Recalled · 2009Coherence Dosimetrist, Model 2.2
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Solutions USA, Inc. · MUJ System, Planning, Radiation Therapy Treatment
Cleared on Jun 7, 2006 as substantially equivalent to K022036 (Siemens Medical Solutions USA, Inc., 2002) and 1 other predicate after a 49-day review. Its family tree reaches back 6 generations to 1992. 19 later devices descend from it. It has been recalled once.
- Decided
- Jun 7, 2006
- Review
- 49 days
- Class
- Class II
- Product code
- MUJ
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 6 generations to K914698 (1992). 23 ancestors in all, the oldest K841997 (1984). 19 later devices descend from it, 1 citing it directly.
- Root · 6 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
- 5 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
- 4 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
- 3 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
- 2 generations backK013112 Focal SimComputerized Medical Systems, Inc. · 2001 · Recalled
- 1 generation backK022036 Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc. · 2002
- This deviceK061097 Coherence Dosimetrist, Model 2.2Siemens Medical Solutions USA, Inc. · 2006
- Descendant · cites it directly · 18 of its ownK101119 Syngo Dosimetrist Workspace V2.7 Model: 1056864Siemens Medical Solutions USA, Inc. · 2010 · Recalled
- Descendant · 2 generations on · 9 of its ownK151380 syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2015
- Descendant · 2 generations on · 8 of its ownK162370 syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2016
- Descendant · 2 generations on · 8 of its ownK151887 syngo VSimSiemens Medical Solutions USA, Inc. · 2015
- Descendant · 3 generations on · 7 of its ownK173635 syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2018
Showing 5 of 19 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 19 · See all- K101119Syngo Dosimetrist Workspace V2.7 Model: 1056864Siemens Medical Solutions USA, Inc. · 2010 · cites it directly18descendants
- K151380syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2015 · 2 generations on9descendants
- K162370syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2016 · 2 generations on8descendants
- K151887syngo VSimSiemens Medical Solutions USA, Inc. · 2015 · 2 generations on8descendants
- K173635syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2018 · 3 generations on7descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K914698 | Setup Computer with Software V-1261,62 and 63 | K.S., Inc. | 1992 | — |
| −5 | K980904 | Mucheck Monitor Unit Validation Program | Medical Dosimetry Services, Inc. | 1998 | — |
| −4 | K990833 | Dose Calc, Model VS. 1.02 | Integrity Medical Software, Inc. | 1999 | — |
| −3 | K993923 | Adac PINNACLE3TM Radiation Therapy Planning System | Adac Laboratories | 2000 | — |
| −2 | K013112 | Focal Sim | Computerized Medical Systems, Inc. | 2001 | 10 |
| −1 | K022036 | Siemens Virtual Simulation (Vsim) | Siemens Medical Solutions USA, Inc. | 2002 | — |
| 0 (this device) | K061097 | Coherence Dosimetrist, Model 2.2 | Siemens Medical Solutions USA, Inc. | 2006 | — |
| +1 | K101119 | Syngo Dosimetrist Workspace V2.7 Model: 1056864 | Siemens Medical Solutions USA, Inc. | 2010 | 1 |
| +2 | K151380 | syngo.via RT Image Suite | Siemens Medical Solutions USA, Inc. | 2015 | — |
| +2 | K162370 | syngo.via RT Image Suite | Siemens Medical Solutions USA, Inc. | 2016 | — |
| +2 | K151887 | syngo VSim | Siemens Medical Solutions USA, Inc. | 2015 | — |
| +3 | K173635 | syngo.via RT Image Suite | Siemens Medical Solutions USA, Inc. | 2018 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1984 | 1 | 0 |
| 1989 | 3 | 0 |
| 1990 | 2 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 2 | 0 |
| 1995 | 1 | 0 |
| 1996 | 2 | 0 |
| 1998 | 2 | 0 |
| 2000 | 1 | 0 |
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 3 | 0 |
| 2014 | 2 | 0 |
| 2015 | 2 | 0 |
| 2016 | 1 | 0 |
| 2018 | 2 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 1 | 0 |
| 2022 | 1 | 0 |
| 2024 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MUJ device
- #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K022036 | Siemens Virtual Simulation (Vsim)Summary page 1…g, Radiation Therapy Treatment Classification: 892.5050 Product Code: MUJ Substantial Equivalence Claimed To: VSIM K022036 (cleared on November 5, 2002) Market name: COHERENCEDosimetrist Workspace v2.0 KonRad, Model v2.0… | 2002 |
| * | K022307 | Konrad, Model V.2.0Summary page 1…M K022036 (cleared on November 5, 2002) Market name: COHERENCEDosimetrist Workspace v2.0 KonRad, Model v2.0 K022307 (cleared on October 8, 2002) The Food and Drug Administration was notified on August 22, 2003 that Siemens Medical So… | 2002 |
| † | K010938 | Syngo Multimodality Workstation | 2001 |
What descends from it? (19)
See all 191 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K101119 | Syngo Dosimetrist Workspace V2.7 Model: 1056864 | 2010 | 18 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 6, 2009 | Not classified | Z-1834-2011 | Software design | A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose, for the OAR (Organs… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 6 companies; a new generation was cleared every 1.7 years on median.
- 3 of the 23 ancestors have a recall of any class.
- Of its descendants, K101119 (Siemens Medical Solutions USA, Inc., 2010) has gone furthest, with 18 descendants of its own.
- 1 other device also cites its predicate K022036.
How does its category compare?
All MUJ devicesMUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MUJ
10in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MUJ
186in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MUJ
2in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MUJ
21in 2025
MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 67 | 4 | 24 |
| 2008 | 9 | 63 | 10 | 51 |
| 2009 | 16 | 84 | 11 | 24 |
| 2010 | 11 | 56 | 23 | 78 |
| 2011 | 9 | 103 | 17 | 73 |
| 2012 | 8 | 59 | 4 | 39 |
| 2013 | 5 | 64 | 6 | 33 |
| 2014 | 9 | 108 | 2 | 35 |
| 2015 | 13 | 78 | 7 | 30 |
| 2016 | 5 | 80 | 13 | 27 |
| 2017 | 7 | 30 | 4 | 16 |
| 2018 | 7 | 53 | 3 | 25 |
| 2019 | 9 | 56 | 4 | 25 |
| 2020 | 9 | 104 | 2 | 26 |
| 2021 | 11 | 87 | 4 | 15 |
| 2022 | 8 | 168 | 3 | 22 |
| 2023 | 10 | 140 | 2 | 76 |
| 2024 | 9 | 122 | 0 | 9 |
| 2025 | 10 | 186 | 0 | 21 |
| 2026 | 9 | 233 | 2 | — |
What is it indicated for?
The COHERENCE Dosimetrist Workspace v2.2 is a comprehensive oncology worldlow software package that allows for both CT simulation as well as inverse radiation therapy treatment planning and optimization in one software package to aid in oncology clinical workflow. This workspace is comprised of two major components, the CT Simulation component (VSIM) and the inverse radiation therapy treatment planning component (KonRad). The VSIM component permits CT…
Read the full statement (242 words)
The COHERENCE Dosimetrist Workspace v2.2 is a comprehensive oncology worldlow software package that allows for both CT simulation as well as inverse radiation therapy treatment planning and optimization in one software package to aid in oncology clinical workflow. This workspace is comprised of two major components, the CT Simulation component (VSIM) and the inverse radiation therapy treatment planning component (KonRad).
The VSIM component permits CT simulation to be performed on the syngo workstation. The CT scans are first loaded into the VSIM software component and the user is able to create three-dimensional models of targets and organs. The user is able to identify the patient isocenter, place treatment beams, and identify beam modifiers (blocks, apertures, and MLCs). The information is then available for radiation treatment planning for dose calculation via the KonRad software component or other treatment planning systems. The plans are then reviewed and approved by the clinician prior to transfer to the delivery system for the actual treatment.
The KonRad software component is intended to optimize multi-leaf (MLC) positions or partial attenuation block shapes for intensity modulated external beam radiation therapy (IMRT). Once the optimization is complete the dose distribution and dose volume histogram curves are displayed for the user to evaluate. After approval the results are exported to the delivery equipment, linear accelerator, or record and verify system, for final verification prior to treatment delivery. The KonRad software component allows for efficient inverse radiation therapy treatment planning and optimization.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K061097 · printed Nov 4, 2026
Cite this page
510(k) Report. “K061097: Coherence Dosimetrist, Model 2.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K061097
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026