CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor

K060065510(k) TraditionalCleared May 4, 2006115-day reviewClass II21 CFR 870.2700Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Oridion Capnography, Inc. · DQA Oximeter

Cleared on May 4, 2006 as substantially equivalent to K040011 (Oridion, Ltd., 2004) and 5 other predicates after a 115-day review. Its family tree reaches back 8 generations to 1997. 2 of the 285 devices in its family, counting itself, have a Class I recall. 250 later devices descend from it. It has been recalled 3 times.

Decided
May 4, 2006
Review
115 days
Class
Class II
Product code
DQA
Recalls
3
Predicates
6

Its longest line of descent runs back 8 generations to K962603 (1997). 34 ancestors in all, the oldest K893221 (1990). 250 later devices descend from it, 7 citing it directly. 2 of the 285 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2004: 1 other ancestor2002: 1 other ancestor2003: 1 other ancestor2000: 1 other ancestor1999: 3 other ancestors1995: 4 other ancestors1990: 2 other ancestors1998: 4 other ancestors1997: 2 other ancestors, 1 with a Class I recall11996: 2 other ancestors1991: 2 other ancestors1993: 1 other ancestor1992: 1 other ancestor2009: 7 descendants cleared2016: 10 descendants cleared2010: 8 descendants cleared2007: 1 descendant cleared2008: 1 descendant cleared2012: 13 descendants cleared2011: 6 descendants cleared2013: 14 descendants cleared2018: 15 descendants cleared2015: 11 descendants cleared2014: 17 descendants cleared, 1 with a Class I recall12022: 24 descendants cleared2017: 12 descendants cleared2019: 16 descendants cleared2020: 15 descendants cleared2021: 11 descendants cleared2023: 16 descendants cleared2024: 20 descendants cleared2026: 15 descendants cleared2025: 18 descendants clearedK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999×2: 2 generations cleared close together, K012891 (2002), K021090 (2002) K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 20051990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 9 other ancestors3 generations back: 7 other ancestors, 1 with a Class I recall14 generations back: 5 other ancestors5 generations back: 2 other ancestors 1 generation forward: 7 descendants2 generations forward: 15 descendants3 generations forward: 26 descendants4 generations forward: 38 descendants5 generations forward: 32 descendants6 generations forward: 36 descendants, 1 with a Class I recall17 generations forward: 26 descendants8 generations forward: 18 descendants9 generations forward: 20 descendants10 generations forward: 26 descendants11 generations forward: 6 descendants K050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK021090 · Oximax N-550 Pulse Oximeter,Model N-550 Pulse Oximeter · Nellcor Puritan Bennett, Inc. · 2002K012891 · Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables (Aka Accessories) · Nellcor Puritan Bennett, Inc. · 2002 · recalledK991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−50+5+10
  1. Root · 8 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 7 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 6 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 5 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 4 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 3 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
  7. 2 generations backK033048 Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS…Cas Medical Systems, Inc. · 2004 · Recalled
  8. 1 generation backK050844 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL…Cas Medical Systems, Inc. · 2005
  9. This deviceK060065 CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One…Oridion Capnography, Inc. · 2006
  10. Descendant · cites it directly · 197 of its ownK072295 Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc. · 2007
  11. Descendant · 2 generations on · 170 of its ownK082268 CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc. · 2009 · Recalled
  12. Descendant · 3 generations on · 155 of its ownK091241 Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse…Masimo Corporation · 2009
  13. Descendant · 4 generations on · 115 of its ownK100428 Masimo Rainbow Set Radical 7R Pulse CO-OximeterMasimo Corporation · 2010
  14. Descendant · 3 generations on · 93 of its ownK094012 CAPNOSTREAM20 with MICROMEDICO2 ModuleOridion Capnography, Inc. · 2010

Showing 5 of 250 descendants. See all descendants

Descendants cleared per year

2007: 12008: 12009: 72010: 82011: 62012: 132013: 142014: 172015: 112016: 102017: 122018: 152019: 162020: 152021: 112022: 242023: 162024: 202025: 182026: 15
2007high 24 (2022) · low 1 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 250 · See all
  • K072295Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc. · 2007 · cites it directly197descendants
  • K082268CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc. · 2009 · 2 generations on170descendants
  • K091241Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse CO-OximeterMasimo Corporation · 2009 · 3 generations on155descendants
  • K100428Masimo Rainbow Set Radical 7R Pulse CO-OximeterMasimo Corporation · 2010 · 4 generations on115descendants
  • K094012CAPNOSTREAM20 with MICROMEDICO2 ModuleOridion Capnography, Inc. · 2010 · 3 generations on93descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−7K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−6K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−5K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−4K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−3K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc.2002—
−2K033048Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL…Cas Medical Systems, Inc.20041
−1K050844750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL…Cas Medical Systems, Inc.2005—
0 (this device)K060065CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter…Oridion Capnography, Inc.2006—
+1K072295Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc.2007—
+2K082268CAPNOSTREAM20 with Integrated Pulmonary IndexOridion Capnography, Inc.20099
+3K091241Masimo Rainbow Set Radical 7 Pulse CO-Oximeter; Rad 87 Pulse CO-OximeterMasimo Corporation2009—
+4K100428Masimo Rainbow Set Radical 7R Pulse CO-OximeterMasimo Corporation2010—
+3K094012CAPNOSTREAM20 with MICROMEDICO2 ModuleOridion Capnography, Inc.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199120
199210
199310
199540
199620
199721
199840
199930
200010
200210
200310
200410
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200810
200970
201080
201160
2012130
2013140
2014171
2015110
2016100
2017120
2018150
2019160
2020150
2021110
2022240
2023160
2024200
2025180
2026150
FIG. 1 — Lineage of K060065, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040011Polaris 2004 Capnograph
Summary page 2
…and alarms. SUBSTANTIAL EQUIVALENCE INFORMATION * CO 2 Module (MiniMediCO 2) used in legally marketed Predicate Devices: V Oridion Polaris 2004, K04001 1, V CAS Medical Systems, Inc Models 750c-2ms, 750cm-2ms, 750c-Nnl, 750cm with MiniMediCO 2-V1 K0…
Oridion, Ltd.2004—
*K050844750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS…
Summary page 2
…V Oridion Polaris 2004, K04001 1, V CAS Medical Systems, Inc Models 750c-2ms, 750cm-2ms, 750c-Nnl, 750cm with MiniMediCO 2-V1 K050844 V Larsen & Toubro Limited Star 50 Monitoring System K051608 * Pulse Oximeter Module, Sp02 Module (MP100)…
Cas Medical Systems, Inc.2005—
*K051608Star 50 Monitoring System
Summary page 2
…stems, Inc Models 750c-2ms, 750cm-2ms, 750c-Nnl, 750cm with MiniMediCO 2-V1 K050844 V Larsen & Toubro Limited Star 50 Monitoring System K051608 * Pulse Oximeter Module, Sp02 Module (MP100) used in legally marketed Predicate Devices: , NP…
Larsen & Toubro Limited2005—
*K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables…
Summary page 2
…Module (MP100) used in legally marketed Predicate Devices: , NPB OxiMax Pulse Oximeter System With N-595 Pulse Oximeter, K012891 v NPB Oximax N-550, K021090 * Nurse Call v NPB OxiMax Pulse Oximeter System With N-595 Pu…
Nellcor Puritan Bennett, Inc.20021
*K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse Oximeter
Summary page 2
…dicate Devices: , NPB OxiMax Pulse Oximeter System With N-595 Pulse Oximeter, K012891 v NPB Oximax N-550, K021090 * Nurse Call v NPB OxiMax Pulse Oximeter System With N-595 Pulse Oximeter, K012891…
Nellcor Puritan Bennett, Inc.2002—
*K022084Atlas Monitor, Models 200,210,220
Summary page 2
…Nurse Call v NPB OxiMax Pulse Oximeter System With N-595 Pulse Oximeter, K012891 · Welch Allyn Atlas Monitor K022084 CLASSIFICATION Capnostream 20 73CCK Class [I This device has two modules that are classified as follows: * 21 CFR 868.1400,…
Welch Allyn, Inc.2002—

Also cleared against K040011: 3 other devices →

What descends from it? (250)

See all 250

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072295Capnostream 20 with A2 Adaptive Averaging SoftwareOridion Capnography, Inc.2007197
K083821Vital Signs MonitorEdan Instruments, Inc.200939
K093016Vital Signs MonitorShenzhen Creative Industry Co., Ltd.201016
K100654Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters201011
K090556Spacelabs Elance Vital Signs Monitoring System and Central Monitor SoftwareSpacelabs Healthcare Medical Equipment (Suzhou) CO200910
K160582WiCapAthena Gtx20160
K091787Model 6000 Two Parameter Bedside MonitorIvy Biomedical Systems, Inc.20090

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 30, 2009Not classifiedZ-2232-2009Device Design1) The monitor's ability to withstand electrostatic discharge (ESD) to exposed metal connectors on the back of the monitor which may cause the monitor's display to freeze. 2) The unit generates electromagnetic radiation above the limits…
Jul 28, 2007Not classifiedZ-0177-2008Process designHandle Cracking - Cracks in the handle will lead to disconnection of the handle from the monitor, causing the monitor to fall.
Jul 28, 2007Not classifiedZ-0178-2008Process designHandle Cracking - Cracks in the handle will lead to disconnection of the handle from the monitor, causing the monitor to fall.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K970209Capnocheck PlusBci Intl., Inc. · 1997 · 3 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall
  • K123690CAPNOSTREAM20P with Hifi C02 MonitoringOridion Medical 1987, Ltd. · 2014 · 6 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 9 years and passes through 5 companies; a new generation was cleared every 11 months on median.
  • 6 of the 34 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K072295 (Oridion Capnography, Inc., 2007) has gone furthest, with 197 descendants of its own.
  • 3 other devices also cite its predicate K040011.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

The Capnostream 2 0 is intended for CO 2 and SPO2 indications. The Capnostream 20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO 2 and pulse…

Read the full statement (147 words)

The Capnostream 2 0 is intended for CO 2 and SPO2 indications. The Capnostream 20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO 2 and pulse rate). It is intended for use with neonatal, pediatric and adult patients in hospitals, hospital type facilities, intra hospital moves and home environments.

The Capnostreamlo is intended for C02 indications only. The Capnostream1 0 is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath. It is intended for use with neonatal, pediatric and adult patients in hospitals, hospital type facilities, intra hospital moves and home environments.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060065: CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K060065

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026