510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

H12+ Holter Recorder

K050896510(k) TraditionalCleared Aug 24, 2005138-day reviewClass II21 CFR 870.2800Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mortara Instrument, Inc. · DSH Recorder, Magnetic Tape, Medical

Cleared on Aug 24, 2005 as substantially equivalent to K911097 (Mortara Instrument, Inc., 1991) after a 138-day review. Its family tree reaches back 1 generation to 1991. 1 of the 12 devices in its family, counting itself, has a Class I recall. 10 later devices descend from it. It has been recalled once.

Decided
Aug 24, 2005
Review
138 days
Class
Class II
Product code
DSH
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K911097 (1991). 10 later devices descend from it, 2 citing it directly. 1 of the 12 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2016: 2 descendants cleared2009: 1 descendant cleared2015: 1 descendant cleared, 1 with a Class I recall12026: 2 descendants cleared2018: 2 descendants cleared2019: 2 descendants clearedK911097 · PR4 · Mortara Instrument, Inc. · 19911990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 3 descendants, 1 with a Class I recall13 generations forward: 4 descendants4 generations forward: 1 descendant K911097 · PR4 · Mortara Instrument, Inc. · 19910
  1. Root · 1 generation backK911097 PR4Mortara Instrument, Inc. · 1991
  2. This deviceK050896 H12+ Holter RecorderMortara Instrument, Inc. · 2005
  3. Descendant · cites it directly · 6 of its ownK091505 COR12+Viasys Healthcare GmbH · 2009
  4. Descendant · 2 generations on · 5 of its ownK150810 Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH · 2015 · Recalled (Class I)
  5. Descendant · cites it directly · 2 of its ownK151787 Holter SystemEdan Instruments, Inc. · 2016
  6. Descendant · 3 generations on · 1 of its ownK181524 Vyntus ONE with CART, Vyntus ONE PFT only with CART, Vyntus ONE with…Vyaire Medical, Inc. · 2018
  7. Descendant · 2 generations on · 0 of its ownK254119 Wearable ECG Recorder (YMHeart XY-12); Wearable ECG Recorder…Shenzhen Xinyan Medical Technology Co.,ltd. · 2026

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2009: 12015: 12016: 22018: 22019: 22026: 2
2009high 2 (2016) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K091505COR12+Viasys Healthcare GmbH · 2009 · cites it directly6descendants
  • K150810Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH · 2015 · 2 generations on5descendants
  • K151787Holter SystemEdan Instruments, Inc. · 2016 · cites it directly2descendants
  • K181524Vyntus ONE with CART, Vyntus ONE PFT only with CART, Vyntus ONE with adjustable CART…Vyaire Medical, Inc. · 2018 · 3 generations on1descendant
  • K254119Wearable ECG Recorder (YMHeart XY-12); Wearable ECG Recorder (YMHeart XY-1201); Wearable…Shenzhen Xinyan Medical Technology Co.,ltd. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K911097PR4Mortara Instrument, Inc.1991—
0 (this device)K050896H12+ Holter RecorderMortara Instrument, Inc.2005—
+1K091505COR12+Viasys Healthcare GmbH2009—
+2K150810Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH2015Class I
+1K151787Holter SystemEdan Instruments, Inc.2016—
+3K181524Vyntus ONE with CART, Vyntus ONE PFT only with CART, Vyntus ONE with…Vyaire Medical, Inc.2018—
+2K254119Wearable ECG Recorder (YMHeart XY-12); Wearable ECG Recorder (YMHeart…Shenzhen Xinyan Medical Technology Co.,ltd.2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201511
201620
201820
201920
202620
FIG. 1 — Lineage of K050896, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSH device

  1. #1 K260157 LifeLens Physiological Monitorscore 17.9 · 22 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K911097PR4
Summary page 1
…to which S. E. is claimed The H1 2+ Holler Recorder is a modification of a legally marketed Mortara predicate device. Mortara PR4 Helter analysis Recorder (K91 10977) Note: For commercial reasons, Mortara changed the proprietary name from PR 4 to H-12 in 1995 and has continued…
Mortara Instrument, Inc.1991—

Also cleared against K911097: 1 other device →

What descends from it? (10)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091505COR12+Viasys Healthcare GmbH20096
K151787Holter SystemEdan Instruments, Inc.20162

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 18, 2020Class IIZ-1812-2020Device DesignImpacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K150810Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K091505 (Viasys Healthcare GmbH, 2009) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K911097.

How does its category compare?

All DSH devices

DSH Recorder, Magnetic Tape, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DSH

4in 2025

2007: 12008: 12009: 32010: 22011: 62012: 22013: 22014: 12015: 22016: 22017: 32018: 32019: 32020: 42021: 42022: 42023: 62024: 22025: 42026 (partial year): 2
2007high 6 (2011) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSH

205in 2025

2009: 84 days median2011: 64 days median2017: 91 days median2018: 269 days median2019: 30 days median2020: 122 days median2021: 263 days median2022: 427 days median2023: 198 days median2025: 205 days median
2007high 427 (2022) · low 30 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSH

No recall names DSH since 2007.

Adverse event reports per year · DSH

213in 2025

2009: 12010: 82011: 32012: 12013: 22014: 32015: 82016: 62017: 312018: 302019: 762020: 912021: 1132022: 462023: 652024: 882025: 213
2007high 213 (2025) · low 0 (2007)2025

MAUDE reports naming product code DSH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
200938401
20102—08
201166403
20122—01
20132—02
20141—03
20152—08
20162—06
2017391031
20183269030
2019330076
20204122091
202142630113
20224427046
20236198065
20242—088
202542050213
20262—0—

What is it indicated for?

The H 12+ Holter recorder is intended to acquire, record and store ECG data of patients thathave been connected to the Mortara H 12+ recorder and are undergoing Holter monitoring. The Hi 2+ acquires, digitizes and stores data to be analyzed by the H-Scribe Hoiter system.

The H12+ is indicated for use:

* in a clinical setting, by qualified medical professionals only, for recording ECG data of symptomatic patients requiring ambulatory (Holter) monitoring of up to 48 hours.

* The H12+ performs no cardiac analysis by itself and is intended to be used with the H-Scribe Hotter analysis system (K004017). ECG data prerecorded by the HI12+ is acquired and analyzed by the H-Scribe.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050896: H12+ Holter Recorder.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050896

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026