510(k) Report

510(k) · Cleared

Recalled · 2011

Gem Premier 4000 with Iqm, Model 4000; Gem CVP 1 and 2 with CO-OX; Gem CVP 3 and 4 Hematocrit

K061974510(k) TraditionalCleared Sep 15, 200665-day reviewClass II21 CFR 862.1120Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Instrumentation Laboratory CO · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Sep 15, 2006 as substantially equivalent to K052121 (Instrumentation Laboratory CO, 2005) and 2 other predicates after a 65-day review. Its family tree reaches back 5 generations to 1980. 14 later devices descend from it. It has been recalled 4 times.

Decided
Sep 15, 2006
Review
65 days
Class
Class II
Product code
CHL
Recalls
4
Predicates
3

Its longest line of descent runs back 5 generations to K801848 (1980). 10 ancestors in all. 14 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1996: 1 other ancestor1994: 1 other ancestor1989: 1 other ancestor2001: 1 other ancestor2011: 1 descendant cleared2012: 1 descendant cleared2010: 1 descendant cleared2016: 4 descendants cleared2014: 1 descendant cleared2017: 1 descendant cleared2019: 2 descendants cleared2021: 1 descendant cleared2023: 2 descendants clearedK801848 · ICA1 Ionized Calcium Analyzer · The London Co. · 1980K963800 · IL Synthesis · Instrumentation Laboratory CO · 1996K992834 · Gem Premier 3000 · Instrumentation Laboratory CO · 1999 · recalledK022158 · Iqm (Intelligent Quality Management) on the Gem Premier 3000, CVP (Calibration Validation Product) · Instrumentation Laboratory CO · 2002K052121 · Gem Premier 3000 · Instrumentation Laboratory CO · 2005 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 2 descendants3 generations forward: 7 descendants4 generations forward: 2 descendants K052121 · Gem Premier 3000 · Instrumentation Laboratory CO · 2005 · recalledK022158 · Iqm (Intelligent Quality Management) on the Gem Premier 3000, CVP (Calibration Validation Product) · Instrumentation Laboratory CO · 2002K992834 · Gem Premier 3000 · Instrumentation Laboratory CO · 1999 · recalledK963800 · IL Synthesis · Instrumentation Laboratory CO · 1996K801848 · ICA1 Ionized Calcium Analyzer · The London Co. · 1980−50
  1. Root · 5 generations backK801848 ICA1 Ionized Calcium AnalyzerThe London Co. · 1980
  2. 4 generations backK963800 IL SynthesisInstrumentation Laboratory CO · 1996
  3. 3 generations backK992834 Gem Premier 3000Instrumentation Laboratory CO · 1999 · Recalled
  4. 2 generations backK022158 Iqm (Intelligent Quality Management) on the Gem Premier 3000, CVP…Instrumentation Laboratory CO · 2002
  5. 1 generation backK052121 Gem Premier 3000Instrumentation Laboratory CO · 2005 · Recalled
  6. This deviceK061974 Gem Premier 4000 with Iqm, Model 4000; Gem CVP 1 and 2 with CO-OX…Instrumentation Laboratory CO · 2006
  7. Descendant · cites it directly · 12 of its ownK093623 Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter)…Instrumentation Laboratory CO · 2010 · Recalled
  8. Descendant · cites it directly · 11 of its ownK112995 Gem Premier 4000Instrumentation Laboratory CO · 2012
  9. Descendant · 2 generations on · 10 of its ownK133407 Gem Premier 4000Instrumentation Laboratory CO · 2014
  10. Descendant · 3 generations on · 2 of its ownK173403 GEM Premier 5000Instrumentation Laboratory CO · 2017
  11. Descendant · 2 generations on · 2 of its ownK160225 GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride…Instrumentation Laboratory CO · 2016

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12012: 12014: 12016: 42017: 12019: 22021: 12023: 2
2010high 4 (2016) · low 0 (2013)2026

Most influential descendants

Showing 5 of 14 · See all
  • K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2 (Addition…Instrumentation Laboratory CO · 2010 · cites it directly12descendants
  • K112995Gem Premier 4000Instrumentation Laboratory CO · 2012 · cites it directly11descendants
  • K133407Gem Premier 4000Instrumentation Laboratory CO · 2014 · 2 generations on10descendants
  • K173403GEM Premier 5000Instrumentation Laboratory CO · 2017 · 3 generations on2descendants
  • K160225GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride, Ionized Calcium), GEM…Instrumentation Laboratory CO · 2016 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K801848ICA1 Ionized Calcium AnalyzerThe London Co.1980—
−4K963800IL SynthesisInstrumentation Laboratory CO1996—
−3K992834Gem Premier 3000Instrumentation Laboratory CO19995
−2K022158Iqm (Intelligent Quality Management) on the Gem Premier 3000, CVP (Calibration…Instrumentation Laboratory CO2002—
−1K052121Gem Premier 3000Instrumentation Laboratory CO20052
0 (this device)K061974Gem Premier 4000 with Iqm, Model 4000; Gem CVP 1 and 2 with CO-OX; Gem CVP 3…Instrumentation Laboratory CO2006—
+1K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2…Instrumentation Laboratory CO20105
+1K112995Gem Premier 4000Instrumentation Laboratory CO2012—
+2K133407Gem Premier 4000Instrumentation Laboratory CO2014—
+3K173403GEM Premier 5000Instrumentation Laboratory CO2017—
+2K160225GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride, Ionized…Instrumentation Laboratory CO2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199410
199510
199610
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201210
201410
201640
201710
201920
202110
202320
FIG. 1 — Lineage of K061974, 1980–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.1 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052121Gem Premier 3000
Summary page 2
…Analytes pH, blood gases, electrolytes (except GEM 3000 with iQM and GEM CVP K052121 chloride), glucose, lactate, hematocrit IL Synthesis with traditional QC K963800 Chloride IL 682 CO-Oxim…
Instrumentation Laboratory CO20052
*K963800IL Synthesis
Summary page 2
…, electrolytes (except GEM 3000 with iQM and GEM CVP K052121 chloride), glucose, lactate, hematocrit IL Synthesis with traditional QC K963800 Chloride IL 682 CO-Oximeter with traditional QC K945677 CO-Oximeter parameters Device Description and Indicati…
Instrumentation Laboratory CO1996—
*K945677IL682, CO-Oximeter
Summary page 2
…K052121 chloride), glucose, lactate, hematocrit IL Synthesis with traditional QC K963800 Chloride IL 682 CO-Oximeter with traditional QC K945677 CO-Oximeter parameters Device Description and Indications for Use: The GEM Premier 4000 is a portable critical care system…
Instrumentation Laboratory CO1995—

Also cleared against K052121: 1 other device →

What descends from it? (14)

See all 14

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2 (Addition…Instrumentation Laboratory CO201012
K112995Gem Premier 4000Instrumentation Laboratory CO201211
K093280Opti Cca-TS, B-Lac Cassette Model BP7561Opti Medical Systems, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 31, 2011Not classifiedZ-2806-2011Nonconforming Material/ComponentK+ results on the GEM Premier 4000 are too low when compared to a reference analyzer, with biases exceeding allowable error claim of a 0.5 moUL
May 31, 2011Not classifiedZ-2805-2011Nonconforming Material/ComponentK+ results on the GEM Premier 4000 are too low when compared to a reference analyzer, with biases exceeding allowable error claim of a 0.5 moUL
May 31, 2011Not classifiedZ-2803-2011Nonconforming Material/ComponentK+ results on the GEM Premier 4000 are too low when compared to a reference analyzer, with biases exceeding allowable error claim of a 0.5 moUL
Mar 27, 2007Not classifiedZ-0778-2007OtherInstrument in Micro Sampling Mode failed to aspirate and no error code was generated and an incorrect patient report could result

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 2 companies; a new generation was cleared every 3.0 years on median.
  • 2 of the 10 ancestors have a recall of any class.
  • Of its descendants, K093623 (Instrumentation Laboratory CO, 2010) has gone furthest, with 12 descendants of its own.
  • 1 other device also cites its predicate K052121.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, PO2, Na +, K +, CI-, Ca ++, glucose, lactate, hematocrit and CO-Oximetry (tHb, O 2Hb, COHb, MetHb, HHb) parameters. These parameters, along with derived parameters, aid in the…

Read the full statement (252 words)

The GEM Premier 4000 is a portable critical care system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory. The instrument provides quantitative measurements of pH, pCO2, PO2, Na +, K +, CI-, Ca ++, glucose, lactate, hematocrit and CO-Oximetry (tHb, O 2Hb, COHb, MetHb, HHb) parameters. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, electrolyte and metabolite balance and oxygen delivery capacity.

Intelligent Quality Management (iQM) is used as the quality control and assessment system for the GEM Premier 4000 system. iQM is an active quality process control program designed to provide continuous monitoring of the analytical process with real-time, automatic error detection, automatic correction of the system and automatic documentation of all corrective actions, replacing the use of traditional external quality controls.

As part of this program, GEM CVP 1 and 2 (Calibration Valuation Product) with CO-Ox and GEM CVP 3 and 4 Hematocrit are external solutions intended to complete the calibration process and final accuracy assessment of the iQM cartridge calibration after initial warm-up. The reported values for the four levels of GEM CVP (two levels for pH, blood gases, electrolytes, metabolites and CO-Oximetry; two levels for hematocrit) must meet specifications before the iQM cartridge can be used for patient sample measurements. Once the cartridge calibration is verified, the internal iQM program monitors the status of the system during the cartridge use life.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061974: Gem Premier 4000 with Iqm, Model 4000; Gem CVP 1 and 2 with CO-OX; Gem CVP 3 and 4 Hematocrit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K061974

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026