510(k) · Decided
Recalled · 2008Ash Advance Hemodialysis Catheter and Procedure Kit, Models; 15FR X 24CM, 15FR X 28CM, 15FR X 32CM
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ash Access Technology, Inc. · MSD Catheter, Hemodialysis, Implanted
Cleared on Jun 26, 2007 as substantially equivalent to K002901 (The Kendall Company, 2001) and 3 other predicates after a 118-day review. Its family tree reaches back 5 generations to 1998. 3 of the 21 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled 3 times.
- Decided
- Jun 26, 2007
- Review
- 118 days
- Class
- Class II
- Product code
- MSD
- Recalls
- 3
- Predicates
- 4
Its longest line of descent runs back 5 generations to K981994 (1998). 13 ancestors in all, the oldest K972207 (1998). 7 later devices descend from it, 2 citing it directly. 3 of the 21 devices in this family, counting this one, have a Class I recall.
- Root · 5 generations backK981994 Vascath Opti-Flow Long Term Dual Lumen Hemodialysis CatheterC.R. Bard, Inc. · 1998
- 4 generations backK012562 Medcomp 14.5F Double Lumen Hemodialysis CatheterMedical Components, Inc. · 2002
- 3 generations backK020465 Medcomp Ash Split-Cath IIMedcomp · 2002
- 2 generations backK022678 Medcomp SC4Medcomp · 2003
- 1 generation backK040318 Medcomp Split-Cath IIMedcomp · 2005
- This deviceK070572 Ash Advance Hemodialysis Catheter and Procedure Kit, Models; 15FR X…Ash Access Technology, Inc. · 2007
- Descendant · cites it directly · 6 of its ownK080400 Angiodynamics, Inc., Duramax Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2008 · Recalled
- Descendant · 2 generations on · 2 of its ownK101843 Duramax Chronic Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2010 · Recalled (Class I)
- Descendant · cites it directly · 2 of its ownK092597 Centros Chronic Hemodialysis Catheter SetAsh Access Technology, Inc. · 2009
- Descendant · 2 generations on · 1 of its ownK141363 Centros Long Term Hemodialysis Catheter, Centrosflo Long Term…Merit Medical Systems, Inc. · 2014
- Descendant · 3 generations on · 0 of its ownK151967 Centros Long-Term Hemodialysis Catheter, CentrosFLO Long-Term…Merit Medical Systems, Inc. · 2016 · Recalled (Class I)
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 7 · See all- K080400Angiodynamics, Inc., Duramax Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2008 · cites it directly6descendants
- K101843Duramax Chronic Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2010 · 2 generations on2descendants
- K092597Centros Chronic Hemodialysis Catheter SetAsh Access Technology, Inc. · 2009 · cites it directly2descendants
- K141363Centros Long Term Hemodialysis Catheter, Centrosflo Long Term Hemodialysis CatheterMerit Medical Systems, Inc. · 2014 · 2 generations on1descendant
- K151967Centros Long-Term Hemodialysis Catheter, CentrosFLO Long-Term Hemodialysis CatheterMerit Medical Systems, Inc. · 2016 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K981994 | Vascath Opti-Flow Long Term Dual Lumen Hemodialysis Catheter | C.R. Bard, Inc. | 1998 | — |
| −4 | K012562 | Medcomp 14.5F Double Lumen Hemodialysis Catheter | Medical Components, Inc. | 2002 | — |
| −3 | K020465 | Medcomp Ash Split-Cath II | Medcomp | 2002 | — |
| −2 | K022678 | Medcomp SC4 | Medcomp | 2003 | — |
| −1 | K040318 | Medcomp Split-Cath II | Medcomp | 2005 | — |
| 0 (this device) | K070572 | Ash Advance Hemodialysis Catheter and Procedure Kit, Models; 15FR X 24CM, 15FR… | Ash Access Technology, Inc. | 2007 | — |
| +1 | K080400 | Angiodynamics, Inc., Duramax Hemodialysis Catheter and Procedure Kit | AngioDynamics, Inc. | 2008 | 5 |
| +2 | K101843 | Duramax Chronic Hemodialysis Catheter and Procedure Kit | AngioDynamics, Inc. | 2010 | Class I |
| +1 | K092597 | Centros Chronic Hemodialysis Catheter Set | Ash Access Technology, Inc. | 2009 | — |
| +2 | K141363 | Centros Long Term Hemodialysis Catheter, Centrosflo Long Term Hemodialysis… | Merit Medical Systems, Inc. | 2014 | — |
| +3 | K151967 | Centros Long-Term Hemodialysis Catheter, CentrosFLO Long-Term Hemodialysis… | Merit Medical Systems, Inc. | 2016 | Class I |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 2 | 0 |
| 2001 | 3 | 0 |
| 2003 | 1 | 0 |
| 2004 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 1 |
| 2011 | 1 | 0 |
| 2014 | 2 | 1 |
| 2016 | 1 | 1 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MSD device
- #1 K241581 Hemodialysis Catheterscore 12.4 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K002901 | Kendall 14.5 FR Maxid Cuffed Dual Lumen CatheterSummary page 1…modialysis Catheter and Procedure Kit is substantially equivalent to the following devices: ICuffed Dual Lumen Catheter System yco/Kendall K002901, cleared T M o Mahurkar MAXIc Kendall/Tyc Healthcare… | 2001 |
| * | K040318 | Medcomp Split-Cath IISummary page 1…T 3/28/2001 medComp® Split-Cath® II/III Hemodialysis medComp® K040318, cleared Catheter 2/3/2005 AngioDynamies® Dynamic Flow TM Chron… | 2005 |
| * | K040402 | Angiodynamics, Inc. More-Flow Hemodialysis Catheter and Procedure KitSummary page 1…K040318, cleared Catheter 2/3/2005 AngioDynamies® Dynamic Flow TM Chronic AngioDynamics® K040402, cleared- Hemodialysis Catheter and Procedure Kit, medComp® 3/19/2004 medComp® ExcellWDynanic Flow Split… | 2004 |
| * | K040992 | 15F X 55CM Excell Split -Tip Catheter, Models 10300606, 10300706Summary page 1…40402, cleared- Hemodialysis Catheter and Procedure Kit, medComp® 3/19/2004 medComp® ExcellWDynanic Flow Split-Tip K040992, cleared Catheter 7/29/2004 Device Description: The Ash Ad… | 2004 |
What descends from it? (7)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (3 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 7, 2008 | Not classified | Z-2397-2008 | Process design | The cuff may be inadequately attached to the catheter, resulting in possible catheter movement or leakage at the insertion site. |
| Jul 7, 2008 | Not classified | Z-2398-2008 | Process design | The cuff may be inadequately attached to the catheter, resulting in possible catheter movement or leakage at the insertion site. |
| Jul 7, 2008 | Not classified | Z-2396-2008 | Process design | The cuff may be inadequately attached to the catheter, resulting in possible catheter movement or leakage at the insertion site. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K101843Duramax Chronic Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2010 · 2 generations on · ◆ in the 2010 descendants bar of FIG. 11 Class I recall
- K131260Nmi Dialysis CatheterNavilyst Medical, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
- K151967Centros Long-Term Hemodialysis Catheter, CentrosFLO Long-Term Hemodialysis CatheterMerit Medical Systems, Inc. · 2016 · 3 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 9 years and passes through 4 companies; a new generation was cleared every 1.9 years on median.
- 3 of the 13 ancestors have a recall of any class.
- Of its descendants, K080400 (AngioDynamics, Inc., 2008) has gone furthest, with 6 descendants of its own.
- 1 other device also cites its predicate K002901.
How does its category compare?
All MSD devicesMSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · MSD
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MSD
153in 2021
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MSD
1in 2013
Class I (4 since 2007) Class II/III not classified
Adverse event reports per year · MSD
1,054in 2025
MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 92 | 0 | 118 |
| 2008 | 3 | 89 | 47 | 174 |
| 2009 | 6 | 81 | 49 | 173 |
| 2010 | 4 | 116 | 11 | 306 |
| 2011 | 7 | 70 | 24 | 268 |
| 2012 | 2 | — | 0 | 367 |
| 2013 | 5 | 105 | 1 | 363 |
| 2014 | 8 | 246 | 0 | 506 |
| 2015 | 1 | — | 0 | 550 |
| 2016 | 2 | — | 0 | 532 |
| 2017 | 2 | — | 0 | 484 |
| 2018 | 2 | — | 0 | 463 |
| 2019 | 1 | — | 0 | 645 |
| 2020 | 4 | 106 | 0 | 401 |
| 2021 | 4 | 153 | 0 | 523 |
| 2022 | 0 | — | 0 | 610 |
| 2023 | 0 | — | 0 | 542 |
| 2024 | 0 | — | 0 | 841 |
| 2025 | 1 | — | 0 | 1,054 |
| 2026 | 0 | — | 4 (4) | — |
What is it indicated for?
The Ash AdvanceTM Hemodialysis Catheter and Procedure Kit is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. This catheter is indicated for > 30 days long-term placement.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K070572 · printed Nov 4, 2026
Cite this page
510(k) Report. “K070572: Ash Advance Hemodialysis Catheter and Procedure Kit, Models; 15FR X 24CM, 15FR X 28CM, 15FR X 32CM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K070572