510(k) · Cleared
Recalled · Class II · 2014Ultegra System Analyzer, Ultegra System Rpfa-Trap Test Cartidges, Ultegra System Rpfa-Trap Level One QC, Ultegra System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Accumetrics, Inc. · JOZ System, Automated Platelet Aggregation
Cleared on Dec 20, 1999 as substantially equivalent to K830749 (Chrono-Log Corp., 1983) and 1 other predicate after a 145-day review. Its family tree reaches back 1 generation to 1976. It has been recalled 3 times.
- Decided
- Dec 20, 1999
- Review
- 145 days
- Class
- Class II
- Product code
- JOZ
- Recalls
- 3
- Predicates
- 2
Its longest line of descent runs back 1 generation to K760198 (1976). 2 ancestors in all. No later device cites it as a predicate yet.
- Root · 1 generation backK760198 Kit, Platelet Aggregation Reagent (#390)Chrono-Log Corp. · 1976
- This deviceK992531 Ultegra System Analyzer, Ultegra System Rpfa-Trap Test Cartidges…Accumetrics, Inc. · 1999
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K760198 | Kit, Platelet Aggregation Reagent (#390) | Chrono-Log Corp. | 1976 | — |
| 0 (this device) | K992531 | Ultegra System Analyzer, Ultegra System Rpfa-Trap Test Cartidges, Ultegra… | Accumetrics, Inc. | 1999 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1983 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JOZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JOZ device
- #1 K191364 T-TAS 01 Total Thrombus-formation Analysis…score -0.6 · 3 generations · no Class I recall
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K830749 | Platelet AggregometerSummary page 1…ccumetrics Ultegra System Analyzer and Rapid Platelet Function Assay have been found to be substantially equivalent to CHRONO-LOG Corporation's Whole Blood Aggregometer (K830749) and CHRONO-PAR Reagent (K760198). Devi escription: The Ultegra System is a turbidimetric based optical detection system which… | 1983 |
| * | K760198 | Kit, Platelet Aggregation Reagent (#390)Summary page 1…r and Rapid Platelet Function Assay have been found to be substantially equivalent to CHRONO-LOG Corporation's Whole Blood Aggregometer (K830749) and CHRONO-PAR Reagent (K760198). Devi escription: The Ultegra System is a turbidimetric based optical detection system which measures platelet induced aggrega… | 1976 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 13, 2014 | Class II | Z-1172-2014 | Nonconforming Material/Component | Accumetrics is recalling the VerifyNow IIb/IIIa test because it may result in the reporting of an erroneous low platelet aggregation unit (PAU) result. An erroneous low PAU result may cause a low percent inhibition calculation or low… |
| Nov 14, 2013 | Class II | Z-0482-2014 | Other | Accumetrics is recalling the VerifyNow IIb/IIIa 10-Test because it contains incorrect information for the amount of time in which to perform the test following collection of the patient sample. The correct instruction is to perform the… |
| Sep 13, 2005 | Not classified | Z-0102-06 | Other | In 2001, separate reports of mild shock, sparking and smoke from Accumetrics PN 29001 power supply. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 2 ancestors has a recall.
- 8 other devices also cite its predicate K830749.
How does its category compare?
All JOZ devicesJOZ System, Automated Platelet Aggregation: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JOZ
1in 2020
2026 to Oct 25, outlined.
Median 510(k) review, days · JOZ
Too few JOZ 510(k)s in any year since 2007 to give a median.
Recalls per year · JOZ
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JOZ
3in 2025
MAUDE reports naming product code JOZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 1 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 1 | — | 0 | 1 |
| 2013 | 1 | — | 1 | 1 |
| 2014 | 0 | — | 0 | 1 |
| 2015 | 2 | — | 0 | 0 |
| 2016 | 1 | — | 0 | 0 |
| 2017 | 1 | — | 0 | 1 |
| 2018 | 0 | — | 0 | 25 |
| 2019 | 1 | — | 0 | 9 |
| 2020 | 1 | — | 0 | 2 |
| 2021 | 0 | — | 0 | 2 |
| 2022 | 0 | — | 0 | 16 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 1 | 1 |
| 2025 | 0 | — | 1 | 3 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Ultegra Rapid Platelet Function Assay (RPFA) is a semi-quantitative, whole blood platelet function assay used to measure glycoprotein (GP) IIb/IIIa receptor blockade in patients treated with abciximab. Ultegra RPFA results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K992531 · printed Nov 5, 2026
Cite this page
510(k) Report. “K992531: Ultegra System Analyzer, Ultegra System Rpfa-Trap Test Cartidges, Ultegra System Rpfa-Trap Level One QC, Ultegra System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K992531
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026