510(k) Report

510(k) · Cleared

Recalled · 2008

Lifepak 20 Defibrillator/ Monitor

K063119510(k) SpecialCleared Dec 22, 200671-day reviewClass III21 CFR 870.5310Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Emergency Response Systems, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 22, 2006 as substantially equivalent to K033275 (Medtronic Physio-Control Corp., 2003) and 1 other predicate after a 71-day review. Its family tree reaches back 6 generations to 1993. 1 of the 25 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 5 times.

Decided
Dec 22, 2006
Review
71 days
Class
Class III
Product code
MKJ
Recalls
5
Predicates
2

Its longest line of descent runs back 6 generations to K921669 (1993). 22 ancestors in all, the oldest K802288 (1980). 2 later devices descend from it, 2 citing it directly. 1 of the 25 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 3 other ancestors1999: 1 other ancestor1980: 1 other ancestor1996: 2 other ancestors1995: 4 other ancestors1992: 1 other ancestor1991: 1 other ancestor1990: 1 other ancestor1984: 1 other ancestor1994: 1 other ancestor2013: 1 descendant cleared, 1 with a Class I recall12008: 1 descendant clearedK921669 · Marquette SL Series Transport Remote Acquisition · Marquette Electronics, Inc. · 1993K964750 · Eagle 4000 Patient Monitor · Marquette Electronics, Inc. · 1997K973486 · Life Pak 12 · Physio-Control Corp. · 1998 · recalledK991910 · Lifepak 12 Biphasic · Physio-Control Corp. · 1999 · recalledK012274 · Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/Pacemaker · Medtronic Physio-Control Corp. · 2002 · recalledK033275 · Lifepak 12, 20, 500, CR Plus · Medtronic Physio-Control Corp. · 2003 · recalled19801990200020102020
  1. Root · 6 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 5 generations backK964750 Eagle 4000 Patient MonitorMarquette Electronics, Inc. · 1997
  3. 4 generations backK973486 Life Pak 12Physio-Control Corp. · 1998 · Recalled
  4. 3 generations backK991910 Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · Recalled
  5. 2 generations backK012274 Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · Recalled
  6. 1 generation backK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled
  7. This deviceK063119 Lifepak 20 Defibrillator/ MonitorMedtronic Emergency Response Systems, Inc. · 2006
  8. Descendant · cites it directly · 1 of its ownK073089 Lifepak 20EPhysio-Control, Inc. · 2008 · Recalled
  9. Descendant · cites it directly · 0 of its ownK130454 Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorPhysio-Control, Inc. · 2013 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K073089Lifepak 20EPhysio-Control, Inc. · 2008 · cites it directly1descendant
  • K130454Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorPhysio-Control, Inc. · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K921669Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc.1993—
−5K964750Eagle 4000 Patient MonitorMarquette Electronics, Inc.1997—
−4K973486Life Pak 12Physio-Control Corp.19983
−3K991910Lifepak 12 BiphasicPhysio-Control Corp.19993
−2K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp.200217
−1K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031
0 (this device)K063119Lifepak 20 Defibrillator/ MonitorMedtronic Emergency Response Systems, Inc.2006—
+1K073089Lifepak 20EPhysio-Control, Inc.20081
+1K130454Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorPhysio-Control, Inc.2013Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
199010
199110
199210
199410
199540
199620
199910
200130

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201311
FIG. 1 — Lineage of K063119, 1993–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033275Lifepak 12, 20, 500, CR Plus
Summary page 2
…res and functions of the modified LIFEPAK 20 defibrillator/monitor are substantially equivalent to the previously cleared LIFEPAK 20 defibrillator/monitor (K033275, K012274). Description: The LIFEPAK 20 defibrillator/monitor was created especially for hospitals and clinics for…
Medtronic Physio-Control Corp.20031
*K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/Pacemaker
Summary page 2
…unctions of the modified LIFEPAK 20 defibrillator/monitor are substantially equivalent to the previously cleared LIFEPAK 20 defibrillator/monitor (K033275, K012274). Description: The LIFEPAK 20 defibrillator/monitor was created especially for hospitals and clinics for use on "…
Medtronic Physio-Control Corp.200217

Also cleared against K033275: 9 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K073089Lifepak 20EPhysio-Control, Inc.20081
K130454Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/Monitor Class IPhysio-Control, Inc.20130

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 13, 2008Not classifiedZ-2010-2008Device DesignA thicker keypad that may prevent the door from fully latching closed. If the door is not fully closed, there is a potential that the defibrillator will not automatically change from automatic external defibrillation (AED) mode to manual…
May 13, 2008Not classifiedZ-2009-2008Device DesignA thicker keypad may prevent the door from fully latching closed. If the door is not fully closed, there is a potential that the defibrillator will not automatically change from automatic external defibrillation (AED) mode to manual mode…
May 13, 2008Not classifiedZ-2011-2008Device DesignA thicker keypad that may prevent the door from fully latching closed. If the door is not fully closed, there is a potential that the defibrillator will not automatically change from automatic external defibrillation (AED) mode to manual…
Apr 16, 2008Not classifiedZ-1906-2008Software designThe LIFEPAK 20e defibrillator / monitors with software versions 058 have an increase in likelihood for an incorrect Shock Advisory Algorithm (SAS) decision if the Auto Analyze setting in AED mode is On, leading to incorrectly render of…
Apr 16, 2008Not classifiedZ-1905-2008Software designThe LIFEPAK 20 defibrillator / monitors with software versions 048, 052 and 054 have an increase in likelihood for an incorrect Shock Advisory Algorithm (SAS) decision if the Auto Analyze setting in AED mode is On, leading to incorrectly…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130454Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorPhysio-Control, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.4 years on median.
  • 10 of the 22 ancestors have a recall of any class.
  • Of its descendants, K073089 (Physio-Control, Inc., 2008) has gone furthest, with 1 descendant of its own.
  • 9 other devices also cite its predicate K033275.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

Manual Defibrillation: Indications: Defibrillation is a recognized means of terminating certain fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia. The biphasic defibrillation waveform used in this device…

Read the full statement (199 words)

Manual Defibrillation:

Indications: Defibrillation is a recognized means of terminating certain fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.

The biphasic defibrillation waveform used in this device has only been clinically tested on adults; it has not been tested on pediatric patients.

Contraindications: Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA) such as idioventricular escape rhythms, and in the treatment of asystole.

Automated External Defibrillation:

Indications: The AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 20 defibrillator/monitor is not intended for use on pediatric patients less than 8 years old.

Noninvasive Pacing

Indications: Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.

Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.

Pulse Oximetry

Indications: The pulse oximeter is indicated for use in any patient who is at risk of developing hypoxemia.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063119: Lifepak 20 Defibrillator/ Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K063119

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.