510(k) · Cleared
Recalled · Class I · 2023High Flow Insufflation Unit Uhi-4
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Olympus Medical Systems Corporation · HIF Insufflator, Laparoscopic
Cleared on Nov 18, 2011 as substantially equivalent to K063367 (W.O.M. World of Medicine AG, 2007) and 1 other predicate after a 290-day review. Its family tree reaches back 5 generations to 1997. 3 of the 50 devices in its family, counting itself, have a Class I recall. 32 later devices descend from it. It has been recalled 2 times, including a Class I recall.
- Decided
- Nov 18, 2011
- Review
- 290 days
- Class
- Class II
- Product code
- HIF
- Recalls
- 2 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 5 generations to K964171 (1997). 17 ancestors in all, the oldest K904993 (1991). 32 later devices descend from it, 1 citing it directly. 3 of the 50 devices in this family, counting this one, have a Class I recall.
- Root · 5 generations backK964171 Vasoview Balloon Dissection SystemOrigin Medsystems, Inc. · 1997
- 4 generations backK981700 Vasoview Dissection/Vessel Harvesting SystemOrigin Medsystems, Inc. · 1998
- 3 generations backK030512 Guidant Vasoview Endoscopic Vessel Harvesting SystemGuidant Corp. · 2003 · Recalled
- 2 generations backK060723 40 L High Flow Insufflator F113W.O.M. World of Medicine AG · 2006
- 1 generation backK063367 45L Core Insufflator, Model F114W.O.M. World of Medicine AG · 2007 · Recalled
- This deviceK110294 High Flow Insufflation Unit Uhi-4Olympus Medical Systems Corporation · 2011
- Descendant · cites it directly · 31 of its ownK122180 High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · Recalled (Class I)
- Descendant · 2 generations on · 27 of its ownK143404 SurgiQuest AirSeal iFS SystemSurgiquest, Inc. · 2015 · Recalled
- Descendant · 3 generations on · 17 of its ownK170784 PneumoclearW.O.M. World of Medicine GmbH · 2017 · Recalled
- Descendant · 4 generations on · 14 of its ownK201361 PneumoclearW.O.M. World of Medicine GmbH · 2020
- Descendant · 5 generations on · 11 of its ownK232610 da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube…Intuitive Surgical, Inc. · 2024 · Recalled
Showing 5 of 32 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 32 · See all- K122180High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · cites it directly31descendants
- K143404SurgiQuest AirSeal iFS SystemSurgiquest, Inc. · 2015 · 2 generations on27descendants
- K170784PneumoclearW.O.M. World of Medicine GmbH · 2017 · 3 generations on17descendants
- K201361PneumoclearW.O.M. World of Medicine GmbH · 2020 · 4 generations on14descendants
- K232610da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube Set with Smoke…Intuitive Surgical, Inc. · 2024 · 5 generations on11descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K964171 | Vasoview Balloon Dissection System | Origin Medsystems, Inc. | 1997 | — |
| −4 | K981700 | Vasoview Dissection/Vessel Harvesting System | Origin Medsystems, Inc. | 1998 | — |
| −3 | K030512 | Guidant Vasoview Endoscopic Vessel Harvesting System | Guidant Corp. | 2003 | 1 |
| −2 | K060723 | 40 L High Flow Insufflator F113 | W.O.M. World of Medicine AG | 2006 | — |
| −1 | K063367 | 45L Core Insufflator, Model F114 | W.O.M. World of Medicine AG | 2007 | 4 |
| 0 (this device) | K110294 | High Flow Insufflation Unit Uhi-4 | Olympus Medical Systems Corporation | 2011 | — |
| +1 | K122180 | High Flow Insufflantion Unit | Olympus Medical Systems Corporation | 2012 | Class I |
| +2 | K143404 | SurgiQuest AirSeal iFS System | Surgiquest, Inc. | 2015 | 11 |
| +3 | K170784 | Pneumoclear | W.O.M. World of Medicine GmbH | 2017 | 1 |
| +4 | K201361 | Pneumoclear | W.O.M. World of Medicine GmbH | 2020 | — |
| +5 | K232610 | da Vinci Surgical System Model IS5000, da Vinci Insufflator and Tube Set with… | Intuitive Surgical, Inc. | 2024 | 6 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1991 | 1 | 0 |
| 1994 | 1 | 0 |
| 1996 | 2 | 0 |
| 1999 | 2 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 1 | 1 |
| 2003 | 2 | 0 |
| 2004 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 1 | 1 |
| 2015 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2019 | 2 | 0 |
| 2020 | 3 | 0 |
| 2021 | 1 | 0 |
| 2023 | 3 | 0 |
| 2024 | 4 | 0 |
| 2025 | 7 | 0 |
| 2026 | 7 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HIF device
- #1 K250388 Endoflator +score 15.8 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K063367 | 45L Core Insufflator, Model F114Summary page 2…Endoscopic Vessel Harvesting * Manufacturer: W. 0. M. WORLD OF MEDICINE AG E 510(k) No. K063367 2 " Device Name: HIGH FLOW INSUFFLATION UNIT UHI-3 * Common Name: IN… | 2007 |
| * | K014166 | Uhi-3 High Flow Insufflation UnitSummary page 2…mmon Name: INSUFFLATOR * Manufacturer: OLYMPUS MEDICAL SYSTEM CORP. * 510(k) No. K014166 4 Device Description The subject device , UHI-4, has new insufflation function, small mode, for insufflation of cavities alon… | 2002 |
What descends from it? (32)
See all 321 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K122180 | High Flow Insufflantion Unit Class I | 2012 | 31 |
Has anything in its family been recalled? (3 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 25, 2023 | Class II | Z-1375-2024 | Process control | Olympus has become aware of an increased trend of both repairs and customer complaints reporting "UHI-4 stopping air supply due to the front panel LED turning off". |
| Sep 22, 2023 | Class I | Z-0075-2024 | Under Investigation by firm | There have been reports of patients suffering arrhythmias, reported as short cardiac arrests, during surgical procedures where UHI-4s were used. These events may have been due to an over insufflation of the abdominal cavity resulting from… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K014166Uhi-3 High Flow Insufflation UnitOlympus Optical Co., Ltd. · 2002 · Direct predicate · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
- K122180High Flow Insufflantion UnitOlympus Medical Systems Corporation · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
- 4 of the 17 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K122180 (Olympus Medical Systems Corporation, 2012) has gone furthest, with 31 descendants of its own.
- 6 other devices also cite its predicate K063367.
How does its category compare?
All HIF devicesHIF Insufflator, Laparoscopic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · HIF
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HIF
210in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HIF
12so far in 2026, the most since 2007
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · HIF
1,507in 2025
MAUDE reports naming product code HIF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 140 | 0 | 8 |
| 2008 | 2 | — | 1 | 14 |
| 2009 | 4 | 191 | 0 | 21 |
| 2010 | 2 | — | 0 | 41 |
| 2011 | 5 | 128 | 0 | 44 |
| 2012 | 5 | 138 | 0 | 38 |
| 2013 | 2 | — | 0 | 40 |
| 2014 | 2 | — | 0 | 43 |
| 2015 | 2 | — | 0 | 30 |
| 2016 | 3 | 88 | 1 | 72 |
| 2017 | 5 | 242 | 0 | 68 |
| 2018 | 3 | 207 | 1 | 97 |
| 2019 | 2 | — | 0 | 128 |
| 2020 | 4 | 64 | 0 | 360 |
| 2021 | 1 | — | 0 | 654 |
| 2022 | 1 | — | 0 | 667 |
| 2023 | 7 | 134 | 1 (1) | 890 |
| 2024 | 4 | 170 | 0 | 1,483 |
| 2025 | 3 | 210 | 0 | 1,507 |
| 2026 | 3 | 265 | 12 | — |
What is it indicated for?
This instrument has been designed for insufflation of abdominal cavity, and provides automatic suction and smoke evacuation to facilitate laparoscopic observation, diagnosis and treatment. This instrument is also designed for controlled CO2 insufflation to create a cavity along the saphenous vein and/or radial artery to facilitate observation during an endoscopic vessel harvesting procedure.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K110294 · printed Nov 3, 2026
Cite this page
510(k) Report. “K110294: High Flow Insufflation Unit Uhi-4.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K110294
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026