Coulter LH 750 Hematology Analyzer, Model 6605632

K011342510(k) TraditionalCleared Jul 9, 200168-day reviewClass II21 CFR 864.5220Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · GKZ Counter, Differential Cell

Cleared on Jul 9, 2001 after a 68-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 12 later devices descend from it. It has been recalled 15 times.

Decided
Jul 9, 2001
Review
68 days
Class
Class II
Product code
GKZ
Recalls
15
Predicates
0

12 later devices descend from it, 8 citing it directly.

Descendants, by year cleared 2003: 3 descendants cleared2004: 3 descendants cleared2002: 1 descendant cleared2006: 2 descendants cleared2008: 2 descendants cleared2005: 1 descendant cleared200020052010201520202025
  1. This deviceK011342 Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc. · 2001
  2. Descendant · cites it directly · 2 of its ownK032342 Coulter LH 750 Hematology AnalyzerBeckman Coulter, Inc. · 2003 · Recalled
  3. Descendant · cites it directly · 2 of its ownK022161 Modification to Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc. · 2002 · Recalled
  4. Descendant · cites it directly · 1 of its ownK061616 Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc. · 2006 · Recalled
  5. Descendant · cites it directly · 1 of its ownK041882 Flowcare SystemPointcare Technologies, Inc. · 2004
  6. Descendant · cites it directly · 1 of its ownK032000 Coulter LH 500 Hematology AnalyzerBeckman Coulter, Inc. · 2003 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2002: 12003: 32004: 32005: 12006: 22008: 2
2002high 3 (2003) · low 0 (2007)2026

Most influential descendants

Showing 5 of 12 · See all
  • K032342Coulter LH 750 Hematology AnalyzerBeckman Coulter, Inc. · 2003 · cites it directly2descendants
  • K022161Modification to Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc. · 2002 · cites it directly2descendants
  • K061616Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc. · 2006 · cites it directly1descendant
  • K041882Flowcare SystemPointcare Technologies, Inc. · 2004 · cites it directly1descendant
  • K032000Coulter LH 500 Hematology AnalyzerBeckman Coulter, Inc. · 2003 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K011342Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.2001—
+1K032342Coulter LH 750 Hematology AnalyzerBeckman Coulter, Inc.20031
+1K022161Modification to Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.20022
+1K061616Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc.20067
+1K041882Flowcare SystemPointcare Technologies, Inc.2004—
+1K032000Coulter LH 500 Hematology AnalyzerBeckman Coulter, Inc.20039

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200330
200430
200510
200620
200820
FIG. 1 — Lineage of K011342, 2001–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (12)

See all 12

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032342Coulter LH 750 Hematology AnalyzerBeckman Coulter, Inc.20032
K022161Modification to Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.20022
K061616Coulter LH 780 Hematology AnalyzerBeckman Coulter, Inc.20061
K041882Flowcare SystemPointcare Technologies, Inc.20041
K032000Coulter LH 500 Hematology AnalyzerBeckman Coulter, Inc.20031
K081930Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc.20080
K061574Coulter LH 750 Hematology Analyzer, Model 6605632Beckman Coulter, Inc.20060
K052905Modification to Coulter LH 750 Hematology AnalyzerBeckman Coulter, Inc.20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 26, 2017Class IIZ-3113-2017Labeling Change ControlBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
Jul 22, 2015Class IIZ-2481-2015Nonconforming Material/ComponentBeckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.
Jul 22, 2015Class IIZ-2480-2015Nonconforming Material/ComponentBeckman Coulter is recalling the LH 750 and LH780 Hematology Systems because the use of the diluent lots may eventually result in a compromised Hgb cuvette.
Nov 8, 2013Class IIZ-0470-2014Under Investigation by firmBeckman Coulter is recalling the LH 750 and LH 780 Hematology Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), white blood cell differential and/or…
Sep 24, 2013Class IIZ-0113-2014Component change controlThe check valves may fail on the Coulter LH 750 Hematology Analyzer (PN 6605632 and A85570) and Coulter LH 780 Hematology Analyzer (PN 723585 and A90728). A failure may result in an air or liquid leak at the connection to the Vacuum…
Dec 1, 2011Not classifiedZ-1242-2012Software designThe recall was initiated because Beckman Coulter has confirmed that the Auto Stop and Auto Transmit features for Quality Control become disabled after scanning assay values from Coulter 5C and Retic-C Cell Controls assay sheets using the…
Jan 31, 2011Not classifiedZ-1953-2011Software designThis recall was initiated because Beckman Coulter identified a problem with some date fields that are transmitted to the Laboratory Information System (LIS). Starting Jan 1, 2011, the date fields will transmit as a blank field to the LIS…
Apr 18, 2007Not classifiedZ-0312-2008Software designPatient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error…
Sep 12, 2006Not classifiedZ-0566-2007OtherThere is a potential for erroneous yet credible results for Body Fluids on the LH700 series hematology analyzers when a cassette label fails to read and the subsequent sample is cycled in Body Fluid mode.
Dec 15, 2005Not classifiedZ-0474-06OtherThere is a risk of sample misidentification when processing samples in the Manual aspiration mode.
Mar 31, 2004Not classifiedZ-0905-04OtherIf a customer enters a dilution factor for Sample B prior to the results from Sample A being transfered to the workstation, the predilute multiplication factor will be applied erroneously to the Sample A's results once they are received…
Jan 10, 2003Not classifiedZ-0541-03OtherBiohazard situation exists for users.
Dec 27, 2002Not classifiedZ-0669-03OtherIncorrect Hemoglobin result can be reported with software version 2A and higher.
Dec 10, 2002Not classifiedZ-0571-03OtherDevice design allows for cross contamination issue to occur, possibly affecting assays and results reports.
Nov 8, 2002Not classifiedZ-0415-03OtherSoftware vote out logic changed. Erroneous white blood cell counts.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K032342 (Beckman Coulter, Inc., 2003) has gone furthest, with 2 descendants of its own.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—
Cite this page

510(k) Report. “K011342: Coulter LH 750 Hematology Analyzer, Model 6605632.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K011342

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026