510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

45L Core Insufflator, Model F114

K063367510(k) TraditionalCleared May 23, 2007197-day reviewClass II21 CFR 884.1730Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

W.O.M. World of Medicine AG · HIF Insufflator, Laparoscopic

Cleared on May 23, 2007 as substantially equivalent to K060723 (W.O.M. World of Medicine AG, 2006) and 3 other predicates after a 197-day review. Its family tree reaches back 4 generations to 1997. 2 of the 59 devices in its family, counting itself, have a Class I recall. 44 later devices descend from it. It has been recalled 4 times.

Decided
May 23, 2007
Review
197 days
Class
Class II
Product code
HIF
Recalls
4
Predicates
4

Its longest line of descent runs back 4 generations to K964171 (1997). 14 ancestors in all, the oldest K904993 (1991). 44 later devices descend from it, 7 citing it directly. 2 of the 59 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 2 other ancestors2001: 1 other ancestor1996: 2 other ancestors2004: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor1994: 1 other ancestor1991: 1 other ancestor2016: 1 descendant cleared2011: 2 descendants cleared, 1 with a Class I recall12012: 4 descendants cleared, 1 with a Class I recall12019: 2 descendants cleared2023: 4 descendants cleared2017: 3 descendants cleared2013: 3 descendants cleared2015: 1 descendant cleared2021: 1 descendant cleared2014: 1 descendant cleared2025: 7 descendants cleared2018: 1 descendant cleared2020: 3 descendants cleared2024: 4 descendants cleared2026: 7 descendants clearedK964171 · Vasoview Balloon Dissection System · Origin Medsystems, Inc. · 1997K981700 · Vasoview Dissection/Vessel Harvesting System · Origin Medsystems, Inc. · 1998K030512 · Guidant Vasoview Endoscopic Vessel Harvesting System · Guidant Corp. · 2003 · recalledK060723 · 40 L High Flow Insufflator F113 · W.O.M. World of Medicine AG · 20061990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 2 other ancestors3 generations back: 5 other ancestors 1 generation forward: 7 descendants, 1 with a Class I recall12 generations forward: 10 descendants, 1 with a Class I recall13 generations forward: 7 descendants4 generations forward: 4 descendants5 generations forward: 9 descendants6 generations forward: 6 descendants7 generations forward: 1 descendant K060723 · 40 L High Flow Insufflator F113 · W.O.M. World of Medicine AG · 2006K030512 · Guidant Vasoview Endoscopic Vessel Harvesting System · Guidant Corp. · 2003 · recalledK981700 · Vasoview Dissection/Vessel Harvesting System · Origin Medsystems, Inc. · 1998K964171 · Vasoview Balloon Dissection System · Origin Medsystems, Inc. · 19970+5
  1. Root · 4 generations backK964171 Vasoview Balloon Dissection SystemOrigin Medsystems, Inc. · 1997
  2. 3 generations backK981700 Vasoview Dissection/Vessel Harvesting SystemOrigin Medsystems, Inc. · 1998
  3. 2 generations backK030512 Guidant Vasoview Endoscopic Vessel Harvesting SystemGuidant Corp. · 2003 · Recalled
  4. 1 generation backK060723 40 L High Flow Insufflator F113W.O.M. World of Medicine AG · 2006
  5. This deviceK063367 45L Core Insufflator, Model F114W.O.M. World of Medicine AG · 2007
  6. Descendant · cites it directly · 35 of its ownK103692 Sugiquest Airsealoptical Trocar & Cannula System with Integrated…Surgiquest, Inc. · 2011
  7. Descendant · cites it directly · 32 of its ownK110294 High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation · 2011 · Recalled (Class I)
  8. Descendant · 2 generations on · 31 of its ownK122180 High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · Recalled (Class I)
  9. Descendant · 2 generations on · 27 of its ownK143404 SurgiQuest AirSeal iFS SystemSurgiquest, Inc. · 2015 · Recalled
  10. Descendant · cites it directly · 22 of its ownK120151 Nebulae INorthgate Technologies, Inc. · 2012 · Recalled

Showing 5 of 44 descendants. See all descendants

Descendants cleared per year

2011: 22012: 42013: 32014: 12015: 12016: 12017: 32018: 12019: 22020: 32021: 12023: 42024: 42025: 72026: 7
2011high 7 (2025) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 44 · See all
  • K103692Sugiquest Airsealoptical Trocar & Cannula System with Integrated Insufflator Dpis 2000Surgiquest, Inc. · 2011 · cites it directly35descendants
  • K110294High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation · 2011 · cites it directly32descendants
  • K122180High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · 2 generations on31descendants
  • K143404SurgiQuest AirSeal iFS SystemSurgiquest, Inc. · 2015 · 2 generations on27descendants
  • K120151Nebulae INorthgate Technologies, Inc. · 2012 · cites it directly22descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K964171Vasoview Balloon Dissection SystemOrigin Medsystems, Inc.1997—
−3K981700Vasoview Dissection/Vessel Harvesting SystemOrigin Medsystems, Inc.1998—
−2K030512Guidant Vasoview Endoscopic Vessel Harvesting SystemGuidant Corp.20031
−1K06072340 L High Flow Insufflator F113W.O.M. World of Medicine AG2006—
0 (this device)K06336745L Core Insufflator, Model F114W.O.M. World of Medicine AG2007—
+1K103692Sugiquest Airsealoptical Trocar & Cannula System with Integrated Insufflator…Surgiquest, Inc.2011—
+1K110294High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation2011Class I
+2K122180High Flow Insufflantion UnitOlympus Medical Systems Corporation2012Class I
+2K143404SurgiQuest AirSeal iFS SystemSurgiquest, Inc.201511
+1K120151Nebulae INorthgate Technologies, Inc.20122

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
199620
199910
200010
200110
200320
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201121
201241
201330
201410
201510
201610
201730
201810
201920
202030
202110
202340
202440
202570
202670
FIG. 1 — Lineage of K063367, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HIF device

  1. #1 K250388 Endoflator +score 15.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K06072340 L High Flow Insufflator F113
Summary page 1
…pe, 21 C.F.R. § 876.1500 Regulatory Class: H1 Product Code: HIP / GCJ Predicate Devices: 1. 40 LHigh FlowlInsufflator F113, (K060723) 2. HI-FLO Thermne Pneu 45, (K03 1014) 3. Stryker Heated Insufflator Tube S…
W.O.M. World of Medicine AG2006—
*K031014HI-Flo Therme Pneu 45, Wisap #7083, 7083V
Summary page 1
…: H1 Product Code: HIP / GCJ Predicate Devices: 1. 40 LHigh FlowlInsufflator F113, (K060723) 2. HI-FLO Thermne Pneu 45, (K03 1014) 3. Stryker Heated Insufflator Tube Set, (K003792) 4. Surgiflator 20, Singl…
Wisap Gesellschaft Fur Wissenschaftl. App. Bau Mbh2003—
*K003792Stryker Heated Insufflator Tube Set
Summary page 1
…igh FlowlInsufflator F113, (K060723) 2. HI-FLO Thermne Pneu 45, (K03 1014) 3. Stryker Heated Insufflator Tube Set, (K003792) 4. Surgiflator 20, Single-Use Heatable Tube Set, (K950035) Device Desc…
Stryker Corp.20011
*K950035Surgiflator 20
Summary page 1
…3. Stryker Heated Insufflator Tube Set, (K003792) 4. Surgiflator 20, Single-Use Heatable Tube Set, (K950035) Device Description: The 45L CORE Insuflator F114 is a microprocessor controlled C02 insuffator designed…
W.O.M. GmbH1996—

What descends from it? (44)

See all 44

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103692Sugiquest Airsealoptical Trocar & Cannula System with Integrated Insufflator Dpis 2000Surgiquest, Inc.201135
K110294High Flow Insufflation Unit Uhi-4 Class IOlympus Medical Systems Corporation201132
K120151Nebulae INorthgate Technologies, Inc.201222
K153513Insufflator 50L FM134W.O.M. World of Medicine GmbH201621
K190303AirSealConmed Corporation20191
K170799AP 50/30 Insufflator with Insuflow PortLexion Medical, LLC20171
K230239AirSeal iFS System; AirSeal dV SolutionConmed Corporation20230

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 7, 2019Class IIZ-1726-2019Under Investigation by firmThe heated tube sets were consistently leaking and detaching from the cassette.
Mar 7, 2019Class IIZ-1727-2019Under Investigation by firmThe heated tube sets were consistently leaking and detaching from the cassette.
Mar 28, 2016Class IIZ-1566-2016Nonconforming Material/ComponentWhen operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display…
Dec 29, 2008Not classifiedZ-1198-2010Component design/selectionUnexpected powering down/failure to boot up: A problem with a transistor in the heater/temperature sensor component in the device may cause the device to power off during surgery or to fail to boot up at all. The risk includes delay and…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K110294High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K122180High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 3.0 years on median.
  • 2 of the 14 ancestors have a recall of any class.
  • Of its descendants, K103692 (Surgiquest, Inc., 2011) has gone furthest, with 35 descendants of its own.

How does its category compare?

All HIF devices

HIF Insufflator, Laparoscopic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · HIF

3in 2025

2007: 42008: 22009: 42010: 22011: 52012: 52013: 22014: 22015: 22016: 32017: 52018: 32019: 22020: 42021: 12022: 12023: 72024: 42025: 32026 (partial year): 3
2007high 7 (2023) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HIF

210in 2025

2007: 140 days median2009: 191 days median2011: 128 days median2012: 138 days median2016: 88 days median2017: 242 days median2018: 207 days median2020: 64 days median2023: 134 days median2024: 170 days median2025: 210 days median2026 (partial year): 265 days median
2007high 265 (2026) · low 64 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HIF

12so far in 2026, the most since 2007

2008: 12016: 12018: 12023: 1, 1 Class I2026 (partial year): 12
2007high 12 (2026) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · HIF

1,507in 2025

2007: 82008: 142009: 212010: 412011: 442012: 382013: 402014: 432015: 302016: 722017: 682018: 972019: 1282020: 3602021: 6542022: 6672023: 8902024: 1,4832025: 1,507
2007high 1,507 (2025) · low 8 (2007)2025

MAUDE reports naming product code HIF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007414008
20082—114
20094191021
20102—041
20115128044
20125138038
20132—040
20142—043
20152—030
2016388172
20175242068
20183207197
20192—0128
20204640360
20211—0654
20221—0667
202371341 (1)890
2024417001,483
2025321001,507
2026326512—

What is it indicated for?

The 45L CORE Insufflator F114 is a C02 insufflator intended for use in diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with gas. The high flow application, the low flow application and the bariatric application of the device are each indicated for use in facilitating the use of a laparoscope by filling the peritoneal cavity with gas to distend it. The low flow application of the device is indicated for pediatric use. The vessel harvesting application of the F114 is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein and/or radial artery during endoscopic vessel harvesting procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063367: 45L Core Insufflator, Model F114.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K063367

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026