510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Syngo Dosimetrist Workspace V2.7 Model: 1056864

K101119510(k) TraditionalCleared Jun 16, 201056-day reviewClass II21 CFR 892.5050Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · MUJ System, Planning, Radiation Therapy Treatment

Cleared on Jun 16, 2010 as substantially equivalent to K061097 (Siemens Medical Solutions USA, Inc., 2006) and 3 other predicates after a 56-day review. Its family tree reaches back 7 generations to 1992. 18 later devices descend from it. It has been recalled once.

Decided
Jun 16, 2010
Review
56 days
Class
Class II
Product code
MUJ
Recalls
1
Predicates
4

Its longest line of descent runs back 7 generations to K914698 (1992). 55 ancestors in all, the oldest K841997 (1984). 18 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2007: 1 other ancestor2006: 3 other ancestors2004: 3 other ancestors2002: 3 other ancestors2001: 5 other ancestors1996: 5 other ancestors2005: 1 other ancestor2003: 1 other ancestor2000: 3 other ancestors1995: 2 other ancestors1992: 2 other ancestors1991: 1 other ancestor1997: 2 other ancestors1984: 1 other ancestor1999: 1 other ancestor1998: 3 other ancestors1994: 2 other ancestors1993: 2 other ancestors1990: 2 other ancestors1989: 3 other ancestors1987: 1 other ancestor2016: 1 descendant cleared2015: 2 descendants cleared2012: 3 descendants cleared2011: 1 descendant cleared2014: 2 descendants cleared2018: 2 descendants cleared2019: 1 descendant cleared2020: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared2024: 1 descendant cleared2026: 2 descendants clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K061097 · Coherence Dosimetrist, Model 2.2 · Siemens Medical Solutions USA, Inc. · 2006 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 8 other ancestors3 generations back: 10 other ancestors4 generations back: 9 other ancestors5 generations back: 12 other ancestors6 generations back: 5 other ancestors7 generations back: 1 other ancestor 1 generation forward: 6 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 2 descendants K061097 · Coherence Dosimetrist, Model 2.2 · Siemens Medical Solutions USA, Inc. · 2006 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992−50+5
  1. Root · 7 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 6 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 5 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 4 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 3 generations backK013112 Focal SimComputerized Medical Systems, Inc. · 2001 · Recalled
  6. 2 generations backK022036 Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc. · 2002
  7. 1 generation backK061097 Coherence Dosimetrist, Model 2.2Siemens Medical Solutions USA, Inc. · 2006 · Recalled
  8. This deviceK101119 Syngo Dosimetrist Workspace V2.7 Model: 1056864Siemens Medical Solutions USA, Inc. · 2010
  9. Descendant · cites it directly · 9 of its ownK151380 syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2015
  10. Descendant · cites it directly · 8 of its ownK162370 syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2016
  11. Descendant · cites it directly · 8 of its ownK151887 syngo VSimSiemens Medical Solutions USA, Inc. · 2015
  12. Descendant · 2 generations on · 7 of its ownK173635 syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2018
  13. Descendant · 3 generations on · 6 of its ownK192065 Syngo.Via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2019 · Recalled

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2011: 12012: 32014: 22015: 22016: 12018: 22019: 12020: 12021: 12022: 12024: 12026: 2
2011high 3 (2012) · low 0 (2013)2026

Most influential descendants

Showing 5 of 18 · See all
  • K151380syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2015 · cites it directly9descendants
  • K162370syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2016 · cites it directly8descendants
  • K151887syngo VSimSiemens Medical Solutions USA, Inc. · 2015 · cites it directly8descendants
  • K173635syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2018 · 2 generations on7descendants
  • K192065Syngo.Via RT Image SuiteSiemens Medical Solutions USA, Inc. · 2019 · 3 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−6K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−5K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−4K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−3K013112Focal SimComputerized Medical Systems, Inc.200110
−2K022036Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc.2002—
−1K061097Coherence Dosimetrist, Model 2.2Siemens Medical Solutions USA, Inc.20061
0 (this device)K101119Syngo Dosimetrist Workspace V2.7 Model: 1056864Siemens Medical Solutions USA, Inc.2010—
+1K151380syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc.2015—
+1K162370syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc.2016—
+1K151887syngo VSimSiemens Medical Solutions USA, Inc.2015—
+2K173635syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc.2018—
+3K192065Syngo.Via RT Image SuiteSiemens Medical Solutions USA, Inc.20192

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198710
198930
199020
199110
199220
199320
199420
199520
199650
199720
199830
199910
200030
200150
200230
200310
200430
200510
200630
200710
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201230
201420
201520
201610
201820
201910
202010
202110
202210
202410
202620
FIG. 1 — Lineage of K101119, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUJ device

  1. #1 K232799 syngo.via RT Image Suitescore 18.6 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061097Coherence Dosimetrist, Model 2.2
Summary page 1
…ode: MUJ Substantial Equivalence Claimed To: PRODUCT Clearance Claim of Equivalence For: SIEMENS COHERENCE TM K061097 The Siemens syngo® Dosirnetrist Workspace v2.7 Dosimetrist Workspace v2.2 SIEMENS syngo®)CT K07131 0…
Siemens Medical Solutions USA, Inc.20061
*K071310Syngo CT Oncology Software Package
Summary page 1
…valence For: SIEMENS COHERENCE TM K061097 The Siemens syngo® Dosirnetrist Workspace v2.7 Dosimetrist Workspace v2.2 SIEMENS syngo®)CT K07131 0 Auto Segmentation using the Random Walker algorithm ONCOLOGY Software for automated volume s…
Siemens AG2007—
*K080799Atlas-Based Autosegmentation
Summary page 1
…mentation using the Random Walker algorithm ONCOLOGY Software for automated volume segmentation. Package Computerized Medical K080799 Auto-segmentation using predefined contours based on Systems, Inc. anatomical libraries…
Computerized Medical Systems, Inc.2008—
*K061006Ikoengelo
Summary page 1
…application for atlas based segmentation using deformable registration algorithms. IKOEtech, LLC, K061006 Automatic contour delineation to support the radiotherapy lKOEngeloTM treatment plann…
Ikoetech, LLC2006—

What descends from it? (18)

See all 18

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151380syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc.20159
K162370syngo.via RT Image SuiteSiemens Medical Solutions USA, Inc.20168
K151887syngo VSimSiemens Medical Solutions USA, Inc.20158
K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc.20114
K120363Syngo RT InterfaceSiemens Medical Solutions USA, Inc.20120
K111861Syngo.MR Onco, Syngo.MR 3D Lesion Segmentation, Syngo.MR Onco EngineSiemens Medical Systems USA20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 15, 2014Class IIZ-2075-2014Software designAn update to the Virtual Simulation software of the syngo RT Dosimetrist is necessary to address a safety issue. In special scenarios within the Structure Operations function the system did not store the selected name from the list of…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 6 companies; a new generation was cleared every 1.7 years on median.
  • 13 of the 55 ancestors have a recall of any class.
  • Of its descendants, K151380 (Siemens Medical Solutions USA, Inc., 2015) has gone furthest, with 9 descendants of its own.

How does its category compare?

All MUJ devices

MUJ System, Planning, Radiation Therapy Treatment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MUJ

10in 2025

2007: 42008: 92009: 162010: 112011: 92012: 82013: 52014: 92015: 132016: 52017: 72018: 72019: 92020: 92021: 112022: 82023: 102024: 92025: 102026 (partial year): 9
2007high 16 (2009) · low 4 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUJ

186in 2025

2007: 67 days median2008: 63 days median2009: 84 days median2010: 56 days median2011: 103 days median2012: 59 days median2013: 64 days median2014: 108 days median2015: 78 days median2016: 80 days median2017: 30 days median2018: 53 days median2019: 56 days median2020: 104 days median2021: 87 days median2022: 168 days median2023: 140 days median2024: 122 days median2025: 186 days median2026 (partial year): 233 days median
2007high 233 (2026) · low 30 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUJ

2in 2023

2007: 42008: 102009: 112010: 232011: 172012: 42013: 62014: 22015: 72016: 132017: 42018: 32019: 42020: 22021: 42022: 32023: 22026 (partial year): 2
2007high 23 (2010) · low 0 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUJ

21in 2025

2007: 242008: 512009: 242010: 782011: 732012: 392013: 332014: 352015: 302016: 272017: 162018: 252019: 252020: 262021: 152022: 222023: 762024: 92025: 21
2007high 78 (2010) · low 9 (2024)2025

MAUDE reports naming product code MUJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007467424
20089631051
200916841124
201011562378
201191031773
2012859439
2013564633
20149108235
20151378730
20165801327
2017730416
2018753325
2019956425
20209104226
20211187415
20228168322
202310140276
2024912209
202510186021
202692332—

What is it indicated for?

The intended use of syngo® Dosimetrist Workspace is as an accessory to the linear accelerator system to aid and support in the planning of delivery of x-ray photon and electron radiation for the therapeutic-treatment of cancer. The syngo®Dosimetrist Workspace v2.7 is a comprehensive oncology workflow software package that allows for both CT simulation as well as inverse radiation therapy treatment planning and optimization to aid in the oncology clinical…

Read the full statement (335 words)

The intended use of syngo® Dosimetrist Workspace is as an accessory to the linear accelerator system to aid and support in the planning of delivery of x-ray photon and electron radiation for the therapeutic-treatment of cancer.

The syngo®Dosimetrist Workspace v2.7 is a comprehensive oncology workflow software package that allows for both CT simulation as well as inverse radiation therapy treatment planning and optimization to aid in the oncology clinical workdflow where indicated. The syngo® Dosimetrist workspace is comprised of two major components, the CT Simulation component (VSIM) and the inverse radiation therapy treatment planning component (KonRad) to be used in the creation, modification, evaluation and approval of radiation treatment plans.

The VSIM component permits CT simulation to be performed on the syngo© Dosimetrist workstation. The CT scans are imported into the VSIM software component and the user is able to create three-dimensional models of targets and organs. The VSIM application supports the use of automatic contouring and segmentation or a model based segmentation (MBS) library of anatomical regions may be used. The contours and segmentation can be manually adjusted prior to use by the treatment planning system.

Additionally, the user is able to identify the patient isocenter, place treatment beams, and identify beam modifiers (blocks, apertures, and ML~s). The simulation information is then available for radiation treatment planning for dose calculation via the KonRad software component or other treatment planning systems. The plans are then reviewed and approved by the clinician prior to transfer to the delivery system for the actual treatment.

The KonRad software component is intended to optimize multi-leaf (MLC) positions or partial attenuation block shapes for intensity modulated external beam radiation therapy (1IMRT). Once the optimization is complete, the dose distribution and dose volume histogram curves are displayed for the user to evaluate. After approval, the results are exported to the delivery equipment, linear accelerator, or record and verify system, for final verification prior to treatment delivery. The KonRad TPS software component allows for efficient inverse radiation therapy treatment planning and optimization.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101119: Syngo Dosimetrist Workspace V2.7 Model: 1056864.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101119

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026