Philips Medical Systems North America, Inc. · KPR System, X-Ray, Stationary

Cleared on Feb 9, 2005 as substantially equivalent to K023441 (Philips Medical Systems, 2002) after a 16-day review. Its family tree reaches back 2 generations to 1996. It has been recalled 21 times.

Decided
Feb 9, 2005
Review
16 days
Class
Class II
Product code
KPR
Recalls
21
Predicates
1

Its longest line of descent runs back 2 generations to K961374 (1996). No later device cites it as a predicate yet.

  1. Root · 2 generations backK961374 Philips Multidiagnost 4Philips Medical Systems, Inc. · 1996 · Recalled
  2. 1 generation backK023441 Philips Multidiagnost ElevaPhilips Medical Systems · 2002 · Recalled
  3. This deviceK050151 Multidiagnost ElevaPhilips Medical Systems North America, Inc. · 2005
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K961374Philips Multidiagnost 4Philips Medical Systems, Inc.19963
−1K023441Philips Multidiagnost ElevaPhilips Medical Systems200215
0 (this device)K050151Multidiagnost ElevaPhilips Medical Systems North America, Inc.2005—
FIG. 1 — Lineage of K050151, 1996–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K023441Philips Multidiagnost Eleva
Summary page 1
…Philips Medical Systems Nederland B.V. The MultiDiagnost Eleva with image intensifier and XTV imaging system received a 510(k) substantially equivalent determination in K023441 on October 30' , 2002. Device description: The MultiDiagnost Eleva (with a flat detector) is a multi-functional/universal, tilti…
Philips Medical Systems200215

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 18, 2026Class IIZ-2411-2026Radiation Control for Health and Safety ActPhilips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
Aug 22, 2024Class IIZ-3168-2024Device DesignPotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
Aug 15, 2023Class IIZ-2587-2023Device DesignA damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
Jul 19, 2023Class IIZ-2404-2023Radiation Control for Health and Safety ActA foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
Dec 30, 2019Class IIZ-1489-2020Component design/selectionA capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Feb 28, 2019Class IIZ-1055-2019Component design/selectionThe locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
Feb 19, 2019Class IIZ-1140-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Feb 19, 2019Class IIZ-1141-2019Component design/selectionAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or…
Mar 26, 2018Class IIZ-1606-2018Nonconforming Material/ComponentThe fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Mar 26, 2018Class IIZ-1608-2018Nonconforming Material/ComponentThe fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Mar 26, 2018Class IIZ-1607-2018Nonconforming Material/ComponentThe fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Mar 26, 2018Class IIZ-1609-2018Nonconforming Material/ComponentThe fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Mar 26, 2018Class IIZ-1603-2018Nonconforming Material/ComponentThe fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Sep 24, 2014Class IIZ-1279-2016Device DesignLoss of key image functionality due to a bent pedal of the Footswitch.
Sep 24, 2014Class IIZ-1280-2016Device DesignLoss of key image functionality due to a bent pedal of the Footswitch.
Oct 14, 2013Class IIZ-1045-2015Radiation Control for Health and Safety ActThe possibility of the Reference Air Kerma Rate for low dose mode will not be reduced to 50% of the value for the normal mode, but will be approximately 70-80%, leading to a higher patient dose than expected.
Jul 11, 2012Class IIZ-2156-2012Software designWhen doing a 3 Dimensional Rotational Angiography (3DRA) the 3D reconstruction does not always start, even when all images have been acquired.
May 21, 2012Class IIZ-1829-2012Device DesignMultiDiagnost Eleva X--Ray system may have damage to System Cables from repeated movement
Jan 9, 2012Not classifiedZ-0936-2012Software designThe skin dose and skin dose rate values are calculated using an incorrect point of reference. Therefore, the AKR as displayed and stored in patient records is too low, although the dose as received by the patient is not affected.
Dec 30, 2008Not classifiedZ-0623-2009Radiation Control for Health and Safety ActSystem fails to meet requirements when it is used at kV levels greater than 115 kV at certain frame rates and in certain acquisition modes, as grid switch leakage may occur.
Nov 29, 2007Not classifiedZ-1137-2008Software designDelay in Imaging: MultiDiagnost Eleva with Flat Detector x-ray system cannot perform fluoroscopy procedures while certain image file housekeeping operations are performed. Fluoroscopy cannot be performed and new exposures are not able to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 6.2 years on median.
  • 2 of the 2 ancestors have a recall of any class.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

As a multi-functional/universal system, general R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas: * Digestive System: Swallowing studies, Oesophagus, Stomach, Small intestine, Colon, Defeacography, ERCP, T-tube cholangiogram, Liver biopsies, Transjugular Intrahepatic Portosystemic Shunts (TIPS). * Skeletal System: Bone studies. * Urinary…

Read the full statement (99 words)

As a multi-functional/universal system, general R/F, Fluoroscopy, Radiography and Angiography can be performed along with more specialized interventional applications. This includes the following general areas:

* Digestive System:

Swallowing studies, Oesophagus, Stomach, Small intestine, Colon, Defeacography, ERCP, T-tube cholangiogram, Liver biopsies, Transjugular Intrahepatic Portosystemic Shunts (TIPS).

* Skeletal System:

Bone studies.

* Urinary System:

IVP, Cystograms, Percutaneous Nephrolithotomy, Nephrostomy tube placement.

* Reproductive System:

Hysterosalpingograms, Vena spermatica, Cavemography.

* Respiratory System:

Thorax, Bronchoscopy, Pulmonary biopsies.

* Circulatory System:

Venography, Arteriography, Thrombolytic Therapy, Embolizations, Embolectomy, TVC filter placement, Dilatations, Stent placement.

* Various:

Arthrograms, Myelograms, Facet joint injections, Discography, Sialography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050151: Multidiagnost Eleva.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050151

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026