510(k) · Cleared
ClearedBP-200 Plus
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Schiller AG · DXN System, Measurement, Blood-Pressure, Non-Invasive
Cleared on Mar 7, 2007 as substantially equivalent to K053209 (Suntech Medical, Inc., 2006) after a 75-day review. Its family tree reaches back 7 generations to 1991. 7 later devices descend from it.
- Decided
- Mar 7, 2007
- Review
- 75 days
- Class
- Class II
- Product code
- DXN
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 7 generations to K910772 (1991). 22 ancestors in all, the oldest K893221 (1990). 7 later devices descend from it, 2 citing it directly.
- Root · 7 generations backK910772 Propaq 102, 104 and 106Protocol Systems, Inc. · 1991
- 6 generations backK972020 Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc. · 1998
- 5 generations backK980879 Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc. · 1998
- 4 generations backK982776 9303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc. · 1998
- 3 generations backK013319 Nonin Model 2120 Pulse Oximeter and Noninvasive Blood Pressure…Nonin Medical, Inc. · 2002
- 2 generations backK031487 Model Avant 2120 Pulse Oximeter and NibpNonin Medical, Inc. · 2004
- 1 generation backK053209 Tango+ Automatic Blood Pressure and Oxygen Saturation Measurement…Suntech Medical, Inc. · 2006
- This deviceK063814 BP-200 PlusSchiller AG · 2007
- Descendant · cites it directly · 4 of its ownK101619 Welch Allyn 1500 Patient MonitorSchiller AG · 2010
- Descendant · 2 generations on · 3 of its ownK133973 Tempus Pro with Enhanced FeaturesRemote Diagnostic Technologies, Ltd. · 2014
- Descendant · 3 generations on · 2 of its ownK170567 Tempus Pro Patient MonitorRemote Diagnostic Technologies, Ltd. · 2017
- Descendant · 4 generations on · 1 of its ownK173768 Tempus Pro Patient MonitorRemote Diagnostic Technologies, Ltd. · 2018
- Descendant · cites it directly · 1 of its ownK152043 Diagnostic Station DS20Schiller AG · 2016
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 7 · See all- K101619Welch Allyn 1500 Patient MonitorSchiller AG · 2010 · cites it directly4descendants
- K133973Tempus Pro with Enhanced FeaturesRemote Diagnostic Technologies, Ltd. · 2014 · 2 generations on3descendants
- K170567Tempus Pro Patient MonitorRemote Diagnostic Technologies, Ltd. · 2017 · 3 generations on2descendants
- K173768Tempus Pro Patient MonitorRemote Diagnostic Technologies, Ltd. · 2018 · 4 generations on1descendant
- K152043Diagnostic Station DS20Schiller AG · 2016 · cites it directly1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −7 | K910772 | Propaq 102, 104 and 106 | Protocol Systems, Inc. | 1991 | — |
| −6 | K972020 | Oscillomate Non-Invasive Blood Pressure Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −5 | K980879 | Oscillomate Nibp / SP02 Ems Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −4 | K982776 | 9303 Neonatal/Adult Vital Signs Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −3 | K013319 | Nonin Model 2120 Pulse Oximeter and Noninvasive Blood Pressure (Nibp) Monitor | Nonin Medical, Inc. | 2002 | — |
| −2 | K031487 | Model Avant 2120 Pulse Oximeter and Nibp | Nonin Medical, Inc. | 2004 | — |
| −1 | K053209 | Tango+ Automatic Blood Pressure and Oxygen Saturation Measurement System | Suntech Medical, Inc. | 2006 | — |
| 0 (this device) | K063814 | BP-200 Plus | Schiller AG | 2007 | — |
| +1 | K101619 | Welch Allyn 1500 Patient Monitor | Schiller AG | 2010 | — |
| +2 | K133973 | Tempus Pro with Enhanced Features | Remote Diagnostic Technologies, Ltd. | 2014 | — |
| +3 | K170567 | Tempus Pro Patient Monitor | Remote Diagnostic Technologies, Ltd. | 2017 | — |
| +4 | K173768 | Tempus Pro Patient Monitor | Remote Diagnostic Technologies, Ltd. | 2018 | — |
| +1 | K152043 | Diagnostic Station DS20 | Schiller AG | 2016 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1990 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1995 | 1 | 0 |
| 1996 | 1 | 0 |
| 1997 | 3 | 0 |
| 1998 | 3 | 0 |
| 1999 | 1 | 0 |
| 2000 | 2 | 0 |
| 2001 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 1 | 0 |
| 2014 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2020 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DXN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DXN device
- #1 K222125 Arm Blood Pressure Monitor, models AOJ-30A…score 18.5 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K053209 | Tango+ Automatic Blood Pressure and Oxygen Saturation Measurement SystemSummary page 1…DXN Regulatory Class: Class II(two) Regulation Number: 21 CFR 870.1130 Legally-marketed predicate devices Tango +, K053209, SunTech Medical Instruments Inc. The SCHILLER BP-200 plus is substantially equivalent to the above mentioned device. The followi… | 2006 |
| † | K990966 | Masimo Set 2000 Pulse Oximeter, Masimo Lnop -Adt-Adult/Pediatric Disposable Sensor… | 1999 |
| † | K973735 | Microvit MT-100 | 1998 |
| † | K012226 | Monitoring System,Model Argus PB-1000 | 2002 |
What descends from it? (7)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 1.8 years on median.
- 2 of the 22 ancestors have a recall of any class.
- Of its descendants, K101619 (Schiller AG, 2010) has gone furthest, with 4 descendants of its own.
- 3 other devices also cite its predicate K053209.
How does its category compare?
All DXN devicesDXN System, Measurement, Blood-Pressure, Non-Invasive: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · DXN
28in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DXN
142in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DXN
3in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DXN
239in 2025
MAUDE reports naming product code DXN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 27 | 118 | 0 | 21 |
| 2008 | 23 | 82 | 1 | 8 |
| 2009 | 46 | 73 | 3 | 17 |
| 2010 | 43 | 94 | 0 | 14 |
| 2011 | 34 | 72 | 1 | 10 |
| 2012 | 36 | 116 | 0 | 19 |
| 2013 | 34 | 150 | 1 | 72 |
| 2014 | 29 | 133 | 0 | 47 |
| 2015 | 30 | 97 | 0 | 62 |
| 2016 | 30 | 144 | 1 | 54 |
| 2017 | 33 | 184 | 0 | 90 |
| 2018 | 22 | 235 | 0 | 617 |
| 2019 | 33 | 156 | 1 | 96 |
| 2020 | 24 | 198 | 0 | 130 |
| 2021 | 23 | 190 | 0 | 162 |
| 2022 | 25 | 137 | 0 | 155 |
| 2023 | 35 | 182 | 0 | 162 |
| 2024 | 17 | 107 | 3 | 270 |
| 2025 | 28 | 142 | 0 | 239 |
| 2026 | 29 | 147 | 0 | — |
What is it indicated for?
The BP-200 plus is intended to be used as an adjunct to exercise stress testing devices. It is intended to measure and display diastolic and systolic blood pressure, heart rate, percentage of oxygen saturation in arterial blood (SpO2) and pulse rate in adult or adolescent patients during stress tests. The measurement cuff of the BP-200 plus in intended to be placed on the upper right arm of the patient. The BP-200 plus can be used for patients of both sexes and all races. The BP-200 plus should not be used with neonates.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K063814 · printed Nov 5, 2026
Cite this page
510(k) Report. “K063814: BP-200 Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K063814