inc., 2006) has gone furthest, with 10 descendants of its own.
  • 4 other devices also cite its predicate K041106.

How does its category compare?

All JWH devices

JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JWH

20in 2025

2007: 182008: 172009: 112010: 192011: 252012: 332013: 312014: 332015: 272016: 182017: 252018: 252019: 282020: 222021: 242022: 172023: 202024: 192025: 202026 (partial year): 12
2007high 33 (2012) · low 11 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JWH

55in 2025

2007: 74 days median2008: 81 days median2009: 168 days median2010: 85 days median2011: 87 days median2012: 85 days median2013: 128 days median2014: 98 days median2015: 87 days median2016: 88 days median2017: 118 days median2018: 90 days median2019: 133 days median2020: 59 days median2021: 61 days median2022: 56 days median2023: 57 days median2024: 89 days median2025: 55 days median2026 (partial year): 63 days median
2007high 168 (2009) · low 55 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JWH

2in 2025

2007: 142008: 102009: 1092010: 322011: 272012: 852013: 312014: 132015: 112016: 172017: 612018: 142019: 52020: 22021: 82022: 122023: 672024: 72025: 22026 (partial year): 4
2007high 109 (2009) · low 2 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JWH

9,306in 2025

2007: 1,3052008: 1,4832009: 1,8702010: 3,2862011: 5,1332012: 5,2032013: 6,1352014: 6,7222015: 6,8812016: 7,5862017: 12,2632018: 14,0712019: 14,9872020: 12,5922021: 9,4082022: 8,3642023: 8,9472024: 10,8282025: 9,306
2007high 14,987 (2019) · low 1,305 (2007)2025

MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071874141,305
20081781101,483
2009111681091,870
20101985323,286
20112587275,133
20123385855,203
201331128316,135
20143398136,722
20152787116,881
20161888177,586
2017251186112,263
201825901414,071
201928133514,987
20202259212,592
2021246189,408
20221756128,364
20232057678,947
20241989710,828
2025205529,306
202612634—

What is it indicated for?

The Revision Knee System Components are indicated for: 1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result. 3. Failed osteotomies, unicompartmental replacement, or total knee replacement. 4. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior…

Read the full statement (100 words)

The Revision Knee System Components are indicated for:

1. Rheumatoid arthritis.

2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result.

3. Failed osteotomies, unicompartmental replacement, or total knee replacement.

4. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.

The Revision Knee Components are for single use only and are intended for implantation with bone cement.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043440: Smith & Nephew, Inc. Revision Knee System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K043440

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026