510(k) Report

510(k) · Cleared

Cleared

Nonin Model 2120 Pulse Oximeter and Noninvasive Blood Pressure (Nibp) Monitor

K013319510(k) TraditionalCleared Jan 3, 200290-day reviewClass II21 CFR 870.2700

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nonin Medical, Inc. · DQA Oximeter

Cleared on Jan 3, 2002 as substantially equivalent to K970763 (Medical Data Electronics, 1997) and 2 other predicates after a 90-day review. Its family tree reaches back 4 generations to 1991. 124 later devices descend from it.

Decided
Jan 3, 2002
Review
90 days
Class
Class II
Product code
DQA
Recalls
None
Predicates
3

Its longest line of descent runs back 4 generations to K910772 (1991). 11 ancestors in all, the oldest K893221 (1990). 124 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1997: 2 other ancestors1995: 1 other ancestor1993: 1 other ancestor1990: 1 other ancestor1992: 1 other ancestor2004: 4 descendants cleared2006: 2 descendants cleared2009: 3 descendants cleared2005: 2 descendants cleared2019: 10 descendants cleared2012: 7 descendants cleared2020: 8 descendants cleared2008: 3 descendants cleared2017: 5 descendants cleared2007: 1 descendant cleared2014: 5 descendants cleared2018: 6 descendants cleared2016: 7 descendants cleared2010: 1 descendant cleared2013: 6 descendants cleared2015: 4 descendants cleared2025: 15 descendants cleared2011: 1 descendant cleared2024: 10 descendants cleared2023: 6 descendants cleared2022: 11 descendants cleared2026: 5 descendants cleared2021: 2 descendants clearedK910772 · Propaq 102, 104 and 106 · Protocol Systems, Inc. · 1991×3: 3 generations cleared close together, K972020 (1998), K980879 (1998), K982776 (1998) 1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 4 other ancestors4 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 6 descendants3 generations forward: 15 descendants4 generations forward: 26 descendants5 generations forward: 19 descendants6 generations forward: 21 descendants7 generations forward: 18 descendants8 generations forward: 15 descendants9 generations forward: 2 descendants K982776 · 9303 Neonatal/Adult Vital Signs Monitor · Cas Medical Systems, Inc. · 1998K980879 · Oscillomate Nibp / SP02 Ems Monitor · Cas Medical Systems, Inc. · 1998K972020 · Oscillomate Non-Invasive Blood Pressure Monitor · Cas Medical Systems, Inc. · 1998K910772 · Propaq 102, 104 and 106 · Protocol Systems, Inc. · 19910+5
  1. Root · 4 generations backK910772 Propaq 102, 104 and 106Protocol Systems, Inc. · 1991
  2. 3 generations backK972020 Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc. · 1998
  3. 2 generations backK980879 Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc. · 1998
  4. 1 generation backK982776 9303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc. · 1998
  5. This deviceK013319 Nonin Model 2120 Pulse Oximeter and Noninvasive Blood Pressure…Nonin Medical, Inc. · 2002
  6. Descendant · cites it directly · 89 of its ownK041156 Avant Digital Pulse Oximetry System, Models 4000, 4001Nonin Medical, Inc. · 2004 · Recalled
  7. Descendant · 2 generations on · 75 of its ownK043359 4100 Patient Oximeter ModuleNonin Medical, Inc. · 2005
  8. Descendant · cites it directly · 42 of its ownK031487 Model Avant 2120 Pulse Oximeter and NibpNonin Medical, Inc. · 2004
  9. Descendant · 3 generations on · 40 of its ownK081285 Finger Pulse Oximeter, Model 9560Nonin Medical, Inc. · 2008
  10. Descendant · 2 generations on · 27 of its ownK040589 Modification to Avant 9600 Pulse OximeterNonin Medical, Inc. · 2004

Showing 5 of 124 descendants. See all descendants

Descendants cleared per year

2004: 42005: 22006: 22007: 12008: 32009: 32010: 12011: 12012: 72013: 62014: 52015: 42016: 72017: 52018: 62019: 102020: 82021: 22022: 112023: 62024: 102025: 152026: 5
2004high 15 (2025) · low 1 (2007)2026

Most influential descendants

Showing 5 of 124 · See all
  • K041156Avant Digital Pulse Oximetry System, Models 4000, 4001Nonin Medical, Inc. · 2004 · cites it directly89descendants
  • K0433594100 Patient Oximeter ModuleNonin Medical, Inc. · 2005 · 2 generations on75descendants
  • K031487Model Avant 2120 Pulse Oximeter and NibpNonin Medical, Inc. · 2004 · cites it directly42descendants
  • K081285Finger Pulse Oximeter, Model 9560Nonin Medical, Inc. · 2008 · 3 generations on40descendants
  • K040589Modification to Avant 9600 Pulse OximeterNonin Medical, Inc. · 2004 · 2 generations on27descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K910772Propaq 102, 104 and 106Protocol Systems, Inc.1991—
−3K972020Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc.1998—
−2K980879Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc.1998—
−1K9827769303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc.1998—
0 (this device)K013319Nonin Model 2120 Pulse Oximeter and Noninvasive Blood Pressure (Nibp) MonitorNonin Medical, Inc.2002—
+1K041156Avant Digital Pulse Oximetry System, Models 4000, 4001Nonin Medical, Inc.20041
+2K0433594100 Patient Oximeter ModuleNonin Medical, Inc.2005—
+1K031487Model Avant 2120 Pulse Oximeter and NibpNonin Medical, Inc.2004—
+3K081285Finger Pulse Oximeter, Model 9560Nonin Medical, Inc.2008—
+2K040589Modification to Avant 9600 Pulse OximeterNonin Medical, Inc.2004—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199210
199310
199510
199720
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200440
200520
200620
200710
200830
200930
201010
201110
201270
201360
201450
201540
201670
201750
201860
2019100
202080
202120
2022110
202360
2024100
2025150
202650
FIG. 1 — Lineage of K013319, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970763Modular Telemetry System (MTS) Option for the Mde Escort II Monitor
Summary page 1
…een established under Section 514 of the Food, Drug and Cosmetic Act for Pulse Oximeters. (3) Legally-marketed predicate devices MTS Option for the ESCORT® O Monitor, K970763, made by Medical Data Electronics, clearance date, July 14, 1997. Model 9303 Neonatal / Adult Vital Signs Monitor, K 982776, mad…
Medical Data Electronics1997—
*K9827769303 Neonatal/Adult Vital Signs Monitor
Summary page 1
…evices MTS Option for the ESCORT® O Monitor, K970763, made by Medical Data Electronics, clearance date, July 14, 1997. Model 9303 Neonatal / Adult Vital Signs Monitor, K 982776, made by CAS Medical Systems, Inc., clearance date, November 2, 1998. OSCAR 2, K003004, made by SunTech Medical Instruments, cle…
Cas Medical Systems, Inc.1998—
*K003004Oscar 2, Model 222
Summary page 1
…clearance date, July 14, 1997. Model 9303 Neonatal / Adult Vital Signs Monitor, K 982776, made by CAS Medical Systems, Inc., clearance date, November 2, 1998. OSCAR 2, K003004, made by SunTech Medical Instruments, clearance date, October 25, 2000. The NONIN® Model 2120 Pulse Oximeter and Noninvasive Blo…
Suntech Medical Instruments, Inc.2000—

Also cleared against K970763: 3 other devices →

What descends from it? (124)

See all 124

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041156Avant Digital Pulse Oximetry System, Models 4000, 4001Nonin Medical, Inc.200489
K031487Model Avant 2120 Pulse Oximeter and NibpNonin Medical, Inc.200442

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 3.2 years on median.
  • 1 of the 11 ancestors has a recall of any class.
  • Of its descendants, K041156 (Nonin Medical, Inc., 2004) has gone furthest, with 89 descendants of its own.
  • 3 other devices also cite its predicate K970763.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

The NONIN® Model 2120 Pulse Oximeter and NIBP Monitor is a portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (Sp0,), pulse rate, and blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The Model 2120 is intended for spot-checking and / or continuous monitoring of patients. Its functions may be used separately or simultaneously. The…

Read the full statement (166 words)

The NONIN® Model 2120 Pulse Oximeter and NIBP Monitor is a portable device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (Sp0,), pulse rate, and blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The Model 2120 is intended for spot-checking and / or continuous monitoring of patients. Its functions may be used separately or simultaneously.

The pulse oximeter is intended for noninvasively monitoring the oxygen saturation and pulse rate of adult, pediatric, infant, and neonatal patients in hospitals, medical facilities, and subacute environments. It may be used for spot-checking and/or continuous monitoring of patients.

The blood pressure monitor is intended for noninvasively monitoring the blood pressure of adult and pediatric patients in hospitals, medical facilities, and subacute environments. The blood pressure monitor is not intended for use with neonates. It is intended for attended care and may be used for spot-checking. The Model 2120 should be used for patients with arm circumferences of 18-42 cm.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013319: Nonin Model 2120 Pulse Oximeter and Noninvasive Blood Pressure (Nibp) Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K013319

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026