510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Nebulae I

K120151510(k) TraditionalCleared Aug 17, 2012212-day reviewClass II21 CFR 884.1730Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Northgate Technologies, Inc. · HIF Insufflator, Laparoscopic

Cleared on Aug 17, 2012 as substantially equivalent to K063367 (W.O.M. World of Medicine AG, 2007) and 1 other predicate after a 212-day review. Its family tree reaches back 5 generations to 1997. 22 later devices descend from it. It has been recalled 2 times.

Decided
Aug 17, 2012
Review
212 days
Class
Class II
Product code
HIF
Recalls
2
Predicates
2

Its longest line of descent runs back 5 generations to K964171 (1997). 16 ancestors in all, the oldest K904993 (1991). 22 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 3 other ancestors2001: 1 other ancestor1996: 2 other ancestors2004: 1 other ancestor2000: 1 other ancestor1999: 1 other ancestor1994: 1 other ancestor1991: 1 other ancestor2016: 1 descendant cleared2017: 3 descendants cleared2023: 1 descendant cleared2020: 1 descendant cleared2025: 6 descendants cleared2019: 1 descendant cleared2024: 3 descendants cleared2026: 6 descendants clearedK964171 · Vasoview Balloon Dissection System · Origin Medsystems, Inc. · 1997K981700 · Vasoview Dissection/Vessel Harvesting System · Origin Medsystems, Inc. · 1998K030512 · Guidant Vasoview Endoscopic Vessel Harvesting System · Guidant Corp. · 2003 · recalledK060723 · 40 L High Flow Insufflator F113 · W.O.M. World of Medicine AG · 2006K063367 · 45L Core Insufflator, Model F114 · W.O.M. World of Medicine AG · 2007 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 2 other ancestors4 generations back: 5 other ancestors 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 4 descendants4 generations forward: 3 descendants5 generations forward: 7 descendants6 generations forward: 4 descendants K063367 · 45L Core Insufflator, Model F114 · W.O.M. World of Medicine AG · 2007 · recalledK060723 · 40 L High Flow Insufflator F113 · W.O.M. World of Medicine AG · 2006K030512 · Guidant Vasoview Endoscopic Vessel Harvesting System · Guidant Corp. · 2003 · recalledK981700 · Vasoview Dissection/Vessel Harvesting System · Origin Medsystems, Inc. · 1998K964171 · Vasoview Balloon Dissection System · Origin Medsystems, Inc. · 1997−50+5
  1. Root · 5 generations backK964171 Vasoview Balloon Dissection SystemOrigin Medsystems, Inc. · 1997
  2. 4 generations backK981700 Vasoview Dissection/Vessel Harvesting SystemOrigin Medsystems, Inc. · 1998
  3. 3 generations backK030512 Guidant Vasoview Endoscopic Vessel Harvesting SystemGuidant Corp. · 2003 · Recalled
  4. 2 generations backK060723 40 L High Flow Insufflator F113W.O.M. World of Medicine AG · 2006
  5. 1 generation backK063367 45L Core Insufflator, Model F114W.O.M. World of Medicine AG · 2007 · Recalled
  6. This deviceK120151 Nebulae INorthgate Technologies, Inc. · 2012
  7. Descendant · cites it directly · 21 of its ownK153513 Insufflator 50L FM134W.O.M. World of Medicine GmbH · 2016
  8. Descendant · 2 generations on · 17 of its ownK170784 PneumoclearW.O.M. World of Medicine GmbH · 2017 · Recalled
  9. Descendant · 3 generations on · 14 of its ownK201361 PneumoclearW.O.M. World of Medicine GmbH · 2020
  10. Descendant · 4 generations on · 11 of its ownK232610 da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube…Intuitive Surgical, Inc. · 2024 · Recalled
  11. Descendant · 5 generations on · 2 of its ownK251739 da Vinci Surgical System (IS5000)Intuitive Surgical, Inc. · 2026

Showing 5 of 22 descendants. See all descendants

Descendants cleared per year

2016: 12017: 32019: 12020: 12023: 12024: 32025: 62026: 6
2016high 6 (2025) · low 0 (2018)2026

Most influential descendants

Showing 5 of 22 · See all
  • K153513Insufflator 50L FM134W.O.M. World of Medicine GmbH · 2016 · cites it directly21descendants
  • K170784PneumoclearW.O.M. World of Medicine GmbH · 2017 · 2 generations on17descendants
  • K201361PneumoclearW.O.M. World of Medicine GmbH · 2020 · 3 generations on14descendants
  • K232610da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube Set with Smoke…Intuitive Surgical, Inc. · 2024 · 4 generations on11descendants
  • K251739da Vinci Surgical System (IS5000)Intuitive Surgical, Inc. · 2026 · 5 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K964171Vasoview Balloon Dissection SystemOrigin Medsystems, Inc.1997—
−4K981700Vasoview Dissection/Vessel Harvesting SystemOrigin Medsystems, Inc.1998—
−3K030512Guidant Vasoview Endoscopic Vessel Harvesting SystemGuidant Corp.20031
−2K06072340 L High Flow Insufflator F113W.O.M. World of Medicine AG2006—
−1K06336745L Core Insufflator, Model F114W.O.M. World of Medicine AG20074
0 (this device)K120151Nebulae INorthgate Technologies, Inc.2012—
+1K153513Insufflator 50L FM134W.O.M. World of Medicine GmbH2016—
+2K170784PneumoclearW.O.M. World of Medicine GmbH20171
+3K201361PneumoclearW.O.M. World of Medicine GmbH2020—
+4K232610da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube Set with…Intuitive Surgical, Inc.20246
+5K251739da Vinci Surgical System (IS5000)Intuitive Surgical, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
199620
199910
200010
200110
200330
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201730
201910
202010
202310
202430
202560
202660
FIG. 1 — Lineage of K120151, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HIF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HIF device

  1. #1 K250388 Endoflator +score 15.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K06336745L Core Insufflator, Model F114
Summary page 2
…fflator for Laparoscopy and Endoscopic Vessel Harvesting Manufacturer: W.OM. World of Medicine 510(k) No. K063367 2. Device Name: 4OLPM ABDOMINAL INSUFFLATOR, CATALOG # 72-00203-0 Com…
W.O.M. World of Medicine AG20074
*K02205240LPM Abdominal Insufflator, Catalog # 72-00203-0
Summary page 2
…, CATALOG # 72-00203-0 Common Name: Insufflator Manufacturer: Northgate Technologies Inc. 510(k) No. K022052 4. Device Description The Nebulae TM 150 LPM Insufflator is microprocessor controlled. This CO2 (Carbon Dioxide) Hig…
Northgate Technologies, Inc.2003—

Also cleared against K063367: 6 other devices →

What descends from it? (22)

See all 22

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153513Insufflator 50L FM134W.O.M. World of Medicine GmbH201621

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 15, 2021Class IIZ-1147-2021Nonconforming Material/ComponentAn internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.
Nov 21, 2016Class IIZ-0866-2017Software designDevice could cause a patient overpressure situation without any visual or auditory indication or warning

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 3 of the 16 ancestors have a recall of any class.
  • Of its descendants, K153513 (W.O.M. World of Medicine GmbH, 2016) has gone furthest, with 21 descendants of its own.
  • 6 other devices also cite its predicate K063367.

How does its category compare?

All HIF devices

HIF Insufflator, Laparoscopic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · HIF

3in 2025

2007: 42008: 22009: 42010: 22011: 52012: 52013: 22014: 22015: 22016: 32017: 52018: 32019: 22020: 42021: 12022: 12023: 72024: 42025: 32026 (partial year): 3
2007high 7 (2023) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HIF

210in 2025

2007: 140 days median2009: 191 days median2011: 128 days median2012: 138 days median2016: 88 days median2017: 242 days median2018: 207 days median2020: 64 days median2023: 134 days median2024: 170 days median2025: 210 days median2026 (partial year): 265 days median
2007high 265 (2026) · low 64 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HIF

12so far in 2026, the most since 2007

2008: 12016: 12018: 12023: 1, 1 Class I2026 (partial year): 12
2007high 12 (2026) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · HIF

1,507in 2025

2007: 82008: 142009: 212010: 412011: 442012: 382013: 402014: 432015: 302016: 722017: 682018: 972019: 1282020: 3602021: 6542022: 6672023: 8902024: 1,4832025: 1,507
2007high 1,507 (2025) · low 8 (2007)2025

MAUDE reports naming product code HIF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007414008
20082—114
20094191021
20102—041
20115128044
20125138038
20132—040
20142—043
20152—030
2016388172
20175242068
20183207197
20192—0128
20204640360
20211—0654
20221—0667
202371341 (1)890
2024417001,483
2025321001,507
2026326512—

What is it indicated for?

The Nebulae TM 1 50 LPM Insufflator provides CO2 gas distention of surgical cavities for diagnostic and/or operative endoscopy. The insufflator has multiple operating modes which can be used for the following procedures:

General Laparoscopic

Pediatric Laparoscopic

Bariatric Laparoscopic

Minimally Invasive Vessel Harvesting

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120151: Nebulae I.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K120151

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026