Recalled · Class II · 2019Cic Pro Clinical Information Center Central Station
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Medical Systems Information Technologies · DSI Detector And Alarm, Arrhythmia
Cleared on Aug 13, 2003 as substantially equivalent to K001112 (General Electric Medical Systems Information Techn, 2000) after a 12-day review. Its family tree reaches back 1 generation to 2000. 23 later devices descend from it. It has been recalled 5 times.
- Decided
- Aug 13, 2003
- Review
- 12 days
- Class
- Class II
- Product code
- DSI
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 1 generation to K001112 (2000). 23 later devices descend from it, 1 citing it directly.
- Root · 1 generation backK001112 GE Marquette Clinical Information Center, GE Marquette CicGeneral Electric Medical Systems Information Techn · 2000
- This deviceK032370 Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2003
- Descendant · cites it directly · 22 of its ownK053356 Modification to: Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2006 · Recalled
- Descendant · 2 generations on · 10 of its ownK073372 Electrophysiology Systems Integrator (Epsi)Systemsone · 2007
- Descendant · 2 generations on · 10 of its ownK062976 Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies · 2007 · Recalled
- Descendant · 3 generations on · 9 of its ownK133882 Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · Recalled
- Descendant · 3 generations on · 9 of its ownK083321 Cardio-ViewCurlview Igt, LLC · 2008
Showing 5 of 23 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 23 · See all- K053356Modification to: Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2006 · cites it directly22descendants
- K073372Electrophysiology Systems Integrator (Epsi)Systemsone · 2007 · 2 generations on10descendants
- K062976Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies · 2007 · 2 generations on10descendants
- K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · 3 generations on9descendants
- K083321Cardio-ViewCurlview Igt, LLC · 2008 · 3 generations on9descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K001112 | GE Marquette Clinical Information Center, GE Marquette Cic | General Electric Medical Systems Information Techn | 2000 | — |
| 0 (this device) | K032370 | Cic Pro Clinical Information Center Central Station | Ge Medical Systems Information Technologies | 2003 | — |
| +1 | K053356 | Modification to: Cic Pro Clinical Information Center Central Station | Ge Medical Systems Information Technologies | 2006 | 2 |
| +2 | K073372 | Electrophysiology Systems Integrator (Epsi) | Systemsone | 2007 | — |
| +2 | K062976 | Modification to Cic Pro Clinical Information Center | Ge Medical Systems Information Technologies | 2007 | 4 |
| +3 | K133882 | Carescape Central Station (Formally Known as Cic Pro) | Ge Medical Systems Information Technologies, Inc. | 2014 | 1 |
| +3 | K083321 | Cardio-View | Curlview Igt, LLC | 2008 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
| 2007 | 2 | 0 |
| 2008 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 3 | 0 |
| 2014 | 2 | 0 |
| 2016 | 3 | 0 |
| 2018 | 2 | 0 |
| 2020 | 3 | 0 |
| 2022 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DSI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DSI device
- #1 K243349 BodyGuardian Remote Monitoring System (BGRMS…score 13.1 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K001112 | GE Marquette Clinical Information Center, GE Marquette Cic | 2000 |
What descends from it? (23)
See all 231 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K053356 | Modification to: Cic Pro Clinical Information Center Central Station | 2006 | 22 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 12, 2019 | Class II | Z-0726-2020 | Software in the Use Environment | When connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information… |
| Feb 27, 2009 | Not classified | Z-1232-2009 | Software design | GE Healthcare is aware of potential network time issues associated with the time change aspects of certain CARESCAPE CIC Pro" software versions. These issues may impact patient safety. No reported patient injuries or illnesses for the use… |
| Jan 24, 2007 | Not classified | Z-0280-2008 | Software design | Two separate issues may occur with operating The CiC Pro v4.1 System with Telemetry. 1) Experience a loss of audible alarms when operating system memory resource reaches approximately 170MG. This will occur if CIC Pro continuously runs… |
| Jan 24, 2007 | Not classified | Z-0891-2007 | Other | Alarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system… |
| Oct 11, 2004 | Not classified | Z-0030-05 | Other | The following two conditions may occur with the printing of full disclosure strips and reports, printing of the alarm history events stored on a GE Unity Network Patient Data Server (PDS), and printing of caliper reports in the GE CIC Pro… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Its one ancestor has no recall.
- Of its descendants, K053356 (Ge Medical Systems Information Technologies, 2006) has gone furthest, with 22 descendants of its own.
How does its category compare?
All DSI devicesDSI Detector And Alarm, Arrhythmia: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DSI
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DSI
243in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DSI
18in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DSI
712in 2025
MAUDE reports naming product code DSI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 151 | 4 | 64 |
| 2008 | 5 | 70 | 8 | 156 |
| 2009 | 6 | 73 | 3 | 279 |
| 2010 | 6 | 111 | 5 | 823 |
| 2011 | 6 | 114 | 3 | 793 |
| 2012 | 3 | 113 | 1 | 766 |
| 2013 | 1 | — | 0 | 795 |
| 2014 | 4 | 200 | 0 | 1,000 |
| 2015 | 9 | 127 | 0 | 1,645 |
| 2016 | 6 | 224 | 0 | 4,352 |
| 2017 | 0 | — | 0 | 3,594 |
| 2018 | 2 | — | 1 | 3,648 |
| 2019 | 3 | 243 | 4 | 3,485 |
| 2020 | 2 | — | 2 | 3,084 |
| 2021 | 0 | — | 0 | 3,092 |
| 2022 | 0 | — | 1 | 1,638 |
| 2023 | 2 | — | 2 | 1,616 |
| 2024 | 1 | — | 0 | 1,320 |
| 2025 | 1 | — | 18 | 712 |
| 2026 | 1 | — | 1 | — |
What is it indicated for?
The CIC Pro Clinical Information Center Central Station is intended to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. The CIC Pro Central Station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and / or other non-medical information. Physiological parameters and waveforms from GE Medical…
Read the full statement (149 words)
The CIC Pro Clinical Information Center Central Station is intended to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment.
The CIC Pro Central Station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and / or other non-medical information.
Physiological parameters and waveforms from GE Medical Systems Information Technologies monitors and telemetry systems can be displayed and printed from the CIC Pro Central Station. Beat to beat patient information for all parameters and waveforms from the bedside and telemetry systems can be displayed.
The CIC Pro Central Station supports the ability to access information from GE Medical Systems Information Technologies’ products in a web browser format. Additionally, CIC Pro Central Station supports the ability to access patient information collected from the Unity network and stored on a network server
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K032370 · printed Nov 6, 2026
Cite this page
510(k) Report. “K032370: Cic Pro Clinical Information Center Central Station.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K032370
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026