510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

Hospira Plum a+ Infusion Pump, Model 12391

K070398510(k) SpecialCleared Apr 24, 200771-day reviewClass II21 CFR 880.5725Recalls 21 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hospira, Inc. · FRN Pump, Infusion

Cleared on Apr 24, 2007 as substantially equivalent to K052052 (Hospira, Inc., 2005) and 1 other predicate after a 71-day review. Its family tree reaches back 4 generations to 2001. 4 of the 7 devices in its family, counting itself, have a Class I recall. It has been recalled 21 times, including 4 Class I recalls.

Decided
Apr 24, 2007
Review
71 days
Class
Class II
Product code
FRN
Recalls
21 (4 Class I)
Predicates
2

Its longest line of descent runs back 4 generations to K011442 (2001). 6 ancestors in all. No later device cites it as a predicate yet. 4 of the 7 devices in this family, counting this one, have a Class I recall.

  1. Root · 4 generations backK011442 Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · Recalled (Class I)
  2. 3 generations backK031185 Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories · 2003 · Recalled
  3. 2 generations backK042081 Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3…Hospira, Inc. · 2004 · Recalled (Class I)
  4. 1 generation backK052052 Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3…Hospira, Inc. · 2005 · Recalled (Class I)
  5. This deviceK070398 Hospira Plum a+ Infusion Pump, Model 12391Hospira, Inc. · 2007

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories2001Class I
−3K031185Abbott Plum a+ Infusion Pump with a HPL/RS Plug-and-Play ModuleAbbott Laboratories20034
−2K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion…Hospira, Inc.2004Class I
−1K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump…Hospira, Inc.2005Class I
0 (this device)K070398Hospira Plum a+ Infusion Pump, Model 12391Hospira, Inc.2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220
FIG. 1 — Lineage of K070398, 2001–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5
Summary page 1
…Product code FRN 4. Trade Name of Proposed Device: Plum A+ ® Infusion Pump, 5. Predicate Devices: K052052 Hospira Plum A+ Infusion Pump System v 1 Cleared 08/24/2005…
Hospira, Inc.2005Class I
*K024084Abbott Plum a+ Infusion Pump, Model 12391
Summary page 1
…K052052 Hospira Plum A+ Infusion Pump System v 1 Cleared 08/24/2005 K024084 Abbott Plum A- Infusion Pump, Model 12391 Cleared 12/31/2002 6. Manufacturer and Est…
Abbott Laboratories2002—

Also cleared against K052052: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 7, 2024Class IZ-2448-2024Process controlDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than…
May 7, 2024Class IZ-2449-2024Process controlDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than…
Mar 22, 2023Class IZ-1566-2023Under Investigation by firmDue to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than…
Jul 2, 2014Class IZ-1072-2015Component change controlOne lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
Feb 1, 2013Class IIZ-0287-2014Component design/selectionPlum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and…
Feb 1, 2013Class IIZ-2681-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2683-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2675-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2678-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2680-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2677-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2679-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2676-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Feb 1, 2013Class IIZ-2682-2017OtherA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Dec 7, 2012Class IIZ-0624-2013Device DesignThe door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Oct 26, 2012Class IIZ-0481-2013Device DesignThe volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a…
Feb 5, 2012Class IIZ-0207-2014Nonconforming Material/ComponentHospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186…
Sep 16, 2011Not classifiedZ-0009-2012Software designHospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This…
Sep 16, 2011Not classifiedZ-3284-2011Nonconforming Material/ComponentHospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged…
Feb 14, 2011Not classifiedZ-1644-2011Nonconforming Material/ComponentHospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.
Aug 11, 2009Not classifiedZ-0024-2010Component design/selectionFire/Shock hazard-- A component used in the device may cause sparks/flashes and poses a fire and shock hazard.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K052052Hospira Plum a+ Infusion Pump System V11.5, Hospira Plum a+ 3 Infusion Pump System, V11.5Hospira, Inc. · 2005 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K042081Plum a+ Infusion System with Hospira Mednet Software , and Plum a+3 Infusion System with Hospira Mednet…Hospira, Inc. · 2004 · 2 generations back · ◆ on the line of descent in FIG. 14 Class I recalls
  • K011442Abbott Plum a+Infusion Pump, Model 11971Abbott Laboratories · 2001 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 1.7 years on median.
  • 5 of the 6 ancestors have a recall of any class; 3 have a Class I recall.
  • 1 other device also cites its predicate K052052.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

Intended for use in parenteral, enteral, and epidural therapies and the

administration of whole blood and blood products

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K070398: Hospira Plum a+ Infusion Pump, Model 12391.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K070398

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026