Cobas 6000 Series System

K060373510(k) SpecialCleared Mar 13, 200628-day reviewClass II21 CFR 862.1600Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics · CEM Electrode, Ion Specific, Potassium

Cleared on Mar 13, 2006 as substantially equivalent to K953239 (Boehringer Mannheim Corp., 1995) and 1 other predicate after a 28-day review. Its family tree reaches back 1 generation to 1995. 18 later devices descend from it. It has been recalled 10 times.

Decided
Mar 13, 2006
Review
28 days
Class
Class II
Product code
CEM
Recalls
10
Predicates
2

Its longest line of descent runs back 1 generation to K953239 (1995). 2 ancestors in all. 18 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor2010: 1 descendant cleared2011: 2 descendants cleared2012: 3 descendants cleared2017: 1 descendant cleared2019: 1 descendant cleared2013: 7 descendants cleared2026: 2 descendants cleared2022: 1 descendant clearedK953239 · Boehringer Mannheim/Hitachi 917 Analyzer · Boehringer Mannheim Corp. · 1995 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 6 descendants2 generations forward: 12 descendants K953239 · Boehringer Mannheim/Hitachi 917 Analyzer · Boehringer Mannheim Corp. · 1995 · recalled0
  1. Root · 1 generation backK953239 Boehringer Mannheim/Hitachi 917 AnalyzerBoehringer Mannheim Corp. · 1995 · Recalled
  2. This deviceK060373 Cobas 6000 Series SystemRoche Diagnostics · 2006
  3. Descendant · cites it directly · 9 of its ownK100853 Cobas 8000 Modular Series AnalyzerRoche Diagnostics Corp. · 2010 · Recalled
  4. Descendant · cites it directly · 3 of its ownK191899 Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro…Roche Diagnostics · 2019 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK253491 ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl…Roche Diagnostics · 2026
  6. Descendant · 2 generations on · 0 of its ownK253490 Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated…Roche Diagnostics · 2026
  7. Descendant · 2 generations on · 0 of its ownK220134 cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na…Roche Diagnostics · 2022

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2010: 12011: 22012: 32013: 72017: 12019: 12022: 12026: 2
2010high 7 (2013) · low 0 (2014)2026

Most influential descendants

Showing 5 of 18 · See all
  • K100853Cobas 8000 Modular Series AnalyzerRoche Diagnostics Corp. · 2010 · cites it directly9descendants
  • K191899Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated solutionsRoche Diagnostics · 2019 · cites it directly3descendants
  • K253491ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas…Roche Diagnostics · 2026 · 2 generations on0descendants
  • K253490Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutionsRoche Diagnostics · 2026 · 2 generations on0descendants
  • K220134cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na for Gen.2, ONLINE DAT…Roche Diagnostics · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K953239Boehringer Mannheim/Hitachi 917 AnalyzerBoehringer Mannheim Corp.199518
0 (this device)K060373Cobas 6000 Series SystemRoche Diagnostics2006—
+1K100853Cobas 8000 Modular Series AnalyzerRoche Diagnostics Corp.201011
+1K191899Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated…Roche Diagnostics20193
+2K253491ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for…Roche Diagnostics2026—
+2K253490Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutionsRoche Diagnostics2026—
+2K220134cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na for Gen.2…Roche Diagnostics2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201120
201230
201370
201710
201910
202210
202620
FIG. 1 — Lineage of K060373, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K953239Boehringer Mannheim/Hitachi 917 Analyzer
Summary page 3
…stantial equivalence to the predicate devices, the Roche/Hitachi Device MODULAR ANALYTICS System with MODULAR P and ISE Module, cleared in K953239 / A005, and the Roche Elecsys MODULAR El70, cleared in K961481 / A003.…
Boehringer Mannheim Corp.199518
*K961481Elecsys 2010 Analyzer
Summary page 2
…he e 601 module is analytically identical to the Elecsys El70 MODULAR ANALYTICS immunoassay analyzer, which was cleared in a letter to file K961481 / A003. Hardware changes are only cosmetic (shape, color) or housing related (sample rack interface), and do not i…
Boehringer Mannheim Corp.19966

Also cleared against K953239: 4 other devices →

What descends from it? (18)

See all 18

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K100853Cobas 8000 Modular Series AnalyzerRoche Diagnostics Corp.20109
K191899Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated solutionsRoche Diagnostics20193
K163569Elecsys CMV IgMRoche Diagnostics20170
K111188Cobas B 123 Poc System, Cobas B 123 Autoqc Pack Tri-Level, Cobas B 123 Autocvc Pack…Roche Diagnostics20120
K110726Roche Acetaminophen AssayRoche Diagnostics20110
K111753Hitachi Clinical Analyzer S Test Reagent Catridge for: Cholestrol, HDL, LDL, and…Hitachi Chemical Diagnostics, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 18, 2019Class IIZ-2476-2019Device DesignQuality issue with high pressure solenoid valves may cause inaccurate results.
Dec 18, 2017Class IIZ-1268-2018Labeling False and MisleadingComplaints have been received concerning questionable results on analyzers due to customers utilizing 13 sample tubes without tube adapters.
Jul 18, 2017Class IIZ-0909-2018Device DesignA warning is being added to the cobas 6000 MODULAR Series System - c501 and the cobas 8000 MODULAR Analyzer Series - c502 Operator Manuals to prevent operator injury during maintenance.
Jun 29, 2017Class IIZ-0378-2018Software designRoche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.
Mar 21, 2017Class IIZ-2049-2017Component change controlDue to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the…
Jul 15, 2016Class IIZ-0096-2017Device DesignRoche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
May 21, 2014Class IIZ-1791-2014Labeling designUnder certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
May 21, 2014Class IIZ-1792-2014Labeling designUnder certain conditions loose ProCell/CleanCell aspiration tube filters may cause: consecutive discrepant results; sipper alarm; foam on reservoirs and system reagent alarms.
Sep 15, 2008Not classifiedZ-0542-2009Package design/selectionTest results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please…
Dec 4, 2007Not classifiedZ-0576-2008Software designIncorrect results: CRP assays may be calculated using incorrect calibration parameters which could result in falsely high or falsely low patient results being reported.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K100853 (Roche Diagnostics Corp., 2010) has gone furthest, with 9 descendants of its own.
  • 4 other devices also cite its predicate K953239.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitiative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE), a photometric analysis unit and an immunoassay analysis module. The cobas c501 analyzer is a fully…

Read the full statement (442 words)

The cobas 6000 series is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitiative in vitro determinations using a wide variety of tests. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE), a photometric analysis unit and an immunoassay analysis module.

The cobas c501 analyzer is a fully automated, discrete clinical chemistry analyzer intended for the in vitro quantitative/ qualitative determination of analytes in body fluids.

The cobas e601 analyzer is a fully automated discrete immunoassay analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.

The a-Amylase EPS ver.2 assay is an in vitro test for the quantitative determination of ot-Amylase in serum, plasma and urine on Roche/Hitachi cobas c systems. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis.

The Uric Acid ver.2 assay is an in vitro test for the quantitative determination of uric acid in human serum, plasma and urine on Roche/Hitachi cobas c systems. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions, and of patients receiving cytotoxic drugs.

The Tina-quant Complement C3c ver.2 assay is an in vitro test for the quantitative determination of Complement C3c in human serum and plasma on Roche/Hitachi cobas c systems. Measurement of complement C3c aids in the diagnosis of immunologic disorders, especially those associated with deficiencies of complement components.

The ISE module of the Roche/Hitachi cobas c systems is intended for the quantitative determination of sodium, potassium and chloride in serum, plasma or urine using ion-selective electrodes. Measurements obtained by this device are used in the diagnosis and treatment of diseases or conditions involving electrolyte imbalance.

The Acetaminophen assay is an in vitro test for the quantitative determination of toxic levels of acetaminophen in serum and plasma on Roche/Hitachi cobas c systems. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose.

The Opiates LI assay is an in vitro diagnostic test for the qualitative and semniquantitative detection of morphine and its metabolites in human urine on Roche/Hitachi cobas c systems at cutoff concentrations of 300 and 2000 ng/mL. Semniquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Opiates II provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060373: Cobas 6000 Series System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K060373

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026