510(k) · Cleared
ClearedXenta Drug Screen Cup, Xenta Drug Screen Dipcard
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Xenta Biomedical Science Co., Ltd. · DKZ Enzyme Immunoassay, Amphetamine
Cleared on May 19, 2023 as substantially equivalent to K122809 (Advin Biotech, 2013) after a 28-day review. Its family tree reaches back 6 generations to 1995.
- Decided
- May 19, 2023
- Review
- 28 days
- Class
- Class II
- Product code
- DKZ
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to K945610 (1995). 27 ancestors in all. No later device cites it as a predicate yet.
- Root · 6 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
- 5 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
- 4 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
- 3 generations backK031759 Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon…ACON Laboratories, Inc. · 2003 · Recalled
- 2 generations backK061718 Innovacon Spectrum II Test CardInnovacon, Inc. · 2006 · Recalled
- 1 generation backK122809 Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech · 2013
- This deviceK231137 Xenta Drug Screen Cup, Xenta Drug Screen DipcardXenta Biomedical Science Co., Ltd. · 2023
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K945610 | Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-Amp | Princeton BioMeditech Corp. | 1995 | — |
| −5 | K011673 | Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test… | ACON Laboratories, Inc. | 2001 | — |
| −4 | K020313 | Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug… | Acon Laboratories Co. | 2002 | — |
| −3 | K031759 | Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum… | ACON Laboratories, Inc. | 2003 | 1 |
| −2 | K061718 | Innovacon Spectrum II Test Card | Innovacon, Inc. | 2006 | 6 |
| −1 | K122809 | Advin Multi-Drug Screen Test Cassette, Dip Card and Cup | Advin Biotech | 2013 | — |
| 0 (this device) | K231137 | Xenta Drug Screen Cup, Xenta Drug Screen Dipcard | Xenta Biomedical Science Co., Ltd. | 2023 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 2 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 7 | 0 |
| 2002 | 2 | 0 |
| 2003 | 2 | 0 |
| 2004 | 3 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DKZ device
- #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K122809 | Advin Multi-Drug Screen Test Cassette, Dip Card and CupSummary page 4…Unclassified Enzyme Immunoassay Phencyclidine Device class: ClassⅡ Panel: Toxicology Product code: DIS,JXM,DKZ,DJR,DJG,LCM 4. The Predicate Devices K122809 Advin Multi-Drug Screen Test Dip Card Advin Multi-Drug Screen Test Cup Advin Multi-Dru… | 2013 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 28 years and passes through 6 companies; a new generation was cleared every 6.3 years on median.
- 2 of the 27 ancestors have a recall of any class.
- 11 other devices also cite its predicate K122809.
How does its category compare?
All DKZ devicesDKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · DKZ
1in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · DKZ
180in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DKZ
1in 2017
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DKZ
2in 2025
MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 1 | 0 |
| 2008 | 2 | — | 3 | 12 |
| 2009 | 2 | — | 5 | 1 |
| 2010 | 4 | 233 | 0 | 1 |
| 2011 | 6 | 123 | 4 | 2 |
| 2012 | 5 | 55 | 0 | 0 |
| 2013 | 8 | 103 | 1 | 1 |
| 2014 | 8 | 128 | 3 | 9 |
| 2015 | 5 | 107 | 0 | 8 |
| 2016 | 5 | 215 | 0 | 8 |
| 2017 | 0 | — | 1 | 0 |
| 2018 | 3 | 128 | 0 | 0 |
| 2019 | 3 | 180 | 0 | 0 |
| 2020 | 1 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 3 |
| 2022 | 0 | — | 0 | 4 |
| 2023 | 2 | — | 0 | 7 |
| 2024 | 1 | — | 0 | 1 |
| 2025 | 0 | — | 0 | 2 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
Xenta Drug Screen Cup and Xenta Drug Screen Dipcard are lateral flow chromatographic immunoassays designed to qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations: Test Calibrator Cut-off level Barbiturates (BAR) Secobarbital 300 ng/mL Benzodiazepines (BZO) Oxazepam 300 ng/mL Amphetamine (AMP) D-Amphetamine 1000 ng/mL Methadone (MTD) Methadone 300 ng/mL Oxycodone (OXY) Oxycodone 100 ng/mL…
Read the full statement (159 words)
Xenta Drug Screen Cup and Xenta Drug Screen Dipcard are lateral flow chromatographic immunoassays designed to
qualitatively detect the presence of drugs and drug metabolites in human urine at the following cut-off concentrations:
Test Calibrator Cut-off level
Barbiturates (BAR) Secobarbital 300 ng/mL
Benzodiazepines (BZO) Oxazepam 300 ng/mL
Amphetamine (AMP) D-Amphetamine 1000 ng/mL
Methadone (MTD) Methadone 300 ng/mL
Oxycodone (OXY) Oxycodone 100 ng/mL
Phencyclidine (PCP) Phencyclidine 25 ng/mL
The tests contain two formats: 1) Test Cup and 2) Test Dipcard. The tests may be configured as single drug tests or
multiple drug tests in any combination of the drug analytes listed in the table above. These tests are intended for in vitro
diagnostics use. They are intended for prescription use.
The assays provide only a preliminary analytical test result. Gas Chromatography/Mass spectrometry (GC/MS) is the
preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse
test result, particularly when preliminary positive results are indicated.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K231137 · printed Nov 6, 2026
Cite this page
510(k) Report. “K231137: Xenta Drug Screen Cup, Xenta Drug Screen Dipcard.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K231137