Unicel DXH 800 Coulter Cellular Analysis System

K081930510(k) AbbreviatedCleared Dec 19, 2008165-day reviewClass II21 CFR 864.5220Recalls 29

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · GKZ Counter, Differential Cell

Cleared on Dec 19, 2008 as substantially equivalent to K011342 (Beckman Coulter, Inc., 2001) and 1 other predicate after a 165-day review. Its family tree reaches back 4 generations to 1991. It has been recalled 29 times.

Decided
Dec 19, 2008
Review
165 days
Class
Class II
Product code
GKZ
Recalls
29
Predicates
2

Its longest line of descent runs back 4 generations to K912494 (1991). 8 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK912494 Sysmex (TM) R-3000 Automated Reticulocyte AnalyzerToa Medical Electronics USA, Inc. · 1991
  2. 3 generations backK964375 Sysmex SE/Ram-1Sysmex Corp. · 1997
  3. 2 generations backK992875 Sysmex XE-2100, Automatated Hematology AnalyzerSysmex Corp. · 1999 · Recalled
  4. 1 generation backK061616 Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc. · 2006 · Recalled
  5. This deviceK081930 Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2008
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K912494Sysmex (TM) R-3000 Automated Reticulocyte AnalyzerToa Medical Electronics USA, Inc.1991—
−3K964375Sysmex SE/Ram-1Sysmex Corp.1997—
−2K992875Sysmex XE-2100, Automatated Hematology AnalyzerSysmex Corp.19991
−1K061616Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc.20067
0 (this device)K081930Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc.2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200110
200310
FIG. 1 — Lineage of K081930, 1991–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011342Coulter LH 750 Hematology Analyzer, Model 6605632
Summary page 1
…Candidate Predicates Manufacturer Number UniCel®)Dxli 800 COULTER® LH 750 Beckman Coulter, Inc. K01 1342 Coulter®~ Cellular Hematology Analyzer Analysis System COULTER® LH 780…
Beckman Coulter, Inc.200115
*K061616Coulter LH 780 Hematology Analyzer
Summary page 1
…1 1342 Coulter®~ Cellular Hematology Analyzer Analysis System COULTER® LH 780 Beckman Coulter, Inc. K061616 Hematology Analyzer UniCelo DxH 800 Analyzer…
Beckman Coulter, Inc.20067

Also cleared against K011342: 7 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2017Class IIZ-2158-2017Device DesignThere is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual…
Apr 24, 2017Class IIZ-2157-2017Device DesignThere is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual…
Dec 19, 2016Class IIZ-1013-2017Under Investigation by firmBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood…
Dec 19, 2016Class IIZ-1014-2017Under Investigation by firmBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood…
Dec 1, 2015Class IIZ-0433-2016Software designBeckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen…
Dec 1, 2015Class IIZ-0434-2016Software designBeckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen…
Dec 1, 2015Class IIZ-0432-2016Software designBeckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen…
Apr 13, 2015Class IIZ-1616-2015Nonconforming Material/ComponentBeckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result…
Nov 8, 2013Class IIZ-1828-2014Under Investigation by firmBeckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell…
Nov 8, 2013Class IIZ-1829-2014Under Investigation by firmBeckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell…
Sep 24, 2013Class IIZ-0147-2014Under Investigation by firmBeckman Coulter is initiating a recall for UniCel DxH 800 Coulter Cellular Analysis System (PN 629029) and UniCel DxH 600 Coulter Cellular Analysis System (PN B23858) because the company's internal investigation indicates that specific…
Apr 3, 2012Class IIZ-0021-2013Device DesignBeckman Coulter is recalling the UniCel DxH 800 Coulter Cellular Analysis System because the NRBC, Differential or Reticulocyte mix chamber drain ports may become plugged by rubber pieces from specimen tube caps, and result in biohazard…
May 9, 2011Not classifiedZ-3125-2011PendingThe recall was initiated because Beckman Coulter has confirmed that the probe wash collar tubing may become stretched or trapped at the Sample Aspiration Module (SAM) during routine use of single-tube presentation station. When the…
Apr 14, 2011Not classifiedZ-2227-2011Software designThe recall was initiated because Beckman Coulter has confirmed the auto prune will not remove patient results if the patient record includes demographics where the results are less than 30 days old.
Feb 11, 2011Not classifiedZ-2172-2011PendingThe recall was initiated because: Specimen tubes may fall out of a Type A cassette (PN A53551, kit of 20) when the analyzer inverts the cassette in the mixing station, or when an operator manually inverts a cassette The impact: - Specimen…
Feb 11, 2011Not classifiedZ-2145-2011OtherThe recall was initiated because the DxH 800 may fail to meet the carryover specification for the WBC Differential (DIFF) when the WBC is greater than 70 X 10 to the 3rd power cells/micro L (70 X 10 to the 9th power cells/L). Beckman…
Feb 11, 2011Not classifiedZ-2157-2011PendingThe recall was initiated because: The handheld barcode scanner does not read barcode labels using NW 7 symbology. Impact: Specimens with NW7 labels cannot be read by the handheld scanner.
Feb 11, 2011Not classifiedZ-2148-2011OtherThe recall was initiated because: The DxH 800 may fail to meet the carryover specification for NRBC% when the WBC is greater than 50 X 10 to the 3rd power cells/micro L (50 X 10 to the 9th power cells/L) Beckman Coulter indicated the…
Oct 20, 2010Not classifiedZ-1978-2011OtherThere is a possibility that a specimen could be diluted if a user requests Diluent Dispense while specimens are being automatically processed in cassette presentation. The dilution may cause erroneous results. IMPACT: Erroneous results…
Oct 20, 2010Not classifiedZ-1980-2011OtherSharp edges on drawer slides within the DxH800 Floor Stand (Cabinet) may expose operators to physical injury while replacing reagents and waste containers. IMPACT: There is potential for exposure to blood borne pathogens such as HIV, HBV…
Oct 20, 2010Not classifiedZ-2128-2011OtherBody Fluids results may be incorrectly interpreted by the Laboratory Information System (LIS). Body Fluid results ~ 1,000 celis/uL are reported using the US numeric separators to demark thousandths places (comma, ",") and partial units…
Oct 20, 2010Not classifiedZ-2121-2011OtherRETIC results may be invalid when the instrument user interface allows reflex processing of a pre-diluted specimen in the RETIC test panel. Beckman Coulter has validated pre-diluted specimen processing in CBC panels. RETIC test panel…
Oct 20, 2010Not classifiedZ-1977-2011OtherThere is a possibility that a specimen could be diluted if a user requests Diluent Dispense while specimens are being automatically processed in cassette presentation. The dilution may cause erroneous results. IMPACT: Erroneous results…
Oct 20, 2010Not classifiedZ-2114-2011OtherFor Panels containing CBC: The RBC and HCT are improperly corrected and not flagged when there is a WBC Pump Failure and the WBC ~ 140x103/uL. For Body Fluid Panels: The RBC is improperly corrected and not flagged when there is a WBC Pump…
Mar 2, 2010Not classifiedZ-1656-2010Software designIssue 1: The Diagnostic procedure, Simulate-CDNR Cycle, is not functioning properly. Impact: There is a potential for inadequate sample preparation of samples following the use of this Diagnostic function impacting WBC, Differential…
Feb 12, 2010Not classifiedZ-2338-2010Software designThe recall was initiated after Beckman Coulter identified four (4) issues with the COULTER LH 500 Series System and the Unicel DxH 800 Coulter Cellular Analysis System: Issue 1: There is a potential for misidentification to occur when the…
Feb 4, 2010Not classifiedZ-1658-2010Software designIssue 1: The system will substitute or omit the characters # @ 1 \ ] ' ( I ) - when the instrument is configured in a language other than English, Chinese or Japanese. Example: Using the character @, a substitution in the string '123@456'…
Dec 8, 2009Not classifiedZ-1680-2010Software designThe Corrective Action was initiated after Beckman Coulter confirmed three issues related to host transmissions with the UniCel DxH 800. This information only affects instruments interfaced to a host system.
Jul 28, 2009Not classifiedZ-1905-2010OtherThe recall was initiated after Beckman Coulter confirmed an issue with the UniCel DxH 800 during in-house testing. It was determined that there is a potential for a false increase in platelet results, due to micro-bubbles. This situation…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 5.5 years on median.
  • 4 of the 8 ancestors have a recall of any class.
  • 7 other devices also cite its predicate K011342.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

The UniCel® DxH 800 Analyzer is a quantitative, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel® DxH 800 Analyzer provides:

· a Complete Blood Count (CBC), Leukocyte 5 Part Differential (Diff), Reticulocyte (Retic) and Nucleated Red Blood Cell (NRBC) on whole blood

· a Total Nucleated Count (TNC) and Red Cell Count (RBC) on Body Fluids (cerebrospinal, serous and synovial) (BF)

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K081930: Unicel DXH 800 Coulter Cellular Analysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K081930

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026