Coulter LH 780 Hematology Analyzer

K061616510(k) AbbreviatedCleared Oct 4, 2006118-day reviewClass II21 CFR 864.5220Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · GKZ Counter, Differential Cell

Cleared on Oct 4, 2006 as substantially equivalent to K011342 (Beckman Coulter, Inc., 2001) and 2 other predicates after a 118-day review. Its family tree reaches back 3 generations to 1991. 1 later device descends from it. It has been recalled 7 times.

Decided
Oct 4, 2006
Review
118 days
Class
Class II
Product code
GKZ
Recalls
7
Predicates
3

Its longest line of descent runs back 3 generations to K912494 (1991). 7 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2001: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2008: 1 descendant clearedK912494 · Sysmex (TM) R-3000 Automated Reticulocyte Analyzer · Toa Medical Electronics USA, Inc. · 1991K964375 · Sysmex SE/Ram-1 · Sysmex Corp. · 1997K992875 · Sysmex XE-2100, Automatated Hematology Analyzer · Sysmex Corp. · 1999 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 1 descendant K992875 · Sysmex XE-2100, Automatated Hematology Analyzer · Sysmex Corp. · 1999 · recalledK964375 · Sysmex SE/Ram-1 · Sysmex Corp. · 1997K912494 · Sysmex (TM) R-3000 Automated Reticulocyte Analyzer · Toa Medical Electronics USA, Inc. · 19910
  1. Root · 3 generations backK912494 Sysmex (TM) R-3000 Automated Reticulocyte AnalyzerToa Medical Electronics USA, Inc. · 1991
  2. 2 generations backK964375 Sysmex SE/Ram-1Sysmex Corp. · 1997
  3. 1 generation backK992875 Sysmex XE-2100, Automatated Hematology AnalyzerSysmex Corp. · 1999 · Recalled
  4. This deviceK061616 Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc. · 2006
  5. Descendant · cites it directly · 0 of its ownK081930 Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2008 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K081930Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K912494Sysmex (TM) R-3000 Automated Reticulocyte AnalyzerToa Medical Electronics USA, Inc.1991—
−2K964375Sysmex SE/Ram-1Sysmex Corp.1997—
−1K992875Sysmex XE-2100, Automatated Hematology AnalyzerSysmex Corp.19991
0 (this device)K061616Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc.2006—
+1K081930Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc.200829

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200110
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
FIG. 1 — Lineage of K061616, 1991–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011342Coulter LH 750 Hematology Analyzer, Model 6605632
Summary page 1
…Number COULTER" LH 780 COULTER ® LH 750 Beckman Coulter, Inc. K011342 Hematology Analyzer Hematology Analyzer COULTER® LH 750…
Beckman Coulter, Inc.200115
*K030606Coulter LH 750 Body Fluids Application
Summary page 1
…c. K011342 Hematology Analyzer Hematology Analyzer COULTER® LH 750 Beckman Coulter, Inc. K030606 Body Fluids Application (for Body Fluids equivalence only)…
Beckman Coulter, Inc.20031
*K992875Sysmex XE-2100, Automatated Hematology Analyzer
Summary page 1
…n (for Body Fluids equivalence only) SysmexTM Automated Sysmex Corporation K992875 Hematology Analyzer XE-2100…
Sysmex Corp.19991

Also cleared against K011342: 7 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081930Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 22, 2015Class IIZ-2481-2015Nonconforming Material/ComponentBeckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.
Jul 22, 2015Class IIZ-2480-2015Nonconforming Material/ComponentBeckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.
Nov 8, 2013Class IIZ-0471-2014Under Investigation by firmBeckman Coulter is recalling the LH 750 and LH 780 Hematology Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), white blood cell differential and/or…
Sep 24, 2013Class IIZ-0113-2014Component change controlThe check valves may fail on the Coulter LH 750 Hematology Analyzer (PN 6605632 and A85570) and Coulter LH 780 Hematology Analyzer (PN 723585 and A90728). A failure may result in an air or liquid leak at the connection to the Vacuum…
Dec 1, 2011Not classifiedZ-1241-2012Software designThe recall was initiated because Beckman Coulter has confirmed that the Auto Stop and Auto Transmit features for Quality Control become disabled after scanning assay values from Coulter 5C and Retic-C Cell Controls assay sheets using the…
Jan 31, 2011Not classifiedZ-1954-2011Software designThis recall was initiated because Beckman Coulter identified a problem with some date fields that are transmitted to the Laboratory Information System (LIS). Starting Jan 1, 2011, the date fields will transmit as a blank field to the LIS…
Apr 18, 2007Not classifiedZ-0313-2008Software designPatient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 5.5 years on median.
  • 3 of the 7 ancestors have a recall of any class.
  • Of its descendants, K081930 (Beckman Coulter, Inc., 2008) has gone furthest, with 0 descendants of its own.
  • 7 other devices also cite its predicate K011342.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

The COULTER LH 780 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER LH 780 Hematology Analyzer provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061616: Coulter LH 780 Hematology Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061616

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.