510(k) · Decided
Recalled · Class I · 2026Duramax Chronic Hemodialysis Catheter and Procedure Kit
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
AngioDynamics, Inc. · MSD Catheter, Hemodialysis, Implanted
Cleared on Oct 20, 2010 as substantially equivalent to K080400 (AngioDynamics, Inc., 2008) and 1 other predicate after a 111-day review. Its family tree reaches back 7 generations to 1998. 2 of the 18 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 30 times, including a Class I recall.
- Decided
- Oct 20, 2010
- Review
- 111 days
- Class
- Class II
- Product code
- MSD
- Recalls
- 30 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 7 generations to K981994 (1998). 15 ancestors in all, the oldest K972207 (1998). 2 later devices descend from it, 2 citing it directly. 2 of the 18 devices in this family, counting this one, have a Class I recall.
- Root · 7 generations backK981994 Vascath Opti-Flow Long Term Dual Lumen Hemodialysis CatheterC.R. Bard, Inc. · 1998
- 6 generations backK012562 Medcomp 14.5F Double Lumen Hemodialysis CatheterMedical Components, Inc. · 2002
- 5 generations backK020465 Medcomp Ash Split-Cath IIMedcomp · 2002
- 4 generations backK022678 Medcomp SC4Medcomp · 2003
- 3 generations backK040318 Medcomp Split-Cath IIMedcomp · 2005
- 2 generations backK070572 Ash Advance Hemodialysis Catheter and Procedure Kit, Models; 15FR X…Ash Access Technology, Inc. · 2007 · Recalled
- 1 generation backK080400 Angiodynamics, Inc., Duramax Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2008 · Recalled
- This deviceK101843 Duramax Chronic Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2010
- Descendant · cites it directly · 0 of its ownK131260 Nmi Dialysis CatheterNavilyst Medical, Inc. · 2014 · Recalled (Class I)
- Descendant · cites it directly · 0 of its ownK110936 Duraflow 2 Chronic Hemodialysis Catheter and Procedure KitAngioDynamics, Inc. · 2011 · Recalled
Class I recalls in this family
- K131260 Nmi Dialysis Cathetercites it directly · 2014
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −7 | K981994 | Vascath Opti-Flow Long Term Dual Lumen Hemodialysis Catheter | C.R. Bard, Inc. | 1998 | — |
| −6 | K012562 | Medcomp 14.5F Double Lumen Hemodialysis Catheter | Medical Components, Inc. | 2002 | — |
| −5 | K020465 | Medcomp Ash Split-Cath II | Medcomp | 2002 | — |
| −4 | K022678 | Medcomp SC4 | Medcomp | 2003 | — |
| −3 | K040318 | Medcomp Split-Cath II | Medcomp | 2005 | — |
| −2 | K070572 | Ash Advance Hemodialysis Catheter and Procedure Kit, Models; 15FR X 24CM, 15FR… | Ash Access Technology, Inc. | 2007 | 3 |
| −1 | K080400 | Angiodynamics, Inc., Duramax Hemodialysis Catheter and Procedure Kit | AngioDynamics, Inc. | 2008 | 5 |
| 0 (this device) | K101843 | Duramax Chronic Hemodialysis Catheter and Procedure Kit | AngioDynamics, Inc. | 2010 | — |
| +1 | K131260 | Nmi Dialysis Catheter | Navilyst Medical, Inc. | 2014 | Class I |
| +1 | K110936 | Duraflow 2 Chronic Hemodialysis Catheter and Procedure Kit | AngioDynamics, Inc. | 2011 | 6 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 2 | 0 |
| 2001 | 3 | 0 |
| 2003 | 1 | 0 |
| 2004 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2014 | 1 | 1 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MSD device
- #1 K241581 Hemodialysis Catheterscore 12.3 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K080400 | Angiodynamics, Inc., Duramax Hemodialysis Catheter and Procedure KitSummary page 1…Manufacture~r 510(k) No.., AngloDynamics% DuraMax" Catheter Kits AngioDyna-mi-cs K080400, cleared _____ ____ ____ ____ ________ ____ ___ ____ _… | 2008 |
| * | K040402 | Angiodynamics, Inc. More-Flow Hemodialysis Catheter and Procedure KitSummary page 1…____ ________ ____ ___ ____ ____ ____ ____ 5/13/08 AngioDynarnicsT EvenMore"' Chronic AngioDynamics"' K040402, cleared H-emodialysis Catheter 4/19/2004 Device Des… | 2004 |
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 13, 2026 | Class I | Z-1578-2026 | Process control | 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for… |
| Mar 18, 2024 | Class II | Z-1771-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Mar 18, 2024 | Class II | Z-1772-2024 | Under Investigation by firm | XXX |
| Mar 18, 2024 | Class II | Z-1766-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Mar 18, 2024 | Class II | Z-1767-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Mar 18, 2024 | Class II | Z-1768-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Mar 18, 2024 | Class II | Z-1769-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Mar 18, 2024 | Class II | Z-1773-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Mar 18, 2024 | Class II | Z-1770-2024 | Under Investigation by firm | Tri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure. |
| Dec 30, 2011 | Not classified | Z-0719-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0720-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0721-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0724-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0725-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0726-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0729-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0730-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0731-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0732-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0718-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0722-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0723-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0727-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0728-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Dec 30, 2011 | Not classified | Z-0733-2012 | Nonconforming Material/Component | The securement cuff on the hemodialysis catheter has the potential to detach from the shaft. |
| Jan 24, 2011 | Not classified | Z-0588-2012 | Nonconforming Material/Component | AngioDynamics received customer complaints for breakage of the tunneler sleeve during product use. |
| Jan 24, 2011 | Not classified | Z-0589-2012 | Nonconforming Material/Component | AngioDynamics received customer complaints for breakage of the tunneler sleeve during product use. |
| Jan 24, 2011 | Not classified | Z-0586-2012 | Nonconforming Material/Component | AngioDynamics received customer complaints for breakage of the tunneler sleeve during product use. |
| Jan 24, 2011 | Not classified | Z-0590-2012 | Nonconforming Material/Component | AngioDynamics received customer complaints for breakage of the tunneler sleeve during product use. |
| Jan 24, 2011 | Not classified | Z-0587-2012 | Nonconforming Material/Component | AngioDynamics received customer complaints for breakage of the tunneler sleeve during product use. |
The 5 latest of 30 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K131260Nmi Dialysis CatheterNavilyst Medical, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 12 years and passes through 5 companies; a new generation was cleared every 1.9 years on median.
- 5 of the 15 ancestors have a recall of any class.
- Of its descendants, K131260 (Navilyst Medical, Inc., 2014) has gone furthest, with 0 descendants of its own.
- 1 other device also cites its predicate K080400.
How does its category compare?
All MSD devicesMSD Catheter, Hemodialysis, Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.
FDA decisions per year · MSD
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MSD
153in 2021
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MSD
1in 2013
Class I (4 since 2007) Class II/III not classified
Adverse event reports per year · MSD
1,054in 2025
MAUDE reports naming product code MSD, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 92 | 0 | 118 |
| 2008 | 3 | 89 | 47 | 174 |
| 2009 | 6 | 81 | 49 | 173 |
| 2010 | 4 | 116 | 11 | 306 |
| 2011 | 7 | 70 | 24 | 268 |
| 2012 | 2 | — | 0 | 367 |
| 2013 | 5 | 105 | 1 | 363 |
| 2014 | 8 | 246 | 0 | 506 |
| 2015 | 1 | — | 0 | 550 |
| 2016 | 2 | — | 0 | 532 |
| 2017 | 2 | — | 0 | 484 |
| 2018 | 2 | — | 0 | 463 |
| 2019 | 1 | — | 0 | 645 |
| 2020 | 4 | 106 | 0 | 401 |
| 2021 | 4 | 153 | 0 | 523 |
| 2022 | 0 | — | 0 | 610 |
| 2023 | 0 | — | 0 | 542 |
| 2024 | 0 | — | 0 | 841 |
| 2025 | 1 | — | 0 | 1,054 |
| 2026 | 0 | — | 4 (4) | — |
What is it indicated for?
The AngioDynamics, Inc. Duralvax~i Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for Hemodialysis and apheresis.
It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient.
Alternative insertion sites include the subclavian vein as required.
Catheters greater than 40cm are intended for femoral vein insertion.
This catheter is indicated for > 30 days long-term placement.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K101843 · printed Nov 6, 2026
Cite this page
510(k) Report. “K101843: Duramax Chronic Hemodialysis Catheter and Procedure Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101843
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026