510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Vantage Titan 3T

K102489510(k) TraditionalCleared Aug 2, 2011336-day reviewClass II21 CFR 892.1000Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshiba Medical Systems Corporation, Japan · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Aug 2, 2011 as substantially equivalent to K080038 (Toshiba America Medical Systems, In.C, 2008) after a 336-day review. Its family tree reaches back 6 generations to 2000. 56 later devices descend from it. It has been recalled once.

Decided
Aug 2, 2011
Review
336 days
Class
Class II
Product code
LNH
Recalls
1
Predicates
1

Its longest line of descent runs back 6 generations to K002531 (2000). 56 later devices descend from it, 2 citing it directly.

Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 5 descendants5 generations forward: 4 descendants6 generations forward: 4 descendants7 generations forward: 3 descendants8 generations forward: 4 descendants9 generations forward: 7 descendants10 generations forward: 7 descendants11 generations forward: 5 descendants12 generations forward: 2 descendants13 generations forward: 2 descendants14 generations forward: 2 descendants15 generations forward: 2 descendants16 generations forward: 2 descendants17 generations forward: 1 descendant K080038 · Vantage Titan, Model MRT-1504/S3 · Toshiba America Medical Systems, In.C · 2008 · recalledK063361 · Excelart Vantage Atlas, Model MRT-1503 · Toshiba America Medical Systems, In.C · 2006 · recalledK060003 · Excelart Vantage ZGV · Toshibamedical Systems Corporation · 2006 · recalledK032490 · Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and MRT-1503/P2 · Toshiba America Medical Systems, In.C · 2003 · recalledK023511 · Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2 · Toshiba America Medical Systems, In.C · 2002 · recalledK002531 · Excelart with Pianissimo, Model MRT-1501/P3 · Toshiba America Mri, Inc. · 2000 · recalled−50+5+10+15
  1. Root · 6 generations backK002531 Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc. · 2000 · Recalled
  2. 5 generations backK023511 Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C · 2002 · Recalled
  3. 4 generations backK032490 Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models…Toshiba America Medical Systems, In.C · 2003 · Recalled
  4. 3 generations backK060003 Excelart Vantage ZGVToshibamedical Systems Corporation · 2006 · Recalled
  5. 2 generations backK063361 Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C · 2006 · Recalled
  6. 1 generation backK080038 Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C · 2008 · Recalled
  7. This deviceK102489 Vantage Titan 3TToshiba Medical Systems Corporation, Japan · 2011
  8. Descendant · cites it directly · 53 of its ownK113093 Vantage Titan 3TToshibamedical Systems Corporation · 2012 · Recalled
  9. Descendant · 2 generations on · 52 of its ownK120487 Vantage Titan 3TToshibamedical Systems Corporation · 2012
  10. Descendant · 3 generations on · 37 of its ownK132160 Vantage Titan 3TToshibamedical Systems Corporation · 2013
  11. Descendant · 4 generations on · 36 of its ownK143008 Vantage Titan 3TToshibamedical Systems Corporation · 2015
  12. Descendant · 5 generations on · 32 of its ownK152371 Vantage Titan 3TToshibamedical Systems Corporation · 2015 · Recalled

Showing 5 of 56 descendants. See all descendants

Descendants cleared per year

2012: 22013: 12015: 42016: 22017: 32018: 72019: 42020: 72021: 92022: 52023: 22024: 52025: 32026: 2
2012high 9 (2021) · low 0 (2014)2026

Most influential descendants

Showing 5 of 56 · See all
  • K113093Vantage Titan 3TToshibamedical Systems Corporation · 2012 · cites it directly53descendants
  • K120487Vantage Titan 3TToshibamedical Systems Corporation · 2012 · 2 generations on52descendants
  • K132160Vantage Titan 3TToshibamedical Systems Corporation · 2013 · 3 generations on37descendants
  • K143008Vantage Titan 3TToshibamedical Systems Corporation · 2015 · 4 generations on36descendants
  • K152371Vantage Titan 3TToshibamedical Systems Corporation · 2015 · 5 generations on32descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K002531Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc.20004
−5K023511Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C20022
−4K032490Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and…Toshiba America Medical Systems, In.C20033
−3K060003Excelart Vantage ZGVToshibamedical Systems Corporation20061
−2K063361Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C20061
−1K080038Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C20082
0 (this device)K102489Vantage Titan 3TToshiba Medical Systems Corporation, Japan2011—
+1K113093Vantage Titan 3TToshibamedical Systems Corporation20121
+2K120487Vantage Titan 3TToshibamedical Systems Corporation2012—
+3K132160Vantage Titan 3TToshibamedical Systems Corporation2013—
+4K143008Vantage Titan 3TToshibamedical Systems Corporation2015—
+5K152371Vantage Titan 3TToshibamedical Systems Corporation20151

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201220
201310
201540
201620
201730
201870
201940
202070
202190
202250
202320
202450
202530
202620
FIG. 1 — Lineage of K102489, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080038Vantage Titan, Model MRT-1504/S3
Summary page 1
…the maximum field of view of 50 x 50'x 45 cm. The Vantage Titan 3T MRI System is comparable to the current 1.ST EXCELART Vantage Titan MR] System (K080038), cleared January 22, 2008 with the following modifications. * Field strength of magnet has been changed to 3T.…
Toshiba America Medical Systems, In.C20082
†K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I

Also cleared against K080038: 2 other devices →

What descends from it? (56)

See all 56

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113093Vantage Titan 3TToshibamedical Systems Corporation201253
K18169716ch Foot/ Ankle SPEEDERQuality Electrodynamics, LLC20181

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 28, 2013Class IIZ-0201-2014Nonconforming Material/ComponentToshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR) values when using the Toshiba Titan 3T System. When the system is used in combination with either the QD Head Coil…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 4 companies; a new generation was cleared every 2.0 years on median.
  • 6 of the 6 ancestors have a recall of any class.
  • Of its descendants, K113093 (Toshibamedical Systems Corporation, 2012) has gone furthest, with 53 descendants of its own.
  • 2 other devices also cite its predicate K080038.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

Vantage Titan ST systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The…

Read the full statement (122 words)

Vantage Titan ST systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

* Proton density (PD) (also called hydrogen density)

* Spin-lattice relaxation time (Ti)

- Spin-spin relaxation time (T2)

* Flow dynamics

* Chemical Shift

Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102489: Vantage Titan 3T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K102489

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026