Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040A

K071255510(k) TraditionalCleared Sep 7, 2007126-day reviewClass II21 CFR 864.7695Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hyphen Biomed · LCO Platelet Factor 4 Radioimmunoassay

Cleared on Sep 7, 2007 as substantially equivalent to K003767 (Diagnostica Stago, Inc., 2001) and 1 other predicate after a 126-day review. Its family tree reaches back 2 generations to 1999. 1 of the 6 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 8 times.

Decided
Sep 7, 2007
Review
126 days
Class
Class II
Product code
LCO
Recalls
8
Predicates
2

Its longest line of descent runs back 2 generations to K983379 (1999). 3 ancestors in all. 2 later devices descend from it, 1 citing it directly. 1 of the 6 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2017: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant clearedK983379 · Gti-PF4 Elisa · Genetic Testing Institute · 1999K053559 · PF4 Enhanced Solid Phase Elisa · Genetic Testing Institute · 2006 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 1 descendant K053559 · PF4 Enhanced Solid Phase Elisa · Genetic Testing Institute · 2006 · recalledK983379 · Gti-PF4 Elisa · Genetic Testing Institute · 19990
  1. Root · 2 generations backK983379 Gti-PF4 ElisaGenetic Testing Institute · 1999
  2. 1 generation backK053559 PF4 Enhanced Solid Phase ElisaGenetic Testing Institute · 2006 · Recalled
  3. This deviceK071255 Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040AHyphen Biomed · 2007
  4. Descendant · cites it directly · 1 of its ownK170854 HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO · 2017 · Recalled (Class I)
  5. Descendant · 2 generations on · 0 of its ownK243374 HemosIL CL HIT-IgG (PF4-H)Instrumentation Laboratory (IL) Co. · 2025

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO · 2017 · cites it directly1descendant
  • K243374HemosIL CL HIT-IgG (PF4-H)Instrumentation Laboratory (IL) Co. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K983379Gti-PF4 ElisaGenetic Testing Institute1999—
−1K053559PF4 Enhanced Solid Phase ElisaGenetic Testing Institute20062
0 (this device)K071255Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040AHyphen Biomed2007—
+1K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO2017Class I
+2K243374HemosIL CL HIT-IgG (PF4-H)Instrumentation Laboratory (IL) Co.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201711
202510
FIG. 1 — Lineage of K071255, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCO device

  1. #1 K243374 HemosIL CL HIT-IgG (PF4-H)score 9.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K003767Asserachrom Hpia Test Kit
Summary page 1
…t Kit (Stago) 2. PF4 ENHANCED Solid Phase ELISA (GTI) Predicate device 510 (k) number: 1. K003767 2. K053559 Description of the Device: The ZYMUTEST HIA is solid phase enzyme…
Diagnostica Stago, Inc.20011
*K053559PF4 Enhanced Solid Phase Elisa
Summary page 1
…2. PF4 ENHANCED Solid Phase ELISA (GTI) Predicate device 510 (k) number: 1. K003767 2. K053559 Description of the Device: The ZYMUTEST HIA is solid phase enzyme linked immunosorbent assay (ELISA) des…
Genetic Testing Institute20062

Also cleared against K003767: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT Controls Class IInstrumentation Laboratory CO20171

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 1, 2026Class IIZ-2171-2026Labeling Change ControlKits from lot FD1265 had been packaged on 17 October 2024 and put on the market with an outdated flyer (8 October 2024), containing incorrect values for the positive control.
Jul 19, 2019Class IIZ-2347-2019Under Investigation by firmResidual crystallization of the microplate and recurrent negative controls out of range.
Jul 19, 2019Class IIZ-2346-2019Under Investigation by firmResidual crystallization of the microplate and recurrent negative controls out of range.
Feb 5, 2018Class IIZ-1743-2018Packaging process controlIncorrect package insert indicating use of the kits with plasma, serum, and other biological fluids. The kits should only be used with plasma.
Feb 5, 2018Class IIZ-1744-2018Packaging process controlIncorrect package insert indicating use of the kits with plasma, serum, and other biological fluids. The kits should only be used with plasma.
Feb 5, 2018Class IIZ-1742-2018Packaging process controlIncorrect package insert indicating use of the kits with plasma, serum, and other biological fluids. The kits should only be used with plasma.
Nov 9, 2011Class IIZ-1062-2018OtherA voluntary recall is initiated for ZYMUTEST HIA lgGAM - Ref RK040D - Lot F1700415 and ZYMUTEST HIA MonoStrip lgGAM - Ref RK041D - Lot F1700416 because they may produce false positive results.
Nov 9, 2011Class IIZ-1063-2018OtherA voluntary recall is initiated for ZYMUTEST HIA lgGAM - Ref RK040D - Lot F1700415 and ZYMUTEST HIA MonoStrip lgGAM - Ref RK041D - Lot F1700416 because they may produce false positive results.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K170854HemosIL AcuStar HIT-IgG (PF4-H), HemosIL AcuStar HIT ControlsInstrumentation Laboratory CO · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 6.9 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K170854 (Instrumentation Laboratory CO, 2017) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K003767.

How does its category compare?

All LCO devices

LCO Platelet Factor 4 Radioimmunoassay: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · LCO

1in 2025

2007: 22016: 12017: 12020: 22025: 1
2007high 2 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCO

Too few LCO 510(k)s in any year since 2007 to give a median.

Recalls per year · LCO

3so far in 2026, the most since 2007

2011: 22013: 12014: 12026 (partial year): 3, 2 Class I
2007high 3 (2026) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LCO

11in 2025

2016: 22019: 42020: 12024: 22025: 11
2007high 11 (2025) · low 0 (2007)2025

MAUDE reports naming product code LCO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20080—00
20090—00
20100—00
20110—20
20120—00
20130—10
20140—10
20150—00
20161—02
20171—00
20180—00
20190—04
20202—01
20210—00
20220—00
20230—00
20240—02
20251—011
20260—3 (2)—

What is it indicated for?

ZYMUTEST HIA IgG and IgGAM kits are designed as a solid phase enzyme- linked immunosorbent assay (ELISA). These products are intended to be used as an in vitro diagnostics kit by Hematology, coagulation or other pathology laboratories to assist in screening patients samples for the presence of heparin- associated antibodies commonly found in patients with heparin induced thrombocytopenia or thrombosis (HIT).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071255: Zymutest Hia Iggam, Zymutest Igg, Model RK040D, RK040A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K071255

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.