510(k) Report

510(k) · Cleared

Recalled · 2011

4D Integrated Treatment Console

K091132510(k) SpecialCleared Jun 17, 200958-day reviewClass II21 CFR 892.5050Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Varian Medical Systems · IYE Accelerator, Linear, Medical

Cleared on Jun 17, 2009 as substantially equivalent to K050479 (Varian Medical Systems, 2005) and 1 other predicate after a 58-day review. Its family tree reaches back 3 generations to 2000. 5 later devices descend from it. It has been recalled 5 times.

Decided
Jun 17, 2009
Review
58 days
Class
Class II
Product code
IYE
Recalls
5
Predicates
2

Its longest line of descent runs back 3 generations to K001643 (2000). 5 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2014: 2 descendants cleared2012: 1 descendant cleared2021: 2 descendants clearedK001643 · Varis 1.4G · Varian Medical Systems, Inc. · 2000 · recalledK050479 · 4D Integrated Treatment Console · Varian Medical Systems · 2005 · recalledK081036 · 4D Integrated Treatment Console with Patient Accessory Verification System (Pavs) · Varian Medical Systems, Inc. · 2008 · recalled200020052010201520202025
  1. Root · 3 generations backK001643 Varis 1.4GVarian Medical Systems, Inc. · 2000 · Recalled
  2. 2 generations backK050479 4D Integrated Treatment ConsoleVarian Medical Systems · 2005 · Recalled
  3. 1 generation backK081036 4D Integrated Treatment Console with Patient Accessory Verification…Varian Medical Systems, Inc. · 2008 · Recalled
  4. This deviceK091132 4D Integrated Treatment ConsoleVarian Medical Systems · 2009
  5. Descendant · cites it directly · 1 of its ownK133331 4D Integrated Treatment ConsoleVarian Medical Systems, Inc. · 2014 · Recalled
  6. Descendant · cites it directly · 1 of its ownK122616 Panther Ois/R & V Model Version 2.0Prowess, Inc. · 2012
  7. Descendant · 2 generations on · 0 of its ownK211760 Panther OISProwess, Inc. · 2021
  8. Descendant · 2 generations on · 0 of its ownK203766 4D Integrated Treatment ConsoleVarian Medical Systems, Inc. · 2021
  9. Descendant · cites it directly · 0 of its ownK142268 Varian Verification System (VVS)Varian Medical Systems, Inc. · 2014

Most influential descendants

Showing 5 of 5 · See all
  • K1333314D Integrated Treatment ConsoleVarian Medical Systems, Inc. · 2014 · cites it directly1descendant
  • K122616Panther Ois/R & V Model Version 2.0Prowess, Inc. · 2012 · cites it directly1descendant
  • K211760Panther OISProwess, Inc. · 2021 · 2 generations on0descendants
  • K2037664D Integrated Treatment ConsoleVarian Medical Systems, Inc. · 2021 · 2 generations on0descendants
  • K142268Varian Verification System (VVS)Varian Medical Systems, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K001643Varis 1.4GVarian Medical Systems, Inc.20008
−2K0504794D Integrated Treatment ConsoleVarian Medical Systems20052
−1K0810364D Integrated Treatment Console with Patient Accessory Verification System…Varian Medical Systems, Inc.20083
0 (this device)K0911324D Integrated Treatment ConsoleVarian Medical Systems2009—
+1K1333314D Integrated Treatment ConsoleVarian Medical Systems, Inc.20141
+1K122616Panther Ois/R & V Model Version 2.0Prowess, Inc.2012—
+2K211760Panther OISProwess, Inc.2021—
+2K2037664D Integrated Treatment ConsoleVarian Medical Systems, Inc.2021—
+1K142268Varian Verification System (VVS)Varian Medical Systems, Inc.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201420
202120
FIG. 1 — Lineage of K091132, 2000–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K0504794D Integrated Treatment Console
Summary page 1
…Class II Product Code: MUJ Common/Usual Name: 4DITC Predicate Devices: 4D Integrated Treatment Console, K050479 4D Integrated Treatment Console with PAVS, K081036 Device Description: The 4D In…
Varian Medical Systems20052
*K0810364D Integrated Treatment Console with Patient Accessory Verification System (Pavs)
Summary page 1
…4DITC Predicate Devices: 4D Integrated Treatment Console, K050479 4D Integrated Treatment Console with PAVS, K081036 Device Description: The 4D Integrated Console provides the user with: a…
Varian Medical Systems, Inc.20083

Also cleared against K050479: 2 other devices →

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K1333314D Integrated Treatment ConsoleVarian Medical Systems, Inc.20141
K122616Panther Ois/R & V Model Version 2.0Prowess, Inc.20121
K142268Varian Verification System (VVS)Varian Medical Systems, Inc.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 12, 2011Not classifiedZ-0691-2012Device DesignVarian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds…
Dec 12, 2011Not classifiedZ-0715-2012Device DesignWhen multiple conical collimator treatment plans are scheduled to be treated in a session, the Plan Label should be displayed on the BCCV dialog box should match the Plan Label on the 4DITC. However, the name of the plan is not being…
Jul 21, 2011Not classifiedZ-3194-2011PendingThe 4D Integrated Treatment Console may erratically fail to load the planned physical wedge accessory without warning or notification when receiving an approved plan from the oncology information system.
Feb 28, 2011Not classifiedZ-2119-2011Software designA software anomaly, in that the Multileaf Collimator (MLC) message displayed on the 4DITC on versions 8.6 and below, does NOT interlock beam delivery.
Oct 16, 2009Not classifiedZ-0891-2011Device DesignSoftware anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 3.1 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K133331 (Varian Medical Systems, Inc., 2014) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K050479.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The 4DITC function is designed to assist the operator of a radiation therapy device in providing accurate treatment setups for each patient by monitoring setup parameters and preventing the radiation therapy device from commencing irradiation when any parameter is out of conformance with the treatment plan.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091132: 4D Integrated Treatment Console.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K091132

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026