510(k) Report

510(k) · Cleared

Cleared

Forsure One Step Dip & Read (X) Drug Screen Test

K061617510(k) TraditionalCleared Nov 22, 2006166-day reviewClass II21 CFR 862.3100

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tianjin New Bay Bioresearch Co., Ltd. · DKZ Enzyme Immunoassay, Amphetamine

Cleared on Nov 22, 2006 as substantially equivalent to K050540 (Ucp Biosciences, Inc., 2005) and 1 other predicate after a 166-day review. Its family tree reaches back 2 generations to 2001.

Decided
Nov 22, 2006
Review
166 days
Class
Class II
Product code
DKZ
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K004034 (2001). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 2 generations backK004034 Monitect Multiple Drug Screen Tests, Model BC12,BC14,TC10,PC11, and…Branan Medical Corp. · 2001
  2. 1 generation backK052882 Forsure Rapid One Step Multiple (X) Abuse Drug Screen Test Cup DeviceTianjin New Bay Bioresearch Co., Ltd. · 2006
  3. This deviceK061617 Forsure One Step Dip & Read (X) Drug Screen TestTianjin New Bay Bioresearch Co., Ltd. · 2006
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K004034Monitect Multiple Drug Screen Tests, Model BC12,BC14,TC10,PC11, and PC12Branan Medical Corp.2001—
−1K052882Forsure Rapid One Step Multiple (X) Abuse Drug Screen Test Cup DeviceTianjin New Bay Bioresearch Co., Ltd.2006—
0 (this device)K061617Forsure One Step Dip & Read (X) Drug Screen TestTianjin New Bay Bioresearch Co., Ltd.2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210
200510
FIG. 1 — Lineage of K061617, 2001–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DKZ device

  1. #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050540Ucp Rapid Drug Screening Tests
Summary page 2
…II Predicate Devices: Forsure One Step Dip & Read Drug Screen Test Device is substantially equivalent to UCP Bioscience Rapid Drug Screening Test Strips cleared by FDA (K050540), or TNB ForsUre One Step Drug Screen Test Cup Device cleared by FDA (K052882) and GC/MS for its stated intended use. Device Des…
Ucp Biosciences, Inc.2005—
*K052882Forsure Rapid One Step Multiple (X) Abuse Drug Screen Test Cup Device
Summary page 2
…substantially equivalent to UCP Bioscience Rapid Drug Screening Test Strips cleared by FDA (K050540), or TNB ForsUre One Step Drug Screen Test Cup Device cleared by FDA (K052882) and GC/MS for its stated intended use. Device Description: Forsure One Step Dip & Read Drug Screen Test consists of a or Multi…
Tianjin New Bay Bioresearch Co., Ltd.2006—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 2 companies; a new generation was cleared every 4.9 years on median.
  • None of the 4 ancestors has a recall.

How does its category compare?

All DKZ devices

DKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DKZ

1in 2024

2008: 22009: 22010: 42011: 62012: 52013: 82014: 82015: 52016: 52018: 32019: 32020: 12023: 22024: 1
2007high 8 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DKZ

180in 2019

2010: 233 days median2011: 123 days median2012: 55 days median2013: 103 days median2014: 128 days median2015: 107 days median2016: 215 days median2018: 128 days median2019: 180 days median
2007high 233 (2010) · low 55 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DKZ

1in 2017

2007: 12008: 32009: 52011: 42013: 12014: 32017: 1
2007high 5 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DKZ

2in 2025

2008: 122009: 12010: 12011: 22013: 12014: 92015: 82016: 82021: 32022: 42023: 72024: 12025: 2
2007high 12 (2008) · low 0 (2007)2025

MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20082—312
20092—51
2010423301
2011612342
201255500
2013810311
2014812839
2015510708
2016521508
20170—10
2018312800
2019318000
20201—00
20210—03
20220—04
20232—07
20241—01
20250—02
20260—0—

What is it indicated for?

Forsure One Step Dip & Read Drug Screen Test is a prescription assay intended for professional use in central laboratories only. It provides qualitative screening results for Amphetamine (AMP), Methamphetamine (MET), Benzodiazepine (BZD), Barbiturate (BAR), Cocaine (COC), Cannabinoids (THC), Opiates (OPI), Phencyclidine (PCP), Methadone (MAD), Oxycodone (OXY), Tricyclic Antidepressant (TCA) and Propoxyphene (PPX) in human urine at the following cutoff…

Read the full statement (166 words)

Forsure One Step Dip & Read Drug Screen Test is a prescription assay intended for professional use in central laboratories only. It provides qualitative screening results for Amphetamine (AMP), Methamphetamine (MET), Benzodiazepine (BZD), Barbiturate (BAR), Cocaine (COC), Cannabinoids (THC), Opiates (OPI), Phencyclidine (PCP), Methadone (MAD), Oxycodone (OXY), Tricyclic Antidepressant (TCA) and Propoxyphene (PPX) in human urine at the following cutoff levels:

Test Calibrator Cutoff

Amphetamine D-Amphetamine 1000 ng/ml

Methamphetamine D-Methamphetamine 1000 ng/ml

Benzodiazepine Oxazepam 300 ng/ml

Barbiturate Secobarbital 300 ng/ml

Cocaine Benzoylecgonine 300 ng/ml

Cannabinoids 1 1-nor-A9-THC-9 COOH 50 ng/ml

Opiates Morphine 2000 ng/ml

Phencyclidine Phencyclidine 25 ng/ml

Methadone Methadone 300 ng/ml

Oxycodone Oxycodone 100 ng/ml

Tricyclic Antidepressant Nortriptyline 1000 ng/ml

Propoxyphene Propoxyphene 300 ng/ml

This assay provides only a preliminary result. Clinical consideration and professional judgment should be applied to any drugs of abuse test result, particularly in evaluating a preliminary positive. To obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectroscopy (GC/MS) is the recommended confirmatory method.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061617: Forsure One Step Dip & Read (X) Drug Screen Test.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061617

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026