Heartstart MRX, Models M3535A or M3536A

K063375510(k) SpecialCleared Jan 11, 200764-day reviewClass III21 CFR 870.5310Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Jan 11, 2007 as substantially equivalent to K031187 (Philips Medical Systems, 2003) and 2 other predicates after a 64-day review. Its family tree reaches back 2 generations to 2001. 3 of the 5 devices in its family, counting itself, have a Class I recall. It has been recalled 11 times.

Decided
Jan 11, 2007
Review
64 days
Class
Class III
Product code
MKJ
Recalls
11
Predicates
3

Its longest line of descent runs back 2 generations to K011865 (2001). 4 ancestors in all. No later device cites it as a predicate yet. 3 of the 5 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2005: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor, 1 with a Class I recall1K011865 · Zoll M Series Ibp Operation; Zoll M Series Temperature Option · Zoll Medical Corp · 2001K061707 · Heartstart MRX, Invasive Pressure & Temperature Options · Philips Medical Systems · 2006 · Class I recall20002002200420062008
  1. Root · 2 generations backK011865 Zoll M Series Ibp Operation; Zoll M Series Temperature OptionZoll Medical Corp · 2001
  2. 1 generation backK061707 Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems · 2006 · Recalled (Class I)
  3. This deviceK063375 Heartstart MRX, Models M3535A or M3536APhilips Medical Systems · 2007

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K011865Zoll M Series Ibp Operation; Zoll M Series Temperature OptionZoll Medical Corp2001—
−1K061707Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems2006Class I
0 (this device)K063375Heartstart MRX, Models M3535A or M3536APhilips Medical Systems2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200311
200511
FIG. 1 — Lineage of K063375, 2001–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031187Heartstart MRX, Model M3535A
Summary page 1
…Clinical Electronic Thermometer 3. The new device is substantially equivalent to the previously cleared HeartStart MRx Monitor/Defibrillator software cleared under K031187 and K051134, as well as K061707). 4. The modification is a change that provides an optional BluetoothTM wireless technology o…
Philips Medical Systems2003Class I
*K051134Heartstart MRX with Q-CPR Option, Models M3535A or M3536A
Summary page 1
…lectronic Thermometer 3. The new device is substantially equivalent to the previously cleared HeartStart MRx Monitor/Defibrillator software cleared under K031187 and K051134, as well as K061707). 4. The modification is a change that provides an optional BluetoothTM wireless technology option to tra…
Philips Medical Systems2005Class I
*K061707Heartstart MRX, Invasive Pressure & Temperature Options
Summary page 1
…r 3. The new device is substantially equivalent to the previously cleared HeartStart MRx Monitor/Defibrillator software cleared under K031187 and K051134, as well as K061707). 4. The modification is a change that provides an optional BluetoothTM wireless technology option to transfer data from the…
Philips Medical Systems2006Class I

Also cleared against K031187: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 10, 2020Class IIZ-2064-2020Process controlDamaged ESD bags used for storing components in Processor PCA Replacement Kits
Feb 26, 2020Class IIZ-1499-2020Device DesignPhilips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the…
Jan 14, 2020Class IIZ-1342-2020Process designM4758A Ambulance Parameter Extension (APE) kit may cause ECG "Leads Off" condition in hot, humid environments
Jul 16, 2019Class IIZ-2488-2019OtherNon-conforming devices are identified, which may not have been included in prior field actions and may not have had a number of corrections applicable to these devices.
Jun 7, 2019Class IIZ-1923-2019Process controlThe AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery…
May 3, 2018Class IIZ-2035-2018Employee errorThe MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.
Sep 28, 2017Class IIZ-0278-2018Nonconforming Material/ComponentPhilips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).
Apr 14, 2016Class IIZ-1612-2016Employee errorMRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component.
Dec 23, 2014Class IIZ-0219-2016Software designMRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed
Nov 19, 2014Class IIZ-0204-2016Process control1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a…
Nov 19, 2014Class IIZ-0963-2015Software designDevice may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K061707Heartstart MRX, Invasive Pressure & Temperature OptionsPhilips Medical Systems · 2006 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K051134Heartstart MRX with Q-CPR Option, Models M3535A or M3536APhilips Medical Systems · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 12 Class I recalls
  • K031187Heartstart MRX, Model M3535APhilips Medical Systems · 2003 · Direct predicate · ◆ in the 2003 ancestors bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 5.0 years on median.
  • 3 of the 4 ancestors have a recall of any class; 3 have a Class I recall.
  • 3 other devices also cite its predicate K031187.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. AED Therapy To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that…

Read the full statement (333 words)

The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.

AED Therapy

To be used in the presence of a suspected cardiac arrest on patients of at least 8 years of age that are unresponsive, not breathing and pulseless.

Manual Defibrillation

Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for termination of atrial fibrillation.

Non-invasive External Pacing Therapy

The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.

Pulse Oximetry

The SpO2 option is intended for use when it is beneficial to assess a patient's oxygen saturation level.

Non-invasive Blood Pressure Monitoring

The NBP option is intended for non-invasive measurement of a patient's arterial blood pressure.

End-tidal C02

The EtCO2 option is intended for non-invasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.

12-Lead ECG

The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead ECG report, which may include measurements and interpretative statements.

Q-CPR

The Q-CPRT M option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate, depth, and duty cycle and ventilation rate, volume and flow rate (inflation time).

The Q-CPR option is contraindicated as follows:

- The Q-CPR option is contraindicated for use on neonatal and pediatric patients (under 8 years of age or weighing less that 25 kg).

- The Q-CPR option is not for use when CPR is contraindicated.

Invasive Pressure

The Invasive Pressure option is indicated for measuring arterial, venous, intracranial and other physiological pressures on patients.

Temperature

The Temperature option is indicated for measuring temperature in patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063375: Heartstart MRX, Models M3535A or M3536A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K063375

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026