Modification to Cic Pro Clinical Information Center

K062976510(k) SpecialCleared Mar 16, 2007168-day reviewClass II21 CFR 870.1025Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems Information Technologies · DSI Detector And Alarm, Arrhythmia

Cleared on Mar 16, 2007 as substantially equivalent to K053356 (Ge Medical Systems Information Technologies, 2006) after a 168-day review. Its family tree reaches back 3 generations to 2000. 10 later devices descend from it. It has been recalled 4 times.

Decided
Mar 16, 2007
Review
168 days
Class
Class II
Product code
DSI
Recalls
4
Predicates
1

Its longest line of descent runs back 3 generations to K001112 (2000). 10 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2014: 1 descendant cleared2016: 2 descendants cleared2020: 1 descendant cleared2018: 1 descendant cleared2025: 2 descendants cleared2024: 2 descendants cleared2022: 1 descendant clearedK001112 · GE Marquette Clinical Information Center, GE Marquette Cic · General Electric Medical Systems Information Techn · 2000K032370 · Cic Pro Clinical Information Center Central Station · Ge Medical Systems Information Technologies · 2003 · recalledK053356 · Modification to: Cic Pro Clinical Information Center Central Station · Ge Medical Systems Information Technologies · 2006 · recalled200020052010201520202025
  1. Root · 3 generations backK001112 GE Marquette Clinical Information Center, GE Marquette CicGeneral Electric Medical Systems Information Techn · 2000
  2. 2 generations backK032370 Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2003 · Recalled
  3. 1 generation backK053356 Modification to: Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies · 2006 · Recalled
  4. This deviceK062976 Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies · 2007
  5. Descendant · cites it directly · 9 of its ownK133882 Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · Recalled
  6. Descendant · 2 generations on · 7 of its ownK162012 CARESCAPE Central Station V2GE Healthcare · 2016 · Recalled
  7. Descendant · 3 generations on · 3 of its ownK201941 Monitor B125, Monitor B105Ge Medical Systems Information Technologies, Inc. · 2020
  8. Descendant · 4 generations on · 2 of its ownK213490 Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor…Ge Medical Systems Information Technologies, Inc. · 2022 · Recalled
  9. Descendant · 3 generations on · 2 of its ownK173107 Vios Central Station Monitor Software, Vios Central Server SoftwareVios Medical, Inc. · 2018

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2014: 12016: 22018: 12020: 12022: 12024: 22025: 2
2014high 2 (2016) · low 0 (2015)2026

Most influential descendants

Showing 5 of 10 · See all
  • K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc. · 2014 · cites it directly9descendants
  • K162012CARESCAPE Central Station V2GE Healthcare · 2016 · 2 generations on7descendants
  • K201941Monitor B125, Monitor B105Ge Medical Systems Information Technologies, Inc. · 2020 · 3 generations on3descendants
  • K213490Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor B125PGe Medical Systems Information Technologies, Inc. · 2022 · 4 generations on2descendants
  • K173107Vios Central Station Monitor Software, Vios Central Server SoftwareVios Medical, Inc. · 2018 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K001112GE Marquette Clinical Information Center, GE Marquette CicGeneral Electric Medical Systems Information Techn2000—
−2K032370Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies20035
−1K053356Modification to: Cic Pro Clinical Information Center Central StationGe Medical Systems Information Technologies20062
0 (this device)K062976Modification to Cic Pro Clinical Information CenterGe Medical Systems Information Technologies2007—
+1K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc.20141
+2K162012CARESCAPE Central Station V2GE Healthcare20166
+3K201941Monitor B125, Monitor B105Ge Medical Systems Information Technologies, Inc.2020—
+4K213490Monitor B105M, Monitor B125M, Monitor B155M, Monitor B105P, Monitor B125PGe Medical Systems Information Technologies, Inc.20226
+3K173107Vios Central Station Monitor Software, Vios Central Server SoftwareVios Medical, Inc.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201620
201810
202010
202210
202420
202520
FIG. 1 — Lineage of K062976, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSI device

  1. #1 K243349 BodyGuardian Remote Monitoring System (BGRMS…score 13.1 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053356Modification to: Cic Pro Clinical Information Center Central Station
Summary page 1
…CIC Pro Clinical Information Center is classified as: 21 CFR 870.2450 Display, Cathode-ray Tube, Medical Predicate Device: K053356:CIC Pro Clinical Information Center Device Description: The CIC Pro Clinical Information Center is based on a standar…
Ge Medical Systems Information Technologies20062

Also cleared against K053356: 1 other device →

What descends from it? (10)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133882Carescape Central Station (Formally Known as Cic Pro)Ge Medical Systems Information Technologies, Inc.20149

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 12, 2019Class IIZ-0726-2020Software in the Use EnvironmentWhen connected to the Mission Critical (MC) and /or Information Exchange (IX) networks, certain versions of the CARESCAPE Telemetry Server, ApexPro Telemetry Server, CARESCAPE Central Station (CSCS) version 1 and Central Information…
Mar 10, 2009Not classifiedZ-1600-2009Component change controlGE Healthcare has become aware of a potential issue associated with the failure to restart CARESCAPE CIC Pro v5.1 on the MP100 hardware. This issue may impact patient safety.
Feb 27, 2009Not classifiedZ-1232-2009Software designGE Healthcare is aware of potential network time issues associated with the time change aspects of certain CARESCAPE CIC Pro" software versions. These issues may impact patient safety. No reported patient injuries or illnesses for the use…
Nov 8, 2007Not classifiedZ-0519-2008Software designPotential for Loss of Audible Alarming and ECG Tab Setting Changes; from Clinical Informaiton Center (CIC) Pro v5.0.3 & 5.0.6 when used in combination with Telemetry and bedside monitors.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 2.7 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K133882 (Ge Medical Systems Information Technologies, Inc., 2014) has gone furthest, with 9 descendants of its own.
  • 1 other device also cites its predicate K053356.

How does its category compare?

All DSI devices

DSI Detector And Alarm, Arrhythmia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · DSI

1in 2025

2007: 72008: 52009: 62010: 62011: 62012: 32013: 12014: 42015: 92016: 62018: 22019: 32020: 22023: 22024: 12025: 12026 (partial year): 1
2007high 9 (2015) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSI

243in 2019

2007: 151 days median2008: 70 days median2009: 73 days median2010: 111 days median2011: 114 days median2012: 113 days median2014: 200 days median2015: 127 days median2016: 224 days median2019: 243 days median
2007high 243 (2019) · low 70 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSI

18in 2025

2007: 42008: 82009: 32010: 52011: 32012: 12018: 12019: 42020: 22022: 12023: 22025: 182026 (partial year): 1
2007high 18 (2025) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DSI

712in 2025

2007: 642008: 1562009: 2792010: 8232011: 7932012: 7662013: 7952014: 1,0002015: 1,6452016: 4,3522017: 3,5942018: 3,6482019: 3,4852020: 3,0842021: 3,0922022: 1,6382023: 1,6162024: 1,3202025: 712
2007high 4,352 (2016) · low 64 (2007)2025

MAUDE reports naming product code DSI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077151464
20085708156
20096733279
201061115823
201161143793
201231131766
20131—0795
2014420001,000
2015912701,645
2016622404,352
20170—03,594
20182—13,648
2019324343,485
20202—23,084
20210—03,092
20220—11,638
20232—21,616
20241—01,320
20251—18712
20261—1—

What is it indicated for?

The CIC Pro Clinical Information Center central station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment. CIC Pro Clinical Information Center central station is intended to collect information from a network and display this data. This data includes…

Read the full statement (172 words)

The CIC Pro Clinical Information Center central station is intended for use under the direct supervision of a licensed healthcare practitioner. The intended use is to provide clinicians with adult, pediatric and neonatal patient data in a centralized location within a hospital or clinical environment.

CIC Pro Clinical Information Center central station is intended to collect information from a network and display this data. This data includes physiological, patient demographic and/or other non-medical information. Physiological parameters and waveforms from monitors and telemetry systems can be displayed and printed from the CIC Pro Clinical Information Center central station. Beat to beat patient information for all parameters and waveforms from the bedside and telemetry systems can be displayed.

The CIC Pro Clinical Information Center central station supports the ability to access information from the CIC Pro Clinical Information Center central stations' products in a web browser format. Additionally, the CIC Pro Clinical Information Center central station supports the ability to access patient information collected from the unity network and stored on a network server.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062976: Modification to Cic Pro Clinical Information Center.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K062976

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026