Excelart Vantage ZGV

K060003510(k) TraditionalCleared Jan 18, 200615-day reviewClass II21 CFR 892.1000Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshibamedical Systems Corporation · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jan 18, 2006 as substantially equivalent to K032490 (Toshiba America Medical Systems, In.C, 2003) after a 15-day review. Its family tree reaches back 3 generations to 2000. 69 later devices descend from it. It has been recalled once.

Decided
Jan 18, 2006
Review
15 days
Class
Class II
Product code
LNH
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K002531 (2000). 69 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2011: 4 descendants cleared2006: 1 descendant cleared2008: 2 descendants cleared2007: 1 descendant cleared2012: 6 descendants cleared2018: 7 descendants cleared2015: 4 descendants cleared2014: 1 descendant cleared2016: 2 descendants cleared2019: 4 descendants cleared2013: 1 descendant cleared2017: 3 descendants cleared2020: 7 descendants cleared2021: 9 descendants cleared2022: 5 descendants cleared2023: 2 descendants cleared2024: 5 descendants cleared2025: 3 descendants cleared2026: 2 descendants clearedK002531 · Excelart with Pianissimo, Model MRT-1501/P3 · Toshiba America Mri, Inc. · 2000 · recalledK023511 · Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2 · Toshiba America Medical Systems, In.C · 2002 · recalledK032490 · Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and MRT-1503/P2 · Toshiba America Medical Systems, In.C · 2003 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 4 descendants4 generations forward: 5 descendants5 generations forward: 3 descendants6 generations forward: 3 descendants7 generations forward: 7 descendants8 generations forward: 5 descendants9 generations forward: 5 descendants10 generations forward: 5 descendants11 generations forward: 4 descendants12 generations forward: 6 descendants13 generations forward: 5 descendants14 generations forward: 3 descendants15 generations forward: 2 descendants16 generations forward: 2 descendants17 generations forward: 1 descendant18 generations forward: 1 descendant19 generations forward: 1 descendant K032490 · Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and MRT-1503/P2 · Toshiba America Medical Systems, In.C · 2003 · recalledK023511 · Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2 · Toshiba America Medical Systems, In.C · 2002 · recalledK002531 · Excelart with Pianissimo, Model MRT-1501/P3 · Toshiba America Mri, Inc. · 2000 · recalled0+5+10+15
  1. Root · 3 generations backK002531 Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc. · 2000 · Recalled
  2. 2 generations backK023511 Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C · 2002 · Recalled
  3. 1 generation backK032490 Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models…Toshiba America Medical Systems, In.C · 2003 · Recalled
  4. This deviceK060003 Excelart Vantage ZGVToshibamedical Systems Corporation · 2006
  5. Descendant · cites it directly · 68 of its ownK063361 Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C · 2006 · Recalled
  6. Descendant · 2 generations on · 61 of its ownK080038 Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C · 2008 · Recalled
  7. Descendant · 3 generations on · 56 of its ownK102489 Vantage Titan 3TToshiba Medical Systems Corporation, Japan · 2011 · Recalled
  8. Descendant · 4 generations on · 53 of its ownK113093 Vantage Titan 3TToshibamedical Systems Corporation · 2012 · Recalled
  9. Descendant · 5 generations on · 52 of its ownK120487 Vantage Titan 3TToshibamedical Systems Corporation · 2012

Showing 5 of 69 descendants. See all descendants

Descendants cleared per year

2006: 12007: 12008: 22011: 42012: 62013: 12014: 12015: 42016: 22017: 32018: 72019: 42020: 72021: 92022: 52023: 22024: 52025: 32026: 2
2006high 9 (2021) · low 0 (2009)2026

Most influential descendants

Showing 5 of 69 · See all
  • K063361Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C · 2006 · cites it directly68descendants
  • K080038Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C · 2008 · 2 generations on61descendants
  • K102489Vantage Titan 3TToshiba Medical Systems Corporation, Japan · 2011 · 3 generations on56descendants
  • K113093Vantage Titan 3TToshibamedical Systems Corporation · 2012 · 4 generations on53descendants
  • K120487Vantage Titan 3TToshibamedical Systems Corporation · 2012 · 5 generations on52descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K002531Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc.20004
−2K023511Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C20022
−1K032490Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and…Toshiba America Medical Systems, In.C20033
0 (this device)K060003Excelart Vantage ZGVToshibamedical Systems Corporation2006—
+1K063361Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C20061
+2K080038Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C20082
+3K102489Vantage Titan 3TToshiba Medical Systems Corporation, Japan20111
+4K113093Vantage Titan 3TToshibamedical Systems Corporation20121
+5K120487Vantage Titan 3TToshibamedical Systems Corporation2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
200820
201140
201260
201310
201410
201540
201620
201730
201870
201940
202070
202190
202250
202320
202450
202530
202620
FIG. 1 — Lineage of K060003, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032490Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and MRT-1503/P2
Summary page 2
…age TM XGV/AGV (No Vantage TM ZGV changes from the previous submission, K032490) a. Static field strength: 1.5 T Same b. Peak and A-weighted…
Toshiba America Medical Systems, In.C20033

Also cleared against K032490: 1 other device →

What descends from it? (69)

See all 69

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063361Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C200668
K1112514CH Flex SpeederToshiba Medical Systems Coporation, Japan20111

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 21, 2012Class IIZ-2315-2012Component change controlThe firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 2.0 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K063361 (Toshiba America Medical Systems, In.C, 2006) has gone furthest, with 68 descendants of its own.
  • 1 other device also cites its predicate K032490.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

Imaging of:

• The Whole Body (including head, abdomen, pelvis, limbs and extremities, spine, neck, TMJ, heart, blood vessels). [Application terms include MRCP (MR Cholangiopancreatography), MR Cisternography, MR Urography, MR Myelography, MR Fluoroscopy, SAS (Surface Anatomy Scan), Dynamic ScanCine Imaging and Cardiac tagging.]

• Fluid Visualization

• 2D/3D Imaging

• MR Angiography / MR Vascular Imaging

• Blood Oxygenation Level Dependent (BOLD) imaging

• Perfusion / Diffusion Imaging

• Proton Spectroscopy

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060003: Excelart Vantage ZGV.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K060003

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026