510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Modification to: Cic Pro Clinical Information Center Central Station

K053356510(k) TraditionalCleared Apr 19, 2006138-day reviewClass II21 CFR 870.2450Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems Information Technologies · DXJ Display, Cathode-Ray Tube, Medical

Cleared on Apr 19, 2006 as substantially equivalent to K032370 (Ge Medical Systems Information Technologies, 2003) after a 138-day review. Its family tree reaches back 2 generations to 2000. 22 later devices descend from it. It has been recalled 2 times.

Decided
Apr 19, 2006
Review
138 days
Class
Class II
Product code
DXJ
Recalls
2
Predicates
1

Its longest line of descent runs back 2 generations to K001112 (2000). 22 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2007: 2 descendants cleared2014: 2 descendants cleared2008: 1 descendant cleared2010: 1 descendant cleared2011: 3 descendants cleared2016: 3 descendants cleared2018: 2 descendants cleared2020: 3 descendants cleared2025: 2 descendants cleared2024: 2 descendants cleared2022: 1 descendant clearedK001112 · GE Marquette Clinical Information Center, GE Marquette Cic · General Electric Medical Systems Information Techn · 2000K032370 · Cic Pro Clinical Information Center Central Station · Ge Medical Systems Information Technologies · 2003 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 5 descendants4 generations forward: 4 descendants5 generations forward: 5 descendants6 generations forward: 3 descendants K032370 · Cic Pro Clinical Information Center Central Station · Ge Medical Systems Information Technologies · 2003 · recalledK001112 · GE Marquette Clinical Information Center, GE Marquette Cic · General Electric Medical Systems Information Techn · 20000+5
  1. Root · 2 generations backK001112 GE Marquette Clinical Information Center, GE Marquette CicGeneral Electric Medical Systems Information Techn · 2000
  2. 1 generation backK032370 Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2003 · Recalled
  3. This deviceK053356 Modification to: Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2006
  4. Descendant · cites it directly · 10 of its ownK073372 Electrophysiology Systems Integrator (Epsi)Systemsone · 2007
  5. Descendant · cites it directly · 10 of its ownK062976 Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies · 2007 · Recalled
  6. Descendant · 2 generations on · 9 of its ownK133882 Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · Recalled
  7. Descendant · 2 generations on · 9 of its ownK083321 Cardio-ViewCurlview Igt, LLC · 2008
  8. Descendant · 3 generations on · 7 of its ownK162012 CARESCAPE Central Station V2GE Healthcare · 2016 · Recalled

Showing 5 of 22 descendants. See all descendants

Descendants cleared per year

2007: 22008: 12010: 12011: 32014: 22016: 32018: 22020: 32022: 12024: 22025: 2
2007high 3 (2011) · low 0 (2009)2026

Most influential descendants

Showing 5 of 22 · See all
  • K073372Electrophysiology Systems Integrator (Epsi)Systemsone · 2007 · cites it directly10descendants
  • K062976Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies · 2007 · cites it directly10descendants
  • K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · 2 generations on9descendants
  • K083321Cardio-ViewCurlview Igt, LLC · 2008 · 2 generations on9descendants
  • K162012CARESCAPE Central Station V2GE Healthcare · 2016 · 3 generations on7descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K001112GE Marquette Clinical Information Center, GE Marquette CicGeneral Electric Medical Systems Information Techn2000—
−1K032370Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies20035
0 (this device)K053356Modification to: Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies2006—
+1K073372Electrophysiology Systems Integrator (Epsi)Systemsone2007—
+1K062976Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies20074
+2K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc.20141
+2K083321Cardio-ViewCurlview Igt, LLC2008—
+3K162012CARESCAPE Central Station V2GE Healthcare20166

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200720
200810
201010
201130
201420
201630
201820
202030
202210
202420
202520
FIG. 1 — Lineage of K053356, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXJ device

  1. #1 K261469 Nexxis 2;Nexxis OR 2;NMS 2;MNA-240;MNA-4x0…score 12.7 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032370Cic Pro Clinical Information Center Central Station
Summary page 1
…DXJ 21 CFR 870.1025 Detector and Alarm, Arrhythmia DSI Predicate Device: K032370 Clinical Information Center (CIC) Central Station Device Description: The CIC Pro Central Station is based on a st…
Ge Medical Systems Information Technologies20035

What descends from it? (22)

See all 22

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K073372Electrophysiology Systems Integrator (Epsi)Systemsone200710
K062976Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies200710

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 12, 2019Class IIZ-0726-2020Software in the Use EnvironmentWhen connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information…
Feb 27, 2009Not classifiedZ-1232-2009Software designGE Healthcare is aware of potential network time issues associated with the time change aspects of certain CARESCAPE CIC Pro" software versions. These issues may impact patient safety. No reported patient injuries or illnesses for the use…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 3.2 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K073372 (Systemsone, 2007) has gone furthest, with 10 descendants of its own.

How does its category compare?

All DXJ devices

DXJ Display, Cathode-Ray Tube, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DXJ

2in 2021

2007: 12008: 12010: 22011: 42012: 12013: 32014: 42015: 22016: 12017: 42018: 42020: 12021: 22026 (partial year): 1
2007high 4 (2011) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXJ

130in 2018

2011: 53 days median2013: 174 days median2014: 242 days median2017: 164 days median2018: 130 days median
2007high 242 (2014) · low 53 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXJ

2in 2025

2008: 22009: 82010: 12011: 12022: 12025: 2
2007high 8 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXJ

3in 2025

2007: 32008: 52009: 292010: 1482011: 1292012: 212013: 42014: 132015: 152016: 72017: 102018: 32019: 292020: 222021: 62022: 22023: 172024: 152025: 3
2007high 148 (2010) · low 2 (2022)2025

MAUDE reports naming product code DXJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20081—25
20090—829
20102—1148
20114531129
20121—021
2013317404
20144242013
20152—015
20161—07
20174164010
2018413003
20190—029
20201—022
20212—06
20220—12
20230—017
20240—015
20250—23
20261—0—

What is it indicated for?

The CIC Pro Clinical Information Center Central Station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. CIC Pro Clinical Information Center Central Station is intended to collect information from a network and display this data. This data includes…

Read the full statement (174 words)

The CIC Pro Clinical Information Center Central Station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment.

CIC Pro Clinical Information Center Central Station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and / or other non-medical information.

Physiological parameters and waveforms trom monitors aund telenrtu-y sy'stems can be displa cd and printed from the CIG Pro Clinical Information Center Central Station. Beat to beat patienm information for all parameters and waveforms from the hedside and tlemetry ssteins caln be displayed.

The CIC1iro Clinical Information CetCer CCenral Station suIpporIts ttr ability to access infornmation from ('K' Pro Clinical Information Center ( cntral Station' products in a web browser format. Additionalhl ('IC Pro Clinica. Informlation Centel ('c<ita! 1tation supporns the ability to access patient infornmation collected Irma the [ oily netork anicl stored on I nct:'ork SCI'

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053356: Modification to: Cic Pro Clinical Information Center Central Station.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K053356

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026