Cleared on Jun 10, 2009 as substantially equivalent to K050896 (Mortara Instrument, Inc., 2005) and 2 other predicates after a 20-day review. Its family tree reaches back 4 generations to 2000. 1 of the 17 devices in its family, counting itself, has a Class I recall. 6 later devices descend from it.

Decided
Jun 10, 2009
Review
20 days
Class
Class II
Product code
DSH
Recalls
None
Predicates
3

Its longest line of descent runs back 4 generations to K002074 (2000). 10 ancestors in all, the oldest K911097 (1991). 6 later devices descend from it, 1 citing it directly. 1 of the 17 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2005: 2 other ancestors2007: 2 other ancestors1991: 1 other ancestor2015: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared2018: 2 descendants cleared2019: 2 descendants clearedK002074 · AB Cardiette Daedalus View Base and AB Cardiettte Daedalus View Hes · H&C Medical Devices Spa · 2000K070614 · Corscreen · Viasys Healthcare GmbH · 2007K080734 · Flowscreen, Flowscreen/Corscreen, Flowscreen CT · Viasys Healthcare GmbH · 2008K082539 · Masterscope, Masterscope Ecg, Masterscope CT · Cardinal Health Germany 234 GmbH · 20081990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 4 descendants3 generations forward: 1 descendant K082539 · Masterscope, Masterscope Ecg, Masterscope CT · Cardinal Health Germany 234 GmbH · 2008K080734 · Flowscreen, Flowscreen/Corscreen, Flowscreen CT · Viasys Healthcare GmbH · 2008K070614 · Corscreen · Viasys Healthcare GmbH · 2007K002074 · AB Cardiette Daedalus View Base and AB Cardiettte Daedalus View Hes · H&C Medical Devices Spa · 20000
  1. Root · 4 generations backK002074 AB Cardiette Daedalus View Base and AB Cardiettte Daedalus View HesH&C Medical Devices Spa · 2000
  2. 3 generations backK070614 CorscreenViasys Healthcare GmbH · 2007
  3. 2 generations backK080734 Flowscreen, Flowscreen/Corscreen, Flowscreen CTViasys Healthcare GmbH · 2008
  4. 1 generation backK082539 Masterscope, Masterscope Ecg, Masterscope CTCardinal Health Germany 234 GmbH · 2008
  5. This deviceK091505 COR12+Viasys Healthcare GmbH · 2009
  6. Descendant · cites it directly · 5 of its ownK150810 Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH · 2015 · Recalled (Class I)
  7. Descendant · 2 generations on · 1 of its ownK181524 Vyntus ONE with CART, Vyntus ONE PFT only with CART, Vyntus ONE with…Vyaire Medical, Inc. · 2018
  8. Descendant · 2 generations on · 0 of its ownK183567 Vyntus/SentrySuite Product LineVyaire Medical · 2019
  9. Descendant · 3 generations on · 0 of its ownK190853 Vyntus BODYVyaire Medical, Inc. · 2019
  10. Descendant · 2 generations on · 0 of its ownK183369 COR12 EcgCorscience GmbH & Co. · 2018

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12018: 22019: 2
2015high 2 (2018) · low 0 (2017)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K150810Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH · 2015 · cites it directly5descendants
  • K181524Vyntus ONE with CART, Vyntus ONE PFT only with CART, Vyntus ONE with adjustable CART…Vyaire Medical, Inc. · 2018 · 2 generations on1descendant
  • K183567Vyntus/SentrySuite Product LineVyaire Medical · 2019 · 2 generations on0descendants
  • K190853Vyntus BODYVyaire Medical, Inc. · 2019 · 3 generations on0descendants
  • K183369COR12 EcgCorscience GmbH & Co. · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K002074AB Cardiette Daedalus View Base and AB Cardiettte Daedalus View HesH&C Medical Devices Spa2000—
−3K070614CorscreenViasys Healthcare GmbH2007—
−2K080734Flowscreen, Flowscreen/Corscreen, Flowscreen CTViasys Healthcare GmbH2008—
−1K082539Masterscope, Masterscope Ecg, Masterscope CTCardinal Health Germany 234 GmbH2008—
0 (this device)K091505COR12+Viasys Healthcare GmbH2009—
+1K150810Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH2015Class I
+2K181524Vyntus ONE with CART, Vyntus ONE PFT only with CART, Vyntus ONE with…Vyaire Medical, Inc.2018—
+2K183567Vyntus/SentrySuite Product LineVyaire Medical2019—
+3K190853Vyntus BODYVyaire Medical, Inc.2019—
+2K183369COR12 EcgCorscience GmbH & Co.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
200520
200720
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201511
201610
201820
201920
FIG. 1 — Lineage of K091505, 2000–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSH device

  1. #1 K260157 LifeLens Physiological Monitorscore 17.9 · 22 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050896H12+ Holter Recorder
Summary page 2
…rt 870 Code DSH 5.8 Reason for Premarket Notification New Device 5.9 Legally predicate marketed device 1. H12+ Holter Recorder K050896 Code DSH 2. Cardiette Microtel ECG K082124 Code DPS 3. MasterScope ECG K082539…
Mortara Instrument, Inc.20051
*K082124Carditte Microtel
Summary page 2
…ion New Device 5.9 Legally predicate marketed device 1. H12+ Holter Recorder K050896 Code DSH 2. Cardiette Microtel ECG K082124 Code DPS 3. MasterScope ECG K082539 Code BTY I DPS 5.10 Predicate Device Company…
Et Medical Devices Spa2009—
*K082539Masterscope, Masterscope Ecg, Masterscope CT
Summary page 2
…ce 1. H12+ Holter Recorder K050896 Code DSH 2. Cardiette Microtel ECG K082124 Code DPS 3. MasterScope ECG K082539 Code BTY I DPS 5.10 Predicate Device Company 1. Mortara Instrument, INC. 2. Et Medical devices SpA…
Cardinal Health Germany 234 GmbH2008—

Also cleared against K050896: 1 other device →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150810Vyntus / SentrySuite Product Line Class ICarefusion Germany 234 GmbH20155

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K150810Vyntus / SentrySuite Product LineCarefusion Germany 234 GmbH · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • 1 of the 10 ancestors has a recall of any class.
  • Of its descendants, K150810 (Carefusion Germany 234 GmbH, 2015) has gone furthest, with 5 descendants of its own.
  • 1 other device also cites its predicate K050896.

How does its category compare?

All DSH devices

DSH Recorder, Magnetic Tape, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DSH

4in 2025

2007: 12008: 12009: 32010: 22011: 62012: 22013: 22014: 12015: 22016: 22017: 32018: 32019: 32020: 42021: 42022: 42023: 62024: 22025: 42026 (partial year): 2
2007high 6 (2011) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSH

205in 2025

2009: 84 days median2011: 64 days median2017: 91 days median2018: 269 days median2019: 30 days median2020: 122 days median2021: 263 days median2022: 427 days median2023: 198 days median2025: 205 days median
2007high 427 (2022) · low 30 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSH

No recall names DSH since 2007.

Adverse event reports per year · DSH

213in 2025

2009: 12010: 82011: 32012: 12013: 22014: 32015: 82016: 62017: 312018: 302019: 762020: 912021: 1132022: 462023: 652024: 882025: 213
2007high 213 (2025) · low 0 (2007)2025

MAUDE reports naming product code DSH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
200938401
20102—08
201166403
20122—01
20132—02
20141—03
20152—08
20162—06
2017391031
20183269030
2019330076
20204122091
202142630113
20224427046
20236198065
20242—088
202542050213
20262—0—

What is it indicated for?

The Cor12+ Holter Recorder is intended to acquire, record and store ECG data of patients that have been connected to the Cor12+ recorder and are undergoing Holter monitoring. The Corl2+ acquires, digitizes and stores data to be analyzed on a computer with special software.

The Cor12+ is indicated for use:

in a clinical setting, by qualified medical professionals only, for recording ECG data of symptomatic patients requiring ambulatory (Holter) monitoring of up to 24 hours.

The Corl2+ performs no cardiac analysis by itself; it solely records the ECG signals. The patient population for this device is from 4 years on and a minimum weight of 10 kg.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091505: COR12+.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091505

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026