510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated solutions

K191899510(k) TraditionalCleared Aug 16, 201931-day reviewClass II21 CFR 862.1345Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics · CFR Hexokinase, Glucose

Cleared on Aug 16, 2019 as substantially equivalent to K060373 (Roche Diagnostics, 2006) and 1 other predicate after a 31-day review. Its family tree reaches back 2 generations to 1995. 3 later devices descend from it. It has been recalled 3 times.

Decided
Aug 16, 2019
Review
31 days
Class
Class II
Product code
CFR
Recalls
3
Predicates
2

Its longest line of descent runs back 2 generations to K953239 (1995). 4 ancestors in all. 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2019: 1 other ancestor1996: 1 other ancestor2026: 2 descendants cleared2022: 1 descendant clearedK953239 · Boehringer Mannheim/Hitachi 917 Analyzer · Boehringer Mannheim Corp. · 1995 · recalledK060373 · Cobas 6000 Series System · Roche Diagnostics · 2006 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 3 descendants K060373 · Cobas 6000 Series System · Roche Diagnostics · 2006 · recalledK953239 · Boehringer Mannheim/Hitachi 917 Analyzer · Boehringer Mannheim Corp. · 1995 · recalled0
  1. Root · 2 generations backK953239 Boehringer Mannheim/Hitachi 917 AnalyzerBoehringer Mannheim Corp. · 1995 · Recalled
  2. 1 generation backK060373 Cobas 6000 Series SystemRoche Diagnostics · 2006 · Recalled
  3. This deviceK191899 Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro…Roche Diagnostics · 2019
  4. Descendant · cites it directly · 0 of its ownK253491 ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl…Roche Diagnostics · 2026
  5. Descendant · cites it directly · 0 of its ownK253490 Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated…Roche Diagnostics · 2026
  6. Descendant · cites it directly · 0 of its ownK220134 cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na…Roche Diagnostics · 2022

Most influential descendants

Showing 3 of 3 · See all
  • K253491ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas…Roche Diagnostics · 2026 · cites it directly0descendants
  • K253490Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutionsRoche Diagnostics · 2026 · cites it directly0descendants
  • K220134cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na for Gen.2, ONLINE DAT…Roche Diagnostics · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K953239Boehringer Mannheim/Hitachi 917 AnalyzerBoehringer Mannheim Corp.199518
−1K060373Cobas 6000 Series SystemRoche Diagnostics200610
0 (this device)K191899Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated…Roche Diagnostics2019—
+1K253491ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for…Roche Diagnostics2026—
+1K253490Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutionsRoche Diagnostics2026—
+1K220134cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na for Gen.2…Roche Diagnostics2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
201910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202620
FIG. 1 — Lineage of K191899, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CFR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CFR device

  1. #1 K253490 Glucose HK Gen.3; ONLINE DAT Methadone II…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060373Cobas 6000 Series System
Summary page 13
…e 2: Substantial Equivalency System Topic cobas 6000 Analyzer Series cobas pro integrated (k060373) solutions (k191899)…
Roche Diagnostics200610
*K190773Elecsys TSH
Summary page 15
…SH Elecsys TSH Item ( e 801 in cobas 8000 core) (e 801 in cobas pro core) k190773 k191899 Proprietary name Elecsys TSH Elecsys TSH Cata…
Roche Diagnostics2019—

Also cleared against K060373: 5 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K253491ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas…Roche Diagnostics20260
K253490Glucose HK Gen.3; ONLINE DAT Methadone II; cobas pro integrated solutionsRoche Diagnostics20260
K220134cobas pure integrated solutions, Glucose HK Gen.3, ISE indirect Na for Gen.2, ONLINE DAT…Roche Diagnostics20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 20, 2026Class IIZ-1476-2026Software designSoftware defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and…
Dec 8, 2020Class IIZ-0755-2021Software designPotential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may…
Nov 23, 2020Class IIZ-0752-2021Device DesignDuring the subsequent pipetting (1st or 2nd measurement), the residual reagent can sporadically be transferred from the conus into the reaction cuvette and lead to inaccurately low results for the affected TDMs (Therapeutic Drug…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 2 companies; a new generation was cleared every 13.4 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K253491 (Roche Diagnostics, 2026) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K060373.

How does its category compare?

All CFR devices

CFR Hexokinase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · CFR

1in 2025

2007: 12008: 52009: 32011: 32012: 42013: 12014: 12016: 22017: 12019: 12022: 12025: 12026 (partial year): 1
2007high 5 (2008) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CFR

115in 2012

2008: 182 days median2009: 101 days median2011: 316 days median2012: 115 days median
2007high 316 (2011) · low 101 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CFR

1in 2016

2008: 72009: 12012: 12014: 12016: 1
2007high 7 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CFR

16in 2025

2007: 12008: 22009: 72010: 112011: 432012: 492013: 362014: 22015: 32016: 12017: 132018: 82019: 102020: 42021: 52022: 32023: 292024: 152025: 16
2007high 49 (2012) · low 1 (2007)2025

MAUDE reports naming product code CFR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008518272
2009310117
20100—011
20113316043
20124115149
20131—036
20141—12
20150—03
20162—11
20171—013
20180—08
20191—010
20200—04
20210—05
20221—03
20230—029
20240—015
20251—016
20261—0—

What is it indicated for?

The cobas pro integrated solutions is an IVD device used for the quantitation of clinical chemistry/ immunochemistry and Ion Selective Electrolyte parameters from various biological fluids. Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on Roche/Hitachi cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including…

Read the full statement (197 words)

The cobas pro integrated solutions is an IVD device used for the quantitation of clinical chemistry/ immunochemistry and Ion Selective Electrolyte parameters from various biological fluids.

Glucose HK Gen.3 is an in vitro test for the quantitative determination of glucose in human serum, plasma, urine and CSF on Roche/Hitachi cobas c systems. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia and pancreatic islet cell tumors.

The ISE indirect Na for Gen. 2 is intended for the quantitative determination of sodium in serum, plasma or urine using ion‑selective electrodes. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.

Elecsys TSH immunoassay is intended for the in vitro quantitative determination of thyrotropin in human serum and plasma. Measurements of TSH are used in the diagnosis of thyroid and pituitary disorders. The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K191899: Glucose HK Gen.3, ISE indirect Na for Gen.2, Elecsys TSH, cobas pro integrated solutions.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191899

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026