510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

PF4 Enhanced Solid Phase Elisa

K053559510(k) SpecialCleared Jan 20, 200630-day reviewClass II21 CFR 864.7695Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Genetic Testing Institute · LCO Platelet Factor 4 Radioimmunoassay

Cleared on Jan 20, 2006 as substantially equivalent to K983379 (Genetic Testing Institute, 1999) after a 30-day review. Its family tree reaches back 1 generation to 1999. 1 of the 8 devices in its family, counting itself, has a Class I recall. 6 later devices descend from it. It has been recalled 2 times.

Decided
Jan 20, 2006
Review
30 days
Class
Class II
Product code
LCO
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K983379 (1999). 6 later devices descend from it, 3 citing it directly. 1 of the 8 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2007: 2 descendants cleared2020: 2 descendants cleared2017: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant clearedK983379 · Gti-PF4 Elisa · Genetic Testing Institute · 19991995200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 1 descendant K983379 · Gti-PF4 Elisa · Genetic Testing Institute · 19990
  1. Root · 1 generation backK983379 Gti-PF4 ElisaGenetic Testing Institute · 1999
  2. This deviceK053559 PF4 Enhanced Solid Phase ElisaGenetic Testing Institute · 2006
  3. Descendant · cites it directly · 2 of its ownK071255 Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040AHyphen Biomed · 2007 · Recalled
  4. Descendant · 2 generations on · 1 of its ownK170854 HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO · 2017 · Recalled (Class I)
  5. Descendant · cites it directly · 1 of its ownK071781 PF4 IGGGenetic Testing Institute · 2007
  6. Descendant · 3 generations on · 0 of its ownK243374 HemosIL CL HIT-IgG (PF4-H)Instrumentation Laboratory (IL) Co. · 2025
  7. Descendant · cites it directly · 0 of its ownK201570 PF4 Enhanced assayImmucor Gti Diagnostics, Inc. · 2020

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2007: 22017: 12020: 22025: 1
2007high 2 (2007) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K071255Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040AHyphen Biomed · 2007 · cites it directly2descendants
  • K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO · 2017 · 2 generations on1descendant
  • K071781PF4 IGGGenetic Testing Institute · 2007 · cites it directly1descendant
  • K243374HemosIL CL HIT-IgG (PF4-H)Instrumentation Laboratory (IL) Co. · 2025 · 3 generations on0descendants
  • K201570PF4 Enhanced assayImmucor Gti Diagnostics, Inc. · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K983379Gti-PF4 ElisaGenetic Testing Institute1999—
0 (this device)K053559PF4 Enhanced Solid Phase ElisaGenetic Testing Institute2006—
+1K071255Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040AHyphen Biomed20078
+2K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO2017Class I
+1K071781PF4 IGGGenetic Testing Institute2007—
+3K243374HemosIL CL HIT-IgG (PF4-H)Instrumentation Laboratory (IL) Co.2025—
+1K201570PF4 Enhanced assayImmucor Gti Diagnostics, Inc.2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200720
201711
202020
202510
FIG. 1 — Lineage of K053559, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCO device

  1. #1 K243374 HemosIL CL HIT-IgG (PF4-H)score 9.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K983379Gti-PF4 Elisa
Summary page 1
…ct Code: 0545 Ill. Name of predicate device for claiming equivalence GTI-PF4 ELISA (K983379) IV. Description of Device: PF4 ENHANCED® S…
Genetic Testing Institute1999—

Also cleared against K983379: 2 other devices →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071255Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040AHyphen Biomed20072
K071781PF4 IGGGenetic Testing Institute20071
K201570PF4 Enhanced assayImmucor Gti Diagnostics, Inc.20200

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 15, 2014Class IIZ-1001-2015Employee errorSome Lots of PF4 Concentrated Wash (10X) are a slightly more dilute concentration. This may lead to slightly increased Optical Density (OD) values. Samples with a higher background reactivity may demonstrate false positive result…
Feb 4, 2013Class IIZ-1022-2013Process controlLifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K071255 (Hyphen Biomed, 2007) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K983379.

How does its category compare?

All LCO devices

LCO Platelet Factor 4 Radioimmunoassay: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LCO

1in 2025

2007: 22016: 12017: 12020: 22025: 1
2007high 2 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCO

Too few LCO 510(k)s in any year since 2007 to give a median.

Recalls per year · LCO

3so far in 2026, the most since 2007

2011: 22013: 12014: 12026 (partial year): 3, 2 Class I
2007high 3 (2026) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LCO

11in 2025

2016: 22019: 42020: 12024: 22025: 11
2007high 11 (2025) · low 0 (2007)2025

MAUDE reports naming product code LCO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20080—00
20090—00
20100—00
20110—20
20120—00
20130—10
20140—10
20150—00
20161—02
20171—00
20180—00
20190—04
20202—01
20210—00
20220—00
20230—00
20240—02
20251—011
20260—3 (2)—

What is it indicated for?

GTI PF4 ENHANCED® is designed as a solid phase enzyme-linked immunosorbent assay (ELISA). The product is intended to be used as an in vitro diagnostics kit by hematology, coagulation or other pathology laboratories to assist in screening patient samples for the presence of heparin-associated antibodies commonly found in patients with heparin induced thrombocytopenia or thrombosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053559: PF4 Enhanced Solid Phase Elisa.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K053559

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026