510(k) Report

510(k) · Decided

Recalled · Class I · 2023

Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid (TM) Dual Lumen Catheter, Palindrome (TM)...

K111372510(k) SpecialDecided Aug 4, 201180-day reviewClass II21 CFR 876.5540Recalls 3 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Covidien · MSD Catheter, Hemodialysis, Implanted

Cleared on Aug 4, 2011 as substantially equivalent to K043272 (Kendall, 2004) and 2 other predicates after an 80-day review. Its family tree reaches back 3 generations to 2001. 3 of the 16 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it. It has been recalled 3 times, including 2 Class I recalls.

Decided
Aug 4, 2011
Review
80 days
Class
Class II
Product code
MSD
Recalls
3 (2 Class I)
Predicates
3

Its longest line of descent runs back 3 generations to K010399 (2001). 5 ancestors in all, the oldest K002901 (2001). 10 later devices descend from it, 4 citing it directly. 3 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2001: 1 other ancestor2019: 1 descendant cleared2011: 2 descendants cleared, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall12014: 1 descendant cleared2021: 1 descendant cleared2017: 1 descendant cleared2025: 1 descendant cleared2020: 1 descendant cleared2018: 1 descendant clearedK010399 · Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC, CC3200PC (32 CM), CC3200PC (36 CM) · Diatek, Inc. · 2001 · recalledK040078 · Cannon II Plus · Arrow International, Inc. · 2004 · recalledK092205 · Palindrome P Hemodialysis Catheter · Covidien Lp, Formerly Registered AS Kendall · 2009200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 4 descendants, 2 with a Class I recall22 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants K092205 · Palindrome P Hemodialysis Catheter · Covidien Lp, Formerly Registered AS Kendall · 2009K040078 · Cannon II Plus · Arrow International, Inc. · 2004 · recalledK010399 · Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC, CC3200PC (32 CM), CC3200PC (36 CM) · Diatek, Inc. · 2001 · recalled0
  1. Root · 3 generations backK010399 Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC…Diatek, Inc. · 2001 · Recalled
  2. 2 generations backK040078 Cannon II PlusArrow International, Inc. · 2004 · Recalled
  3. 1 generation backK092205 Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall · 2009
  4. This deviceK111372 Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R)…Covidien · 2011
  5. Descendant · cites it directly · 5 of its ownK112477 Palindrome H, SI, Hsi Chronic CatheterCovidien · 2011 · Recalled (Class I)
  6. Descendant · cites it directly · 4 of its ownK123196 Palindrome Precision H Chronic Catheter, Palindrome Precision SI…Covidien · 2013 · Recalled (Class I)
  7. Descendant · 2 generations on · 3 of its ownK140884 Symetrex Long Term Hemodialysis CatheterPhase One Medical · 2014
  8. Descendant · cites it directly · 2 of its ownK182443 Pristine Hemodialysis CatheterPristine Access Technologies, Ltd. · 2019
  9. Descendant · 3 generations on · 2 of its ownK171618 Symetrex Long Term Hemodialysis CatheterMedcomp (Dba Medical Components, Inc.) · 2017

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2011: 22013: 12014: 12017: 12018: 12019: 12020: 12021: 12025: 1
2011high 2 (2011) · low 0 (2012)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K112477Palindrome H, SI, Hsi Chronic CatheterCovidien · 2011 · cites it directly5descendants
  • K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic Catheter…Covidien · 2013 · cites it directly4descendants
  • K140884Symetrex Long Term Hemodialysis CatheterPhase One Medical · 2014 · 2 generations on3descendants
  • K182443Pristine Hemodialysis CatheterPristine Access Technologies, Ltd. · 2019 · cites it directly2descendants
  • K171618Symetrex Long Term Hemodialysis CatheterMedcomp (Dba Medical Components, Inc.) · 2017 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K010399Diatek Cannon-Cath #CC2400, CC2800, CC3200, CC3600,CC5500. CC2400PC, CC2800PC…Diatek, Inc.20012
−2K040078Cannon II PlusArrow International, Inc.200419
−1K092205Palindrome P Hemodialysis CatheterCovidien Lp, Formerly Registered AS Kendall2009—
0 (this device)K111372Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid…Covidien2011—
+1K112477Palindrome H, SI, Hsi Chronic CatheterCovidien2011Class I
+1K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic…Covidien2013Class I
+2K140884Symetrex Long Term Hemodialysis CatheterPhase One Medical2014—
+1K182443Pristine Hemodialysis CatheterPristine Access Technologies, Ltd.2019—
+3K171618Symetrex Long Term Hemodialysis CatheterMedcomp (Dba Medical Components, Inc.)2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201121
201311
201410
201710
201810
201910
202010
202110
202510
FIG. 1 — Lineage of K111372, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MSD device

  1. #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043272Pallindrome 14.5FR Chronic Hemodialysis Catheter with Slotted Symmetrical Tip
Summary page 1
…on Number: 21 CER 876.5540 Product Code: MSD Classification: Class Ill 3. Predicate Devices: Tat PalindromeTM Symmetric Tip Dual Lumen Catheter (K043272) Mahurkar® MaxidTM Dual Lumen Catheter (K002901) PalindromeTM RT Reverse-Tunneled Catheter (K092205) 4. Device…
Kendall2004—
*K002901Kendall 14.5 FR Maxid Cuffed Dual Lumen Catheter
Summary page 1
…Classification: Class Ill 3. Predicate Devices: Tat PalindromeTM Symmetric Tip Dual Lumen Catheter (K043272) Mahurkar® MaxidTM Dual Lumen Catheter (K002901) PalindromeTM RT Reverse-Tunneled Catheter (K092205) 4. Device Description: The Tal PalindromeTM Symmetric Ti…
The Kendall Company2001—
*K092205Palindrome P Hemodialysis Catheter
Summary page 1
…Tat PalindromeTM Symmetric Tip Dual Lumen Catheter (K043272) Mahurkar® MaxidTM Dual Lumen Catheter (K002901) PalindromeTM RT Reverse-Tunneled Catheter (K092205) 4. Device Description: The Tal PalindromeTM Symmetric Tip Dual Lumen Catheter has a radiopaque shaft with two large i…
Covidien Lp, Formerly Registered AS Kendall2009—

Also cleared against K043272: 3 other devices →

What descends from it? (10)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112477Palindrome H, SI, Hsi Chronic Catheter Class ICovidien20115
K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic Catheter… Class ICovidien20134
K182443Pristine Hemodialysis CatheterPristine Access Technologies, Ltd.20192
K111817Palindrome RT Chronic CatheterCovidien20110

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 22, 2023Class IIZ-2632-2023Under Investigation by firmSome catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.
Jun 8, 2022Class IZ-1320-2022Process controlThere is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.
Jun 8, 2022Class IZ-1316-2022Process controlThere is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K112477Palindrome H, SI, Hsi Chronic CatheterCovidien · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K123196Palindrome Precision H Chronic Catheter, Palindrome Precision SI Chronic Catheter, Palindrome Precision Hsi…Covidien · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 10 years and passes through 4 companies; a new generation was cleared every 2.5 years on median.
  • 2 of the 5 ancestors have a recall of any class.
  • Of its descendants, K112477 (Covidien, 2011) has gone furthest, with 5 descendants of its own.
  • 3 other devices also cite its predicate K043272.

How does its category compare?

All MSD devices

MSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · MSD

1in 2025

2007: 72008: 32009: 62010: 42011: 72012: 22013: 52014: 82015: 12016: 22017: 22018: 22019: 12020: 42021: 42025: 1
2007high 8 (2014) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MSD

153in 2021

2007: 92 days median2008: 89 days median2009: 81 days median2010: 116 days median2011: 70 days median2013: 105 days median2014: 246 days median2020: 106 days median2021: 153 days median
2007high 246 (2014) · low 70 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MSD

1in 2013

2008: 472009: 492010: 112011: 242013: 12026 (partial year): 4, 4 Class I
2007high 49 (2009) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · MSD

1,054in 2025

2007: 1182008: 1742009: 1732010: 3062011: 2682012: 3672013: 3632014: 5062015: 5502016: 5322017: 4842018: 4632019: 6452020: 4012021: 5232022: 6102023: 5422024: 8412025: 1,054
2007high 1,054 (2025) · low 118 (2007)2025

MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077920118
200838947174
200968149173
2010411611306
201177024268
20122—0367
201351051363
201482460506
20151—0550
20162—0532
20172—0484
20182—0463
20191—0645
202041060401
202141530523
20220—0610
20230—0542
20240—0841
20251—01,054
20260—4 (4)—

What is it indicated for?

The device is indicated for acute and chronic heinodialysis, apheresis, and infusion. It may be inserted percutaneously or by cut down. Catheters with greater than 40 cm implant length are indicated for femnoral placement.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111372: Tal Pallindrome (TM) Symmetric Tip Dual Lumen Catheter, Mahurkar (R) Maxid (TM) Dual Lumen Catheter, Palindrome (TM)....” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111372

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026