510(k) Report

510(k) · Cleared

Recalled · Class I · 2020

Heartstart MRX with Q-CPR Option, Models M3535A or M3536A

K051134510(k) TraditionalCleared Sep 8, 2005128-day reviewClass III21 CFR 870.5310Recalls 39 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Sep 8, 2005 after a 128-day review. The FDA summary names its predicate without a number. 3 of the 14 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 39 times, including 2 Class I recalls.

Decided
Sep 8, 2005
Review
128 days
Class
Class III
Product code
MKJ
Recalls
39 (2 Class I)
Predicates
0

13 later devices descend from it, 6 citing it directly. 3 of the 14 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2006: 2 descendants cleared, 2 with a Class I recall22007: 1 descendant cleared2012: 1 descendant cleared2013: 3 descendants cleared2015: 2 descendants cleared2018: 2 descendants cleared2020: 1 descendant cleared2021: 1 descendant cleared20052010201520202025
  1. This deviceK051134 Heartstart MRX with Q-CPR Option, Models M3535A or M3536APhilips Medical Systems · 2005
  2. Descendant · cites it directly · 5 of its ownK122050 Cprmeter CPR Feedback DeviceLaerdal Medical A/S · 2012
  3. Descendant · cites it directly · 4 of its ownK061707 Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems · 2006 · Recalled (Class I)
  4. Descendant · 2 generations on · 3 of its ownK151702 Reusable Full Silicone CoverLaerdal Medical AS · 2015
  5. Descendant · 3 generations on · 2 of its ownK173886 CPRmeter 2 CPR Feedback DeviceLaerdal Medical AS · 2018
  6. Descendant · 2 generations on · 2 of its ownK130153 Hearstart MRX with Airway Confimration AssistPhilips Medical Systems · 2013

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2006: 22007: 12012: 12013: 32015: 22018: 22020: 12021: 1
2006high 3 (2013) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K122050Cprmeter CPR Feedback DeviceLaerdal Medical A/S · 2012 · cites it directly5descendants
  • K061707Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems · 2006 · cites it directly4descendants
  • K151702Reusable Full Silicone CoverLaerdal Medical AS · 2015 · 2 generations on3descendants
  • K173886CPRmeter 2 CPR Feedback DeviceLaerdal Medical AS · 2018 · 3 generations on2descendants
  • K130153Hearstart MRX with Airway Confimration AssistPhilips Medical Systems · 2013 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K051134Heartstart MRX with Q-CPR Option, Models M3535A or M3536APhilips Medical Systems2005—
+1K122050Cprmeter CPR Feedback DeviceLaerdal Medical A/S2012—
+1K061707Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems2006Class I
+2K151702Reusable Full Silicone CoverLaerdal Medical AS2015—
+3K173886CPRmeter 2 CPR Feedback DeviceLaerdal Medical AS2018—
+2K130153Hearstart MRX with Airway Confimration AssistPhilips Medical Systems2013—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200622
200710
201210
201330
201520
201820
202010
202110
FIG. 1 — Lineage of K051134, 2005–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Philips Medical Systems HeartStart MRx monitor/defibrillator
Named without a submission number
—
*—ZOLL Medical Corp. Zoll AED Plus with CPR-D Padz external defibrillator
Named without a submission number
—

What descends from it? (13)

See all 13

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122050Cprmeter CPR Feedback DeviceLaerdal Medical A/S20125
K061707Heartstart MRX, Invasive Pressure & Temperature Options Class IPhilips Medical Systems20064
K130684Philips Heartstart FR3 Automated External DefibrillatorPhilips Medical Systems20130
K123597Truecpr Coaching DevicePhysio-Control, Inc.20130
K063375Heartstart MRX, Models M3535A or M3536APhilips Medical Systems20070
K062233Heartstart MRX Monitor/Defibrillator Class IPhilips Medical Systems North America, Inc.20060

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 27, 2020Class IIZ-2157-2020Process change controlHeartStart XL MonitorlDefibrillators may not have been included in prior field actions and received required corrections
Nov 23, 2018Class IIZ-0603-2019Nonconforming Material/ComponentAffected lithium-ion batteries may contain a defective component (Thermal Cut-off or TCO). Should the component fail, the battery will no longer charge or deliver power to the Monitor/Defibrillator, which may prevent operation of the…
Jan 29, 2018Class IIZ-0684-2018Device DesignInform users of correct placement of the disposable adhesive pad on the QCPR meter and inform users of injuries associated with proper performance of CPR. As required by Consent Decree of Permanent Injunction No. 1:17-cv-11955-DJC…
Sep 28, 2017Class IIZ-0278-2018Nonconforming Material/ComponentPhilips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).
Jul 17, 2017Class IIZ-0348-2018Device DesignIssue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference…
Jul 12, 2017Class IZ-0489-2018Nonconforming Material/ComponentAffected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Feb 24, 2017Class IZ-1417-2017Device DesignHeartStart MRx Monitor/Defibrillator may may exhibit abnormal behavior when operating using a single battery as the sole source of power may fail to deliver therapy or monitoring
Aug 26, 2016Class IIZ-2842-2016Device DesignThe handle can separate from the MRx housing due to breakage of mounts on the rear case.
Apr 14, 2016Class IIZ-1612-2016Employee errorMRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component.
Oct 14, 2015Class IIZ-0320-2016Software designThe following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earlier, and model M3536A with version T.00.05 and earlier may stop the automated Ready-For-Use (RFU) test in an abnormal state when…
Dec 23, 2014Class IIZ-0219-2016Software designMRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed
Nov 19, 2014Class IIZ-0204-2016Process control1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a…
Nov 19, 2014Class IIZ-0273-2016Device DesignWhen a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs) while…
Nov 19, 2014Class IIZ-0262-2016Device DesignMRx Defib can be susceptible to one or both issues: 1. The C02 Inlet Port associated with end-tidal carbon dioxide (EtCO2) monitoring on MRx Monitor/Defibrillators can be pushed into the MRx housing, making it inaccessible. 2. The handle…
Nov 19, 2014Class IIZ-0963-2015Software designDevice may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently.
Feb 18, 2014Class IIZ-1108-2014Software designWireless Link may be unable to transmit data when configured for a specific data flow.
Feb 18, 2014Class IIZ-1120-2014Device DesignPhilips HeartStart MRx Monitor/Defibrillator, when operating on battery power only, may experience an unexpected shutdown if exposed to elevated levels of electromagnetic interference from RF energy source
Feb 18, 2014Class IIZ-1090-2014Device DesignA component of the MRx Processor Board may be susceptible to damage from electrostatic discharge (ESD), which can disrupt ECG and SpO2 functionality leading to an inability to perform.
Feb 18, 2014Class IIZ-1118-2014Component design/selectionECG trunk cable and connector block of the MRx could be susceptible to accelerated wear, which could result in an interrupted ECG signal
Feb 18, 2014Class IIZ-1194-2014Software Design ChangeWhen the HeartStart MRx.is used with the Q-CPR Meter in defibrillation mode, the Q-CPR Meter may incorrectly display the Do Not Touch the Patient icon.
Feb 14, 2014Class IIZ-1109-2014Component design/selectionPhilips HeartStart XL may experience a power board failure causing failure to defibrillate
Feb 4, 2014Class IIZ-1110-2014Device DesignHeartStart MRx Defibrillator Monitor therapy connection could experience accelerated wear causing risk for patients and/or caregivers
Dec 10, 2013Class IIZ-0570-2014Software designMRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
Jun 19, 2013Class IIZ-1797-2013Software designThe MRx may deliver a non-synchronized cardioversion shock resulting in the delivery of incorrect therapy, which may induce ventricular fibrillation.
Jun 7, 2013Class IIZ-0593-2014Labeling Change ControlError in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10
Apr 10, 2013Class IIZ-0613-2014Environmental controlDevice Operating on Battery Power May Shutdown without Warning if exposed to elevated levels of electromagnetic interference from other radio frequency (RF) energy sources and
Apr 8, 2013Class IIZ-1189-2013Software designDevice may fail to deliver defibrillation therapy in either Manual Defib or AED mode.
Jun 20, 2012Class IIZ-1896-2012Process controlHeartStart MRx Unexpected Pads/Paddles ECG Failure
Jun 20, 2012Class IIZ-1986-2012Nonconforming Material/ComponentDefibrillator/monitor cannot analyze an ECG report during acquisition of 12 lead ECG
Jun 20, 2012Class IIZ-2347-2012Equipment maintenanceTherapy Cable Connection wear may lead to a malfunction in detection of defibrillation pads/paddles therapy Cables
Feb 7, 2011Not classifiedZ-1609-2011Software designSoftware: EtCO2 and Ventilation Rate values may be labeled incorrectly.
May 6, 2010Not classifiedZ-2293-2010Device DesignReduced sensitivity in the force measurement sensor can result in inaccurate feedback to the user on how well he or she is releasing pressure after each chest compression. This can result in incomplete release of chest compression, which…
Apr 7, 2010Not classifiedZ-1618-2010Component design/selectionECG signal processing chip may fail causing failure to deliver therapy in AED mode or the failure to indicate poor electrical contact (very high impedance) with the patient in Manual Defib mode.
Mar 4, 2010Not classifiedZ-1317-2010Component design/selectionBatteries older than 2 years have a higher potential for intermittent electrical contact, may contribute to an interruption in monitoring or therapy if the monitor/defibrillator is operating on battery.
Nov 19, 2009Not classifiedZ-1260-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1259-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1257-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 19, 2009Not classifiedZ-1258-2010Nonconforming Material/ComponentPower cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may…
Nov 12, 2009Not classifiedZ-0562-2010Device DesignHeartStart Vehicle Wall Mount (Model M5528A) with DC Power Module (Model M5529A)--Incorrect wiring during installation by user may result in a smoke or fire hazard.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K061707Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems · 2006 · Cites it directly · ◆ in the 2006 descendants bar of FIG. 12 Class I recalls
  • K062233Heartstart MRX Monitor/DefibrillatorPhilips Medical Systems North America, Inc. · 2006 · Cites it directly · ◆ in the 2006 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K122050 (Laerdal Medical A/S, 2012) has gone furthest, with 5 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The HeartStart MRx monitor/defibrillator with Q-CPR™ option is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. AED Therapy To be used in the presence of a suspected cardiac arrest on…

Read the full statement (309 words)

The HeartStart MRx monitor/defibrillator with Q-CPR™ option is for use for the termination of ventricular tachycardia and ventricular fibrillation.

The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

AED Therapy

To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.

Manual Defibrillation Therapy

Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronized defibrillation is indicated for termination of atrial fibrillation.

Non-Invasive External Pacing Therapy

The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.

Pulse Oximetry

The SpO2 option is intended for use when it is beneficial to assess a patient's oxygen saturation level.

Non-Invasive Blood Pressure Monitoring

The NIBP option is intended for noninvasive measurement of a patient's arterial blood pressure.

Endtidal CO2 Monitoring

The EtCO2 option is intended for noninvasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.

12-Lead ECG

The 12-Lead ECG option is intended to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements.

Q-CPR

The Q-CPR™ option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume, and flow rate (inflation time).

The Q-CPR option is contraindicated as follows:

* The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less than 25 kg).

* The Q-CPR option is not for use when CPR is contraindicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051134: Heartstart MRX with Q-CPR Option, Models M3535A or M3536A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051134

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026