510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Vitros Chemistry Products Amph Reagent; Calibrator Kit 26; FS Calibrator 1; Dat Performance Verifiers

K062077510(k) TraditionalCleared Nov 22, 2006124-day reviewClass II21 CFR 862.3100Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · DKZ Enzyme Immunoassay, Amphetamine

Cleared on Nov 22, 2006 as substantially equivalent to K031004 (Dade Behring, Inc., 2003) and 1 other predicate after a 124-day review. Its family tree reaches back 2 generations to 2000. 1 later device descends from it. It has been recalled 5 times.

Decided
Nov 22, 2006
Review
124 days
Class
Class II
Product code
DKZ
Recalls
5
Predicates
2

Its longest line of descent runs back 2 generations to K993982 (2000). 3 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2015: 1 descendant clearedK993982 · Syva Emit II Plus Amphetamine/Methamphetamine Assay 9C029UL/9C129UL · Syva Co. · 2000K031004 · Syva Emit II Plus Amphetamines Assay · Dade Behring, Inc. · 2003200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant K031004 · Syva Emit II Plus Amphetamines Assay · Dade Behring, Inc. · 2003K993982 · Syva Emit II Plus Amphetamine/Methamphetamine Assay 9C029UL/9C129UL · Syva Co. · 20000
  1. Root · 2 generations backK993982 Syva Emit II Plus Amphetamine/Methamphetamine Assay 9C029UL/9C129ULSyva Co. · 2000
  2. 1 generation backK031004 Syva Emit II Plus Amphetamines AssayDade Behring, Inc. · 2003
  3. This deviceK062077 Vitros Chemistry Products Amph Reagent; Calibrator Kit 26; FS…Ortho-Clinical Diagnostics, Inc. · 2006
  4. Descendant · cites it directly · 0 of its ownK143500 Immunalysis Amphetamine Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2015

Most influential descendants

Showing 1 of 1 · See all
  • K143500Immunalysis Amphetamine Urine Enzyme Immunoassay, Immunalysis Amphetamine Urine…Immunalysis Corporation · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K993982Syva Emit II Plus Amphetamine/Methamphetamine Assay 9C029UL/9C129ULSyva Co.2000—
−1K031004Syva Emit II Plus Amphetamines AssayDade Behring, Inc.2003—
0 (this device)K062077Vitros Chemistry Products Amph Reagent; Calibrator Kit 26; FS Calibrator 1…Ortho-Clinical Diagnostics, Inc.2006—
+1K143500Immunalysis Amphetamine Urine Enzyme Immunoassay, Immunalysis Amphetamine…Immunalysis Corporation2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
FIG. 1 — Lineage of K062077, 2000–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DKZ device

  1. #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031004Syva Emit II Plus Amphetamines Assay
Summary page 3
…ROS Chemistry Products to predicate DAT Performance Verifiers are substantially equivalent to the Syva® EMIT LI Plus devices: Amphetamines assay (K031004) and Bio-Rad Liquichecku" Urine Toxicology Controls (K022707) predicate devices which were cleared by the…
Dade Behring, Inc.2003—
*K022707Liquichek Urine Toxicology Control Level S1E, Model 438; Level S2E, Model 439
Summary page 3
…ially equivalent to the Syva® EMIT LI Plus devices: Amphetamines assay (K031004) and Bio-Rad Liquichecku" Urine Toxicology Controls (K022707) predicate devices which were cleared by the FDA for IVD use.…
Bio-Rad2002—

Also cleared against K031004: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K143500Immunalysis Amphetamine Urine Enzyme Immunoassay, Immunalysis Amphetamine Urine…Immunalysis Corporation20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 22, 2017Class IIZ-0911-2018Under Investigation by firmThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
Sep 22, 2017Class IIZ-0913-2018Under Investigation by firmThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
Sep 22, 2017Class IIZ-0912-2018Under Investigation by firmThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
Sep 22, 2017Class IIZ-0914-2018Under Investigation by firmThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
Sep 22, 2017Class IIZ-0915-2018Under Investigation by firmThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 3.5 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K143500 (Immunalysis Corporation, 2015) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K031004.

How does its category compare?

All DKZ devices

DKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DKZ

1in 2024

2008: 22009: 22010: 42011: 62012: 52013: 82014: 82015: 52016: 52018: 32019: 32020: 12023: 22024: 1
2007high 8 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DKZ

180in 2019

2010: 233 days median2011: 123 days median2012: 55 days median2013: 103 days median2014: 128 days median2015: 107 days median2016: 215 days median2018: 128 days median2019: 180 days median
2007high 233 (2010) · low 55 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DKZ

1in 2017

2007: 12008: 32009: 52011: 42013: 12014: 32017: 1
2007high 5 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DKZ

2in 2025

2008: 122009: 12010: 12011: 22013: 12014: 92015: 82016: 82021: 32022: 42023: 72024: 12025: 2
2007high 12 (2008) · low 0 (2007)2025

MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20082—312
20092—51
2010423301
2011612342
201255500
2013810311
2014812839
2015510708
2016521508
20170—10
2018312800
2019318000
20201—00
20210—03
20220—04
20232—07
20241—01
20250—02
20260—0—

What is it indicated for?

For in vitro diagnostic use only. VITROS Chemistry Products AMPH Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of amphetamines (AMPH) in human urine using a cutoff of either 500 or 1000 ng/mL. Measurements obtained with the VITROS AMPH method are used in the diagnosis and treatment of amphetamines use or overdose. The VITROS Chemistry Products AMPH assay is intended for use by professional…

Read the full statement (218 words)

For in vitro diagnostic use only. VITROS Chemistry Products AMPH Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of amphetamines (AMPH) in human urine using a cutoff of either 500 or 1000 ng/mL. Measurements obtained with the VITROS AMPH method are used in the diagnosis and treatment of amphetamines use or overdose.

The VITROS Chemistry Products AMPH assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of drugs of abuse.

For in vitro diagnostic use only. VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems.

For in vitro diagnostic use only. VITROS Chemistry Products DAT Performance Verifiers are assayed controls used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062077: Vitros Chemistry Products Amph Reagent; Calibrator Kit 26; FS Calibrator 1; Dat Performance Verifiers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K062077

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026